GC/MSD, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies
Importance of Topic
Chloramination in water treatment can generate nitrosamines such as NDMA, classified as probable human carcinogens. Monitoring these trace contaminants in drinking water at sub-part-per-trillion levels is essential for public health protection and compliance with regulatory guidelines worldwide.
Objectives and Overview of Study
This study evaluates the Agilent 5977A Series GC/MSD System for detecting six nitrosamines in drinking water. The aim is to achieve limits of detection below those cited in EPA Method 521 without large-volume injection or MS/MS instrumentation.
Methodology and Instrumentation Used
Sample Preparation:
- 500 mL water samples extracted by solid-phase extraction following EPA 521 procedures.
- NDMA-d6 surrogate added pre-extraction; NDPA-d14 internal standard added post-concentration.
- Elution in methylene chloride; concentrated to 1 mL (500× enrichment).
- GC: Agilent 7890B with multimode inlet (MMI).
- MS: Agilent 5977A MSD in positive chemical ionization (PCI) mode using methane.
- Column: VF-624ms (20 m × 0.15 mm, 0.84 µm).
- Injection: 2 µL splitless; inlet at 250 °C.
- Oven: 35 °C (4 min) →195 °C at 10 °C/min →240 °C at 15 °C/min (2 min hold).
- Acquisition: Full scan m/z 40–200 and SIM for target/qualifier ions.
Main Results and Discussion
Calibration and Linearity:
- 0.5–50 ppb in methylene chloride; correlation coefficients ≥0.9985.
- LODs (3.3σ) ranged 0.018–0.166 ppb in extract (equivalent to 0.036–0.332 ppt in water).
- LOQs (10× LOD) ranged 0.182–1.662 ppb.
- Agilent GC/MSD LODs were 1.9–11× lower than EPA 521 despite a 10× smaller injection and use of single-quadrupole MS.
Benefits and Practical Applications
The method enables:
- High-sensitivity quantitation of nitrosamines without MS/MS.
- Smaller injection volumes for simpler operation.
- Reproducible results via automated sample preparation.
Future Trends and Possibilities
Potential developments include:
- On-line SPE integration for higher throughput.
- PCI/MS adaptation to other disinfection by-products.
- High-resolution mass spectrometry for enhanced selectivity.
Conclusion
The Agilent 5977A GC/MSD system in PCI mode offers reliable sub-ppt detection of nitrosamines, outperforming EPA Method 521 LODs with minimal injection volume and single-quadrupole instrumentation.
Used Instrumentation
- Agilent 7890B GC with MMI.
- Agilent 5977A MSD in PCI mode.
- Agilent 7696A Sample Prep WorkBench.
References
- EPA Method 521: Determination of Nitrosamines in Drinking Water by SPE-GC/MS/MS.
- Health Canada. Guidelines for Canadian Drinking Water Quality: NDMA.
- World Health Organization. Guidelines for Drinking-water Quality (2008).
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.