Consumables, LC columns
IndustriesManufacturerShimadzu
Significance of the topic
Reliable sample trapping and pre-concentration are essential steps in modern HPLC workflows when analyte levels are low or matrix complexity is high. Trap columns designed for reversed-phase extraction protect analytical columns, concentrate target compounds, and improve sensitivity and reproducibility in bioanalytics, environmental testing, and quality control. The Shim-pack MAYI 2 C18-60 is presented as a dedicated high-performance trap column that combines robust silica chemistry with a protective hydrophilic coating to address these practical needs.
Objectives and overview
The document is an instruction manual for the Shim-pack MAYI 2 C18-60 trap column. Its aims are to describe the stationary phase composition, provide specifications and safe operating ranges, and deliver handling, installation, storage, and maintenance recommendations to ensure consistent performance and longevity of the column when used in HPLC systems.
Methodology and used instrumentation
The product is a packed reversed-phase trap column intended to be used with standard HPLC equipment. Key methodological points and recommended system configuration include:
- Column function: use as a trap/extraction column to retain and concentrate analytes prior to analytical separation.
- Mobile phases: typical reversed-phase solvents—water mixed with acetonitrile or methanol. Buffered aqueous phases may be used but pH must remain within the column stability range.
- System plumbing: use short pieces of 1.6 mm outer diameter PEEK or stainless-steel tubing with 0.1–0.3 mm inner diameter between injector and column to minimize dead volume and peak broadening.
- Filtration: mobile phase and samples must be filtered through a membrane filter ≤0.2 µm (select filter media compatible with organic solvents) to prevent particulate clogging.
- Initial conditioning: flush the column from the outlet with initial mobile phase at 0.5 mL/min for ~10 minutes before putting into standard flow-path connection.
Použitá instrumentace
Details on instrumentation and materials referenced in the manual:
- Column: Shim-pack MAYI 2 C18-60 packed column (trap-type).
- Available physical formats: 1.0 mm I.D. × 50 mm (catalog 228-59567-01) and 2.0 mm I.D. × 30 mm (catalog 228-59567-02).
- HPLC system: standard binary or quaternary pump, injector, and detector compatible with the stated pressure and solvent recommendations.
- Tubing: PEEK or SUS (stainless steel) tubing, 1.6 mm O.D., 0.1–0.3 mm I.D.
- Filters: 0.2 µm membrane filters compatible with planned organic solvents.
Specifications and operating conditions
Key stationary phase and column specifications summarized from the manual:
- Packing material: totally porous, high-purity spherical silica particles with outer hydrophilic polymer coating (methylcellulose) and inner pore-bound octadecyl (C18, monofunctional) groups that are endcapped.
- Particle size: 50 µm; pore size: ~6 nm.
- Column dimensions: 1.0 mm I.D. × 50 mm and 2.0 mm I.D. × 30 mm (two catalog variants).
- Maximum operating pressure: up to 60 MPa (1.0 mm I.D. format) and up to 50 MPa (2.0 mm I.D. format); set flow according to system backpressure.
- Usable pH range: 2.0–7.5.
- Maximum operating temperature: 50 °C.
Main recommendations, results and discussion
The manual emphasizes practices that preserve column performance and prevent common failure modes rather than reporting experimental performance data. Important operational guidance and rationale:
- Avoid introducing contaminants or air into the column by purging the flow path with mobile phase before connection; particulates and bubbles degrade performance and raise backpressure.
- Prevent sudden pressure, pH, or temperature excursions, and avoid repeated use near the stated limits to reduce risk of irreversible damage and shortened lifetime.
- Minimize dead volume and tubing length to reduce peak broadening—critical for trap columns intended to preserve chromatographic resolution downstream.
- When handling potentially precipitating samples (e.g., protein-rich matrices), adapt mobile phase organic content to avoid in-line precipitation and clogging.
- For columns used with buffers or ion-pair reagents, perform thorough desalting washes (≥5 column volumes of salt-free eluent) followed by 5 column volumes of strong reversed-phase solvent prior to storage in acetonitrile or methanol.
Benefits and practical applications
The Shim-pack MAYI 2 C18-60 design combines protected silica chemistry and trap-column geometry to deliver several practical benefits:
- Effective on-line extraction and pre-concentration of analytes from dilute or complex matrices prior to analytical separation.
- Hydrophilic outer coating reduces undesired surface interactions and may improve robustness toward matrix components.
- Endcapped monofunctional C18 chemistry provides conventional reversed-phase retention with reduced silanol activity, aiding reproducibility for basic or polar analytes.
- Compact formats compatible with modern HPLC/UHPLC systems and easy integration as a trap ahead of analytical columns.
Storage, disposal and technical support
Storage and end-of-life handling recommendations:
- Cap both column ends when removing from a system to prevent solvent evaporation and contamination. Store at stable temperature away from high temperature and humidity.
- For long-term storage, flush then replace the mobile phase with methanol or acetonitrile before capping.
- Dispose of columns per local regulations; do not mix with general household waste.
- Technical support is provided by the manufacturer (Shimadzu); the manual notes that lifetime is not guaranteed if operating recommendations are not followed or with normal aging.
Future trends and possibilities for use
While this product is a conventional packed-silica trap column, broader trends and opportunities relevant to trap-column technology include:
- Developments in hybrid and polymeric supports that extend usable pH ranges and thermal stability.
- Smaller particle sizes and superficially porous (core-shell) geometries to improve mass transfer and reduce backpressure for higher throughput.
- Integration with automated sample-preparation and online SPE workflows for higher laboratory productivity and reduced sample handling error.
- Green-solvent strategies and solvent-reduction techniques to decrease analytical environmental footprint and operating costs.
Conclusion
The Shim-pack MAYI 2 C18-60 is specified as a robust reversed-phase trap column for extraction and concentration in HPLC workflows. Its spherical porous silica packing with an outer hydrophilic coating and endcapped C18 chemistry is tailored for routine trapping use within defined pH, pressure, and temperature limits. Adherence to the recommended plumbing, conditioning, filtration, washing, and storage practices is essential to achieve consistent performance and acceptable column lifetime.
Reference
Shim-pack MAYI 2 C18-60 Instruction Manual, Shimadzu Corporation, Document 228-97417B, September 2026.