A simplified approach for nicotine quantification in electronic cigarette liquids using GC-Orbitrap mass spectrometry

Applications | 2019 | Thermo Fisher ScientificInstrumentation
GC/MSD, GC/HRMS, GC/Orbitrap
Industries
Food & Agriculture
Manufacturer
Thermo Fisher Scientific

Significance of the topic


The growing use of electronic cigarettes has created a critical need for reliable nicotine quantification methods to protect consumer health and comply with regulatory limits. High-resolution GC-Orbitrap mass spectrometry delivers the sensitivity and specificity necessary to analyze complex e-liquid formulations and ensure adherence to EU Tobacco Products Directive guidelines.

Objectives and Study Overview


This work evaluates the Thermo Scientific Exactive GC Orbitrap system for quantitative analysis of nicotine in commercial e-liquids. Key goals include assessing calibration linearity, mass accuracy, repeatability, and the method’s applicability to both regulated and unregulated products.

Methodology


E-liquid samples with declared nicotine contents of 0, 6, or 12 mg/mL—including shortfill products containing 0 mg/mL—were diluted with LC/MS-grade acetonitrile and spiked with 8-hydroxyquinoline as internal standard. Calibration standards ranged from 46 to 13 792 ng/mL nicotine. Direct liquid split/splitless injection was performed in electron ionization mode. Data were acquired in full-scan at 60 000 resolution, and compounds were identified by retention time, accurate mass (±5 ppm), and ion ratio criteria.

Used Instrumentation


  • Exactive GC Orbitrap mass spectrometer
  • TRACE 1310 gas chromatograph
  • TriPlus RSH autosampler
  • Instant Connect split/splitless injector
  • TraceGOLD TG-WaxMS capillary column (30 m×0.25 mm×0.25 μm)

Key Results and Discussion


Chromatographic separation yielded clear nicotine and internal standard peaks at 13.77 min. Calibration exhibited excellent linearity (R²=0.9991) across 11 levels with residuals under 4.4%. Mass accuracy remained below 1 ppm for all ions. Repeatability tests on 6 mg/mL samples showed peak area RSDs <4% and concentration RSDs <3%. Analysis of ten commercial e-liquids revealed nicotine levels differing from label claims by –8% to +21%, highlighting variability in unregulated shortfill products.

Benefits and Practical Applications


This GC-Orbitrap approach offers rapid, high-precision nicotine quantitation for regulatory laboratories and industry quality control. The full-scan data supports retrospective screening for additional flavoring agents or impurities without reanalysis.

Future Trends and Potential Applications


Emerging developments may integrate automated multi-component screening workflows and machine learning for data interpretation. Expanding non-targeted analyses could enable comprehensive profiling of e-liquid constituents and support evolving regulatory requirements.

Conclusion


The Exactive GC Orbitrap method provides robust quantitative performance for nicotine in e-liquids, combining rapid calibration, exceptional mass accuracy, and excellent precision. It is a reliable platform for ensuring consumer safety and compliance with regulatory standards.

References


  1. United States Department of Health and Human Services, How Tobacco Smoke Causes Disease, 2010.
  2. Goriounova NA, Mansvelder HD, Cold Spring Harbor Perspect Med, 2012.
  3. WHO Framework Convention on Tobacco Control, 2003.
  4. EU Tobacco Products Directive 2014/14/EU.
  5. MHRA e-cigarette regulations for consumer products.
  6. Flora JW et al., J Liq Chromatogr Relat Technol, 2016.
  7. Liu X et al., Beiträge Tabakforsch Int, 2017.
  8. Aszyk J et al., J Chromatogr A, 2017.
  9. Herrington BJ et al., Restek Application Note, 2015.
  10. CORESTA RM 62, 2005.
  11. Ghalop VV et al., J Anal Methods Chem, 2018.
  12. Cooper J et al., Thermo Scientific App Note 10708, 2019.
  13. Davis B et al., Nicotine Tob Res, 2015.
  14. Trehy ML et al., J Liq Chromatogr Relat Technol, 2011.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Comprehensive chemical characterization of e-cigarette liquids using high-resolution Orbitrap GC-MS
Qualitative and Quantitative Analysis of Electronic Cigarette Liquids using Gas Chromatography Orbitrap Mass Spectrometry
Analysis of Nicotine and Impurities in Electronic Cigarette Solutions and Vapor
Comprehensive Characterization of E-Cigarette Liquids
Comprehensive Characterization of E-Cigarette Liquids
2026|Agilent Technologies|Applications