Air Sampling of VOCs by SPME for Analysis by Capillary GC

Applications | 1998 | MerckInstrumentation
GC/MSD, SPME, GC/IT
Industries
Environmental
Manufacturer
Merck

Significance of the Topic


Volatile organic compounds VOCs in air and water pose health and environmental risks requiring reliable field sampling methods

Objectives and Study Overview


This study evaluates a portable SPME field sampler with a Carboxen PDMS 75µm fiber for sampling and storing VOCs under field and laboratory conditions

Methodology and Instrumentation


Air and water samples were extracted following US EPA Method 624 at 20ppb and in a nine compound mixture covering 400ppt to 400ppb range
  • SPME portable field sampler with Carboxen PDMS 75µm fiber
  • Storage conditions: sealed at –4 °C sealed ambient and unsealed ambient for three days
  • Desorption: five minutes at 300 °C
  • Capillary GC with SPB-1 Sulfur column 30 m × 0.32 mm ID, 4.0 µm film and GC/MS ion trap detection m/z 45–260

Key Results and Discussion


The Carboxen PDMS fiber retained analytes with negligible loss at –4 °C sealed and average loss less than 15 % when stored unsealed at ambient for three days, compared with over 90 % loss on a 100 µm PDMS fiber
Calibration curves for nine VOCs showed excellent linearity (R² ≥ 0.98) except 1,2-dichloroethane (R² = 0.953)
The unique pore structure of Carboxen-1006 prevented displacement of lighter compounds even at higher concentrations

Benefits and Practical Applications


The sampler eliminates gas bag shipment, simplifies field sampling and enables delayed lab analysis of trace level VOCs in air and water, providing a convenient and reliable workflow

Future Trends and Opportunities


Advances may include integration with portable GC/MS, real-time sensors, expanded adsorbent coatings and automated sampling platforms to broaden analyte coverage and improve on-site decision making

Conclusion


The portable SPME sampler with Carboxen PDMS fiber offers robust VOC sampling and storage capabilities supporting accurate trace analysis and streamlined field-to-lab workflows

Reference


US EPA Method 624
Sigma Aldrich Application Note 141
US Patent 5691206 and European Patent 0523092

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Air Sampling of VOCs Using New SPME Portable Field Sampler
SPME Portable Field Sampler with Carboxen/PDMS Fiber
SPME Portable Field Sampler with Carboxen/PDMS Fiber
1997|Merck|Brochures and specifications
Solid Phase Microextraction: Theory and Optimization of Conditions
SPME/GC Analyses of Sulfur Gases and VOCs, Using a New Carboxen/PDMS Fiber