LC/MS, LC/SQ
IndustriesPharma & Biopharma
ManufacturerWaters
Significance of the topic
Reliable impurity profiling of active pharmaceutical ingredients that lack UV chromophores is critical for product quality, safety, and regulatory compliance. Ursodeoxycholic acid (UDCA) and its nine European Pharmacopoeia related substances (impurities A–I) are structurally similar and largely non‑chromophoric, which limits applicability of conventional UV methods. This work demonstrates a practical UHPLC workflow leveraging charged aerosol detection (CAD) combined with volatile mobile phases and orthogonal detectors to enable simultaneous, quantitative analysis of UDCA and its related substances with improved sensitivity, linearity, and regulatory readiness.
Study objectives and overview
• Develop an efficient UHPLC method for simultaneous separation and quantification of UDCA and impurities A–I.
• Use a two‑stage workflow: (1) establish chromatographic separation and peak identity with volatile mobile phases using PDA, CAD and mass detection; (2) optimize CAD operating parameters to maximize response, S/N and calibration linearity for quantitative use.
• Deliver a compliance‑ready solution compatible with Empower CDS for regulated laboratory environments.
Methodology
• Sample and standards: UDCA (2 mg/mL stock) and individual impurities (1 mg/mL stocks) prepared in methanol and diluted in 75:25 water/acetonitrile; calibration range 0.05–50 µg/mL.
• Two‑stage development:
- Separation screening: evaluate multiple reversed‑phase column chemistries (CSH C18, CSH phenyl‑hexyl, BEH C18, HSS PFP) and both acetonitrile and methanol organics using volatile, CAD‑compatible additives to identify optimal selectivity and retention.
- CAD optimization: systematically vary CAD parameters impacting aerosol generation, signal smoothing, ion removal, and response linearization (evaporator temperature, sampling rate, filter time constant, ion trap voltage, power function value).
Used instrumentation
• LC system: ACQUITY Premier UHPLC with Quaternary Solvent Manager and Column Manager.
• Column: ACQUITY Premier BEH C18, 2.1 × 100 mm, 1.7 µm; column temp 25 °C; sample temp 10 °C.
• Detectors: PDA detector (for UV check), Waters Charged Aerosol Detector (CAD) as primary quantitative detector, ACQUITY QDa II mass detector for peak identification.
• Typical LC conditions: flow 0.400 mL/min; injection 5.0 µL; mobile phases A/B = 0.05% formic acid in water / 0.05% formic acid in acetonitrile; gradient optimized to separate analytes within 10 minutes.
• CAD final settings: evaporator temperature 50 °C; sampling rate 10 pts/sec; filter time constant 1.4 s; ion trap voltage 80 V; power function value (PFV) 1.10.
• Data system: Empower CDS 3.6.0 for acquisition and reporting.
Main results and discussion
• Column/solvent screening selected the ACQUITY Premier BEH C18 with acetonitrile as providing the best selectivity and peak shape for resolving UDCA and all nine related substances under a 10‑minute gradient.
• PDA confirmed the analytes are effectively non‑chromophoric (no useful UV response), validating the need for non‑UV detection.
• QDa mass data supported peak identification but was not sufficient alone due to structural isomers; individual standards were used to confirm elution order. Mass spectra commonly showed dehydration adducts (loss of one or two water molecules) under ESI(+).
• CAD optimization outcomes:
- Evaporator temperature: 50 °C provided best balance between solvent evaporation (lower baseline noise) and analyte retention of semi‑volatile species.
- Sampling rate: 10 pts/sec chosen as default since S/N was comparable across rates while maintaining adequate peak definition.
- Filter time constant: 1.4 s produced the highest S/N and improved response stability across impurities.
- Ion trap voltage: 80 V reduced baseline noise without excessive loss of analyte signal; improved calibration linearity versus lower voltages.
- PFV: 1.10 improved linearity across the 0.05–50 µg/mL range versus the default PFV of 1.00; a simulation tool predicted PFV=1.10 and experimental verification matched the prediction with low error.
• Performance: optimized method achieved USP resolution ≥2.0 for adjacent peaks, full separation of UDCA plus impurities A–I within 10 minutes, and acceptable S/N and calibration linearity across the validated range.
Benefits and practical applications
• Broad detection of non‑chromophoric, non‑volatile and semi‑volatile impurities without derivatization.
• Compatibility with volatile mobile phases allows straightforward coupling to mass detection for peak confirmation and minimizes mobile‑phase induced CAD baseline artifacts.
• Systematic CAD parameter optimization (evaporator temp, filter time constant, ion trap voltage, PFV) meaningfully improves sensitivity, noise performance and quantitative linearity, enabling robust impurity profiling.
• Integration with Empower CDS supports audit trails, secure data handling, and compliance for regulated QC environments.
Key practical recommendations for method transfer
- Use high‑purity LC‑MS grade solvents and volatile additives (e.g., low % formic acid) to minimize CAD background and drift.
- Confirm peak identity with orthogonal detectors or individual standards when structural isomers are present; mass data alone may be ambiguous.
- Optimize CAD PFV experimentally and/or with simulation tools to maximize linear operating range for target analytes.
- Document CAD settings and data processing parameters in the CDS to ensure reproducible, compliance‑ready results.
Future trends and opportunities
• Wider adoption of CAD coupled with mass detection for regulated impurity testing of non‑chromophoric APIs where UV lacks sensitivity.
• Increased use of predictive simulation tools (e.g., PFV modeling) to accelerate CAD calibration optimization and reduce empirical iteration.
• Further automation of CAD‑centric workflows and method transfer across laboratories, supported by CDS templates and vendor‑provided method packages.
• Extension of the described workflow to other semi‑volatile impurities, degradation products and biologically relevant lipophilic compounds where derivative chemistry is undesirable.
• Potential method harmonization and formal validation pathways for CAD methods to support pharmacopeial inclusion and regulatory acceptance.
Conclusion
The application note presents a pragmatic, two‑stage UHPLC workflow that combines volatile mobile phases, orthogonal detection (PDA/CAD/MS), and targeted CAD parameter optimization to enable fast, sensitive and linear quantification of UDCA and nine related non‑chromophoric impurities. The final method—run on an ACQUITY Premier BEH C18 column with CAD settings tuned (50 °C evaporator, 1.4 s filter, 80 V ion trap, PFV 1.10)—achieves complete separation within 10 minutes and demonstrates improved S/N and calibration linearity suitable for regulated laboratories when managed through Empower CDS.
References
- Peepliwal A, Bonde CG, Bothara KG. A validated RP‑HPLC Method for Quantitative Determination of Related Impurities of UDCA (API) by Refractive Index Detection. Journal of Pharmaceutical and Biomedical Analysis. 2011;54:845–849.
- Pan S, Zhang Y, Meng F, Wang FS. Can UDCA Prevent SARS‑CoV‑2 Infection or Reduce the COVID‑19 Severity? Current Knowledge and Unresolved Issues. Infectious Diseases and Immunity. 2023;3(3).
- Roda A, Gatti R, Cavrini V, Cerre C, Simoni P. HPLC Study of the Impurities Present in Different UDCA Preparations: Comparative Evaluation of Four Detectors. Journal of Pharmaceutical and Biomedical Analysis. 1993;11:751–760.
- Birdsall R, et al. Predicting Linear Operating Range for Charged Aerosol Detection Using an Inverse Power Function Framework. Waters Application Note. 2026;720009440.