ICP-OES, Elemental Analysis
IndustriesSemiconductor Analysis
ManufacturerThermo Fisher Scientific
Significance of the topic
Lithium-ion battery performance, lifetime and safety critically depend on the chemical composition and purity of cathode active materials (CAM). NMC (nickel–manganese–cobalt) materials are widely used in modern batteries for EVs and energy storage; accurate simultaneous quantification of major components (Li, Ni, Mn, Co) and low-level elemental impurities is essential for materials development, quality control and process optimization. Analytical workflows that measure both majors and traces in a single run increase throughput and reduce sample handling errors while maintaining detection capability for critical impurities.
Goals and overview of the study
The study demonstrates a single-measurement ICP-OES method for simultaneous quantification of four major NMC components (Li, Ni, Mn, Co) and 21 trace impurities in digested NMC cathode materials (NMC-111 and NMC-811). The method aims to: achieve extended linear range for majors without liquid dilution; retain sensitivity for trace analytes; deliver high accuracy, precision and robustness across extended analytical sequences; and reduce maintenance needs through improved sample introduction and aerosol handling.
Methodology
Sample preparation
- Approximately 0.1 g aliquots of CAM were acid-digested on a hotplate using 4 mL aqua regia (HNO3 + HCl) and brought to a final volume of 25 mL with ultrapure water. Digests were analyzed without further dilution.
Calibration and QC
- Calibration for major components used standards up to 2,000 mg·L-1 (to cover high dissolved majors). Trace impurity calibration was prepared from intermediate stocks in the low µg·L-1 range (0.02–5 mg·L-1 typical). Sulfur was calibrated via a separate series; a QC sulfur solution (5 mg·L-1) and multi-analyte QC were run regularly.
Instrument mode and measurement strategy
- Intelligent Full Range (iFR) acquisition used: radial viewing to measure major elements with minimized interferences at high concentration, and axial viewing for trace elements to maximize sensitivity.
- An online internal standard (yttrium, 30 mg·L-1) was continuously added to correct for drift and physical interferences.
- Methd optimized to monitor multiple wavelengths per element where appropriate and to evaluate linearity and detection limits across the range of interest.
Used instrumentation
The method was implemented on a Thermo Scientific iCAP PRO XP ICP-OES Duo configured as follows:
- iCAP PRO XP ICP-OES Duo (DUO view)
- Ceramic D-Torch with 2.0 mm quartz center tube
- Glass cyclonic spray chamber and glass concentric nebulizer
- Sheath gas adaptor providing additional argon flow (front outlet, ~0.2 L·min-1 additional gas) to reduce aerosol dispersion
- Thermo Scientific iSC-65 autosampler for automated sample introduction
- Qtegra Intelligent Scientific Data Solution software for acquisition, QC and automated accuracy calculations
Main results and discussion
Linearity, detection limits and calibration quality
- Correlation coefficients (R2) for all monitored lines exceeded 0.999 in most cases, demonstrating excellent linearity across the calibration ranges used for majors and traces.
- Instrumental detection limits were low for trace analytes (typical IDLs: sub-µg·L-1 for many metals; examples include Mg <0.001 mg·L-1, K 0.001 mg·L-1, many transition metals <0.001–0.01 mg·L-1).
Accuracy and precision
- Certified reference material (BAM‑S014) was analyzed (n=10 aliquots) to assess accuracy. Average accuracy for major components and impurities was within approximately 85–110% with RSDs <5% for most analytes. Major components in NMC-111 matched certified values with recoveries ≈97–100%.
- NMC-811 results showed recovered majors close to expected values (Ni ~98% recovery, Co ~95%). Mn recovery in NMC-811 was reported as 124% relative to the nominal certificate value, attributed to high variability in the certified manganese value and sample transfer uncertainties due to electrostatic effects.
Robustness and long-run performance
- Use of a sheath gas improved aerosol delivery, extended the linear dynamic range for majors up to 2,000 mg·L-1, and reduced matrix load on plasma and inlet components, minimizing maintenance frequency.
- Internal standard (Y) responses measured over 6 hours and 140 samples remained within ~85–105% of the initial response, indicating stable performance. QC solutions analyzed every 20 samples read within 95–105% of expected values, confirming batch stability.
Trace impurities in CAMs
- Most impurities were found well below typical specification limits for precursor materials and finished CAM (examples: Fe, Cu, Zn often <0.25–1 mg·L-1 or below specified thresholds; Na and S were quantified at levels consistent with certificate limits where provided).
Benefits and practical applications
The described workflow enables laboratories to: reduce sample handling and the need for multiple dilutions or separate analyses for majors and traces; achieve higher throughput by analyzing majors and impurities in a single run; maintain low detection limits for impurities while covering large concentration ranges for majors; and reduce instrument downtime thanks to improved aerosol control and robust torch configuration. Practical users include battery materials producers, incoming raw material QC, battery R&D groups, and contract testing labs.
Future trends and potential applications
Adoption of combined high-dynamic-range ICP-OES workflows will support faster iterations in materials development and tighter QC for battery manufacturing. Expected future directions include:
- Automation of digestion and online dilution to further streamline workflows for varied CAM compositions.
- Integration with chemometrics and process control systems for real-time materials feedback.
- Broader adoption of sheath-gas and advanced torch geometries to push dynamic range further and minimize maintenance in high-throughput environments.
- Cross-validation with ICP-MS for elements requiring lower detection limits or isotope-specific information.
Conclusion
The optimized iCAP PRO XP ICP-OES method demonstrates that a single analytical workflow can reliably quantify both high-concentration NMC major components and trace-level impurities with good accuracy, precision and robustness. Key enablers are the DUO viewing configuration (radial for majors, axial for traces), addition of sheath gas to extend linear range and reduce matrix effects, and continuous internal standardization. The approach increases laboratory efficiency and is suitable for routine QC and research on cathode active materials.