Aromatic hydrocarbons - Air-toxics analysis in chemical plant

Applications | 2011 | Agilent TechnologiesInstrumentation
GC/MSD, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies

Importance of the Topic


Rapid and accurate monitoring of volatile aromatic hydrocarbons and 1,3-butadiene in chemical plant atmospheres is critical for ensuring worker safety, regulatory compliance, and environmental protection. Traditional methods often require lengthy analysis times, delaying decision-making and response to potential air-toxic exposures.

Study Objectives and Overview


This application note presents a fast, selective gas chromatography–mass spectrometry method for simultaneous determination of a range of aromatic hydrocarbons and 1,3-butadiene collected from plant air using carbon-disc badges. The primary goal is to reduce analysis time while maintaining sensitivity and resolution.

Methodology and Instrumentation Used


A wide-bore GC system equipped with an Agilent Lowox 0.53 mm × 10 m fused-silica PLOT column (part no. CP8587) was employed. The temperature program ramped from 60 °C to 250 °C at 20 °C/min. Helium carrier gas was maintained at 20 kPa (0.2 bar) in constant pressure mode. A split injector (1:20) at 150 °C introduced samples prepared in carbon disulfide at 1 µg/mL. Detection was performed using mass spectrometry to achieve high selectivity. Airborne compounds were preconcentrated on 3M carbon-disc badges, achieving a detection limit of 0.15 ppm.

Main Results and Discussion


The optimized method achieved baseline separation of eight target analytes—including 1,3-butadiene, benzene, toluene, dicyclopentadiene, xylenes, and styrene—within an 11-minute runtime, a reduction from the typical 25–30 minutes. The MS detector also revealed trace artifacts such as limonene, acetone, and alcohols. Reproducibility and sensitivity met industry requirements for routine air-toxic surveillance.

Benefits and Practical Applications


  • Run time reduced by over 50 %, increasing sample throughput.
  • High selectivity for structurally similar aromatics and 1,3-butadiene.
  • Low detection limit (0.15 ppm) suitable for workplace monitoring.
  • Simple sampling using 3M carbon-disc badges facilitates on-site collection.
  • Method can be integrated into QA/QC protocols for chemical and petrochemical facilities.

Future Trends and Potential Applications


Advances in column materials and micro-GC technologies promise even faster separations. Integration with time-of-flight MS or alternative detectors could enhance mass accuracy and dynamic range. Online, real-time monitoring systems and wireless sensor networks may leverage this approach for continuous air-toxic surveillance. Machine learning algorithms applied to spectral data are likely to further improve compound identification and quantification.

Conclusion


The fast GC–MS method detailed here offers a robust, efficient solution for monitoring aromatic hydrocarbons and 1,3-butadiene in chemical plant environments. The significant reduction in analysis time, coupled with high sensitivity and selectivity, supports improved safety practices and regulatory compliance.

References


  • Agilent Technologies, Inc. (2011). Application Note A01652: Aromatic hydrocarbons Air-toxics analysis in chemical plant. First published prior to May 11, 2010; printed October 31, 2011.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Solutions for light hydrocarbons and gasses: PLOT columns
Solutions for light hydrocarbons and gasses: PLOT columns
2019|Agilent Technologies|Presentations
PLOT Columns Separation Solutions for  Light Hydrocarbons  & Gases
Laboratory-Based Air Monitoring Solutions from Agilent Technologies
Downstream Petrochemical Processes - Application Compendium