Significance of the topic
The contamination of aquatic environments by emerging contaminants (EC)—including pharmaceuticals, personal care products (PPCP), endocrine disruptors and chemicals adsorbed to microplastics (MP)—poses ecological and public‑health risks that are only recently being understood. Sensitive, reproducible workflows that combine efficient extraction from complex matrices (sand, sediments, beached plastics) with selective LC‑MS/MS detection are required for monitoring, risk assessment and remediation planning.
Objectives and overview of the study
This work develops and evaluates a comprehensive multicomponent workflow (overflow separation → microwave‑assisted extraction (MAE) → solid‑phase extraction (SPE) → HPLC‑MS/MS) for determination of 24 EC in beach sand and model microplastic fragments. Goals included: (i) isolating potential microplastics from sand by density overflow separation, (ii) extracting a chemically diverse analyte suite with MAE, (iii) cleanup and concentration by SPE, and (iv) selective quantification by UHPLC coupled to triple‑quadrupole MS using scheduled MRM.
Methodology
Sampling and physical preparation:
- Surface sand collected at multiple northern European beaches; one beached plastic rope sampled and cut into <5 mm fragments to model MP.
- Samples frozen at -20 °C, dried to constant weight, sieved to 0.1–1.6 mm, and separated by overflow in saturated ZnCl2 (ρ = 1.7 g·cm⁻³) to obtain a float fraction potentially containing MP.
Extraction and cleanup:
- Microwave‑assisted extraction (MARS 6TM) used with methanol:acetonitrile (4:1 v/v). Typical MAE cycle: 10 min heating, 10 min extraction, 10 min cooling at 60 °C and 400 W; three sequential extractions per sample (sand: 10 mL each; MP: 5 mL each).
- Combined extracts evaporated in a vacuum centrifuge and reconstituted in water:methanol (95:5, pH 2) for SPE.
- SPE employed a polymeric reversed‑phase cartridge (PEP, 60 mg, 3 mL); conditioning with MeOH and acidified water, loading in 95:5 water:MeOH pH 2, washing, drying and elution with MeOH:NH4OH (95:5, pH 11.5).
Chromatography and mass spectrometry:
- UHPLC: AZURA® UHPLC system with Eurospher Pro C18 column (3 µm, 2.1 × 100 mm), column temperature 40 °C, 1 µL injection.
- MS: SCIEX Triple Quad 5500+ in scheduled MRM; two methods used—positive ESI for 17 analytes and negative ESI for 7 analytes; internal standard D4‑acetaminophen spiked prior to MAE to monitor losses.
- Method calibration: eight concentration levels; linearity R² > 0.997 for all analytes; LOD by S/N = 3 and LOQ by S/N = 10; 15/24 analytes had LOD < 1 ng·mL⁻¹.
Used instrumentation
- Pump: AZURA P 8.1 L (stainless steel)
- Autosampler: AZURA AS 6.1 L with cooling
- Column thermostat: AZURA CT 2.1
- UHPLC column: Eurospher® Pro C18, 3 µm, 2.1 × 100 mm
- Mass spectrometer: SCIEX Triple Quad 5500+ with Turbo V source
- Microwave extractor: CEM MARS 6TM (provided)
- Vacuum centrifuge for evaporation and SPE manifold for cleanup
Main results and discussion
Method performance and recoveries:
- MAE extraction from blank sand achieved >85 % recovery for 19 of 24 analytes; from polyethylene MP model, 17 analytes exceeded 85 % recovery; some analytes (e.g., BPAF) showed lower recovery (~74–77 %).
- SPE recoveries were variable: 12 analytes had >85 % recovery under the single SPE protocol used; some compounds require tailored cleanup to improve yields.
- Precision and accuracy: intra‑ and inter‑assay deviations were generally <15 % (≤16 % for all).
- Absolute recovery from spiked, matrix‑loaded sand samples varied widely between sampling sites and analytes—median absolute recoveries were often below 50 %, reflecting strong matrix effects and sample heterogeneity.
Environmental findings from 19 sand samples:
- Float fraction (potential MP) was low at sampled beaches—typically <0.05 ‰ in many freshwater/Baltic samples and up to 2 ‰ in the most contaminated sand—limiting direct analysis of environmental MP in these samples.
- Caffeine (COF) was detected in every sand sample at concentrations between ~21 and 175 ng·g⁻¹, often the most abundant analyte; this aligns with literature reporting caffeine as a common tracer of anthropogenic contamination.
- Other frequently detected analytes across regions included bisoprolol (BIS), DEET, desipramine (DEZ) and ketoprofen (KET); several analytes were present at low ng·g⁻¹ levels when measured with sufficient sensitivity.
- Some compounds (e.g., BPA) were not detected in environmental samples despite good recoveries in spiked tests, suggesting low environmental levels or losses during sample processing.
Benefits and practical applications of the workflow
- Integrated, modular workflow allows simultaneous assessment of potential MP load and a broad suite of EC from the same sand sample.
- MAE shortens extraction time and improves reproducibility relative to conventional solid–liquid extractions and sonication, and is readily parallelizable for higher throughput.
- Scheduled MRM on a triple‑quad provides selectivity to resolve partial co‑elution and control for matrix interferences.
- Applicable for routine environmental monitoring programs, screening of recreational beaches, and comparative studies across regions.
Limitations
- Single SPE and extraction conditions represent a compromise across chemically diverse analytes; some analytes require alternative solvents or stationary phases for optimal recovery.
- Strong matrix effects produced variable absolute recoveries between real samples; results are currently more qualitative than fully quantitative.
- Low float‑fraction yields constrained direct analysis of environmental microplastics; targeted sampling of larger plastic debris or upscaling density separation is necessary for robust MP analysis.
Future trends and opportunities for application
- Adopt matrix‑matched calibration and multiple isotopically labelled internal standards to correct for matrix effects and improve quantitation across matrices.
- Optimize fractionated SPE strategies or parallel cleanup protocols tailored to analyte classes (polar vs. hydrophobic) to increase overall recovery and reduce bias.
- Upscale density separation procedures or target larger beached plastic items to obtain sufficient environmental MP mass for direct analysis of sorbed contaminants.
- Integrate this workflow into long‑term monitoring networks and combine chemical data with ecological indicators to assess impact and inform policy (wastewater treatment upgrades, banning persistent chemicals).
- Explore high‑resolution mass spectrometry or suspect/non‑target screening extensions to detect additional emerging pollutants not included in the targeted panel.
Conclusion
The presented MAE‑SPE‑UHPLC‑MS/MS workflow demonstrates a practicable and relatively high‑throughput approach to detect a broad panel of emerging contaminants in sand and model microplastic fragments. While MAE and scheduled MRM deliver good sensitivity and selectivity, matrix variability and the need for a compromise SPE protocol limit quantitative accuracy for all analytes across diverse field samples. The method is valuable for qualitative screening and comparative regional surveys, and further refinements (isotope-labelled standards, matrix‑matched calibration, tailored cleanup) will increase quantitative robustness for monitoring programs.
References
- Salaudeen A. O., Olawore Y. A., Yakubu H., Abdulkadiri H., Aduloju E. I. Osong Public Health Research Perspectives, 2026. DOI as provided in source.
- Ivar do Sul J. A., Costa M. F. Environmental Pollution, 2014, 185, 352–364.
- Jiménez‑Skrzypek G., Ortega‑Zamora C., González‑Sálamo J., Hernández‑Sánchez C., Hernández‑Borges J. Journal of Chromatography Open, 2021, 1, 100001.
- Llorca M., Farré M., Karapanagioti H. K., Barceló D. Marine Pollution Bulletin, 2014, 87, 286–291.
- Aouant A. D., Hela D. Sustainability, 2025, 17, 10025.