GC/MSD, Sample Preparation, LC/MS
IndustriesFood & Agriculture
ManufacturerCTC Analytics
Significance of the Topic
QuEChERS extraction is a globally accepted approach for pesticide residue analysis in food matrices. Despite standardized extraction protocols, the clean-up of complex extracts remains a challenge. An automated micro-solid phase extraction (µSPE) workflow enhances throughput, reproducibility, and sustainability in GC-MS and LC-MS analyses.
Objectives and Overview of the Study
- Establish a unified, automated clean-up method for QuEChERS extracts using µSPE cartridges.
- Integrate the workflow on a PAL RTC autosampler compatible with GC-MS and LC-MS platforms.
- Validate performance across a range of simple, complex, and high-lipid food matrices for multi-residue pesticide analysis.
Methodology and Instrumentation
- Sample Preparation: Standard QuEChERS extraction followed by centrifugation to yield a clear raw extract.
- Automated µSPE Clean-up: Miniaturized cartridges (33 mm × 8 mm, 45 mg sorbent) conditioned optionally, loaded with 150 µL extract at ~2 µL/s, and eluted with ~100 µL of 100 mM formate buffer in acetonitrile at the same flow rate.
- Post-Clean-up Processing: Addition of 20 µL analyte protectant, 5 µL internal standard, and 20 µL acetonitrile followed by mixing and direct injection.
- Instrumentation: PAL RTC autosampler with syringe-based flow control, coupled to GC-MS and LC-MS systems for detection.
Main Results and Discussion
- Linearity demonstrated with R² ≥ 0.99 for all 208 target pesticides.
- Recoveries ranged between 70 % and 130 % with RSDs below 10 %, meeting regulatory criteria.
- µSPE performance surpassed traditional dispersive SPE in matrix removal and analyte purity.
- No evaporation step required, reducing processing time and solvent use.
- Parallel sample preparation during chromatographic runs increased overall throughput.
Benefits and Practical Applications
- A single cartridge type accommodates diverse food matrices, simplifying lab workflows.
- Automated operation on the autosampler minimizes manual handling and variability.
- Green chemistry advantages: reduced solvent consumption and waste.
- Seamless integration with existing GC-MS and LC-MS analytical setups for regulatory and quality control laboratories.
Future Trends and Potential Applications
- Extension of µSPE clean-up to other contaminant classes such as veterinary drugs or mycotoxins.
- Coupling with high-resolution mass spectrometry for non-targeted screening and emerging residues.
- Further miniaturization for field-deployable or on-site testing platforms.
- Data-driven optimization of sorbent chemistries and flow parameters tailored to specific matrix challenges.
Conclusion
The automated µSPE-based clean-up workflow offers a robust, green, and universal solution for QuEChERS extracts. It provides high recoveries, excellent reproducibility, and superior matrix removal across diverse food samples while streamlining GC-MS and LC-MS multi-residue pesticide analysis.
Reference
- Bruce D. Morris & Richard B. Schriner, Journal of Agricultural and Food Chemistry, 2015, 63, 5107–5119.
- Steven J. Lehotay et al., Chromatographia, 2016; DOI: 10.1007/s10337-016-3116-y.
- M. Hayward & J. Ho, American Laboratory, 2016.
- H.-J. Hübschmann et al., Poster at LAPRW 2017, San Jose, Costa Rica.