GC/MSD, GC/SQ, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Importance of the topic
Capillary gas chromatography analysis of semi-volatile environmental analytes requires high column inertness to avoid peak tailing, adsorption and loss of sensitivity for active compounds such as benzoic acid, 2,4-dinitrophenol, aniline and benzidine. Consistent inert performance ensures reliable quantification at trace levels and reduces reanalysis costs.
Study objectives and overview
This work compares inertness performance of two premium 20 m × 0.18 mm × 0.18 μm 5 ms GC columns: Agilent J&W HP-5ms Ultra Inert and Vendor R Rxi-5ms. The study aims to evaluate peak shapes, recovery and relative response ratio (RRR) for a standard semi-volatile mix modeled on EPA Method 8270D.
Methodology and used instrumentation
- GC system: Agilent 6890N network GC with 5975B MSD detector
- Sampler: Agilent 7683B autosampler, splitless injection (0.5 μL)
- Inlet: Deactivated direct connect liner at 280 °C
- Carrier gas: Helium ramped from 0.7 to 1.3 mL/min
- Oven program: initial 35 °C, ramp to 300 °C in multi-step sequence
- Columns: two 20 m × 0.18 mm × 0.18 μm formats from Agilent and Vendor R
- MSD conditions: source 300 °C, quad 180 °C, scan 50–550 m/z
Main results and discussion
- Agilent column delivered sharp, symmetric peaks for acidic and basic analytes at 0.5 ng loading, while Vendor R showed pronounced tailing and reduced responses.
- RRR for 2,4-dinitrophenol on Agilent columns averaged 0.192–0.198, well above the 0.05 minimum, compared to 0.144–0.146 for Vendor R, indicating faster loss of inertness on Vendor R.
- Faster analysis: the 5 ms phase column provided equivalent resolution to a conventional 30 m × 0.25 mm column with 30–40% time savings.
Benefits and practical applications
- Reliable peak shapes and quantitation for active environmental contaminants.
- Improved throughput and lower per-sample cost by using high-efficiency fast columns.
- Reduced reanalysis and maintenance by ensuring consistent inertness through rigorous testing.
Future trends and application opportunities
- Development of even more inert stationary phases and advanced surface treatments.
- Integration with software for automated method translation and real-time inertness monitoring.
- Expansion of fast high-efficiency columns to other compound classes and multidimensional GC workflows.
Conclusion
A direct comparison demonstrates that Agilent J&W HP-5ms Ultra Inert columns maintain superior inertness, peak shape and sensitivity for semi-volatile analytes compared to Vendor R Rxi-5ms. Fast 5 ms columns offer robust performance and increased laboratory productivity for environmental GC/MS analyses.
References
- US EPA Method 8270D Revision 4, Semivolatile Organic Compounds by GC/MS, 2007.
- M. Hastings et al., Inertness Comparison of 5% Phenyldimethylpolysiloxane Columns, Pittcon, 2003.
- J. Luong et al., Advanced Solventless Column Test for Capillary GC Columns, J. Sep. Sci., 2007, 30, 2480–2492.
- Agilent J&W Ultra Inert GC Columns Technical Overview, 5989-8685EN, 2008.
- Agilent Method Translation Software, www.agilent.com/chem/gcmethodtranslation.