No Pairing Required: Oligonucleotide Analysis on Phenyl Stationary Phases

Applications | 2026 | KNAUERInstrumentation
HPLC
Industries
Pharma & Biopharma
Manufacturer
KNAUER

No Pairing Required: Oligonucleotide Analysis on Phenyl Stationary Phases — Expert Summary


Importance of the topic:
Oligonucleotides (ONs) are central to modern therapeutics and require robust analytical methods for identity, purity and impurity profiling. Conventional IP-RPLC uses ion-pairing reagents that improve retention but compromise MS compatibility and system cleanliness. Aromatic (phenyl-based) stationary phases offer an ion-pair-free alternative that exploits hydrophobic, dispersive and π–π interactions with exposed nucleobases, potentially improving MS compatibility while retaining useful selectivity for length, sequence and modification differences.

Objectives and study overview:
  • Compare three phenyl-based phases (pentafluoro-phenyl PFP, conventional phenyl, and phenyl-hexyl) for reversed-phase, ion-pair-free separation of oligonucleotides.
  • Evaluate how oligonucleotide length, chemical modifications and secondary structure influence retention, peak shape and elution order.
  • Demonstrate gradient and method transfers between phases and assess practical performance for routine analysis and MS compatibility.

Methodology and experimental design (summary):
  • Four representative oligonucleotides were tested (a 64mer DNA, a dT24 with C12 linker, a 21-nt antisense RNA with 2′-dTs and a thioate, and a modified 20mer with multiple thiolations/DMT protection).
  • Mobile phases: Eluent A = 50 mM ammonium acetate, pH 8.0; Eluent B = 100 mM ammonium acetate : methanol 50:50 (v/v), pH 8.0.
  • Detection: UV at 260 nm; flow 0.5 mL/min; column temperature 50 °C. Various linear gradients (G1–G5) were explored to optimize retention and peak shape.
  • Columns tested: Eurospher II 100-3 PFP (50 × 2 mm), Eurospher II 100-2 Phenyl (50 × 2 mm) and Eurospher II 100-3 Phenyl-Hexyl (100 × 2 mm).

Instrumentation used:
  • Pump: AZURA P 6.1L Pump LPG (stainless steel).
  • Autosampler/liquid handler: LH 8.1 Liquid Handler with robotic cooler and fast wash station.
  • Detector: AZURA DAD 2.1L Diode Array Detector (UV 190–700 nm) with 10 mm, 2 μL stainless-steel flow cell.
  • Column thermostat: AZURA CT 2.1 (up to 8 columns, 5–85 °C).
  • Columns: Eurospher II 100-3 PFP (05BE071E2G), Eurospher II 100-2 Phenyl (05BE050E2F), Eurospher II 100-3 Phenyl-Hexyl (10BE161E2G).
  • Software: ClarityChrom 10.1 (instrument and PDA licenses).

Main results and discussion:
  • PFP phase: Starting from 20% B produced incomplete retention and poor shapes for some samples. Reducing initial organic to 10% B improved retention (notably for the antisense RNA), while increasing maximum B to 50% enabled complete elution of the dT24 C12 sample and improved peak shapes for several analytes. The electron-deficient PFP ring likely contributes additional dipolar/donor–acceptor interactions beyond π–π stacking.
  • Conventional phenyl: Transferring the optimized PFP gradient produced comparable retention and the same elution order. Raising the maximum organic to 60% B caused earlier elution of dT24 C12 and improved symmetry for the 64mer; other analytes were only minimally affected. This indicates similar dominant π–π and hydrophobic contributions, with subtle selectivity differences versus PFP.
  • Phenyl-hexyl: Direct method transfer yielded the same elution order but generally earlier elution. The phenyl-hexyl ligand (phenyl attached via a C6 linker) presents a different balance of hydrophobic vs aromatic interactions, but under the tested conditions it did not show a clear advantage in peak shape or selectivity compared with PFP or conventional phenyl.
  • Across all phases the elution order remained consistent, indicating that oligonucleotide properties (length, base exposure, chemical modification and secondary structure) strongly govern retention. Folded structures or protected/modified bases reduce nucleobase accessibility and thus attenuate aromatic interactions.
  • No single phenyl-based phase proved universally superior for these samples; method optimization (initial %B, gradient slope, maximum %B, temperature) is necessary for best results with each analyte.

Benefits and practical applications:
  • Ion-pair-free reversed-phase methods on phenyl-based phases avoid alkylammonium reagents, reducing MS contamination risk and ion suppression—improving downstream LC–MS compatibility.
  • Phenyl, PFP and phenyl-hexyl phases provide complementary selectivity to IP-RPLC and HILIC, enabling separation of length variants, some sequence/chemically modified species and certain degradation products.
  • These phases can be integrated into workflows for purity testing, impurity profiling (shortmers, sequence variants), method transfer and routine QC with reduced system maintenance associated with ion-pair reagents.

Limitations and considerations:
  • Retention is heavily dependent on nucleobase accessibility; structured or highly modified ONs may show low retention on aromatic phases.
  • Stationary-phase properties beyond chemistry—ligand density, spacer length, particle morphology, pore size, endcapping and hardware—affect performance and must be considered during method transfer.
  • Method optimization remains necessary; phenyl-hexyl did not universally improve peak shape despite offering a distinct interaction balance.

Future trends and opportunities:
  • Further tuning of aromatic stationary phases (controlled ligand density, mixed-ligand sorbents) to increase selective retention for specific ON classes.
  • Integration with sensitive LC–MS/MS workflows for high-resolution impurity mapping, using ion-pair-free mobile phases to enhance MS cleanliness and sensitivity.
  • Use of elevated temperature or denaturing additives (compatible with MS) to manipulate secondary structure and improve nucleobase exposure when needed.
  • Adoption of UHPLC particle formats and optimized gradient profiles to increase resolution and throughput for therapeutic oligonucleotide analysis.
  • Development of decision trees combining IP-RPLC, HILIC and aromatic RP methods to select the optimal approach based on sequence length, modification and desired MS compatibility.

Conclusion:
Phenyl-based stationary phases (PFP, conventional phenyl and phenyl-hexyl) enable ion-pair-free reversed-phase separation of diverse oligonucleotides by exploiting hydrophobic, dispersive and aromatic interactions. They offer a viable, more MS-compatible alternative to IP-RPLC for many applications, though no single phase dominated for all tested analytes. Practical method development—adjusting initial %B, gradient steepness, and maximum organic content—is required to accommodate differences in length, modifications and secondary structure. These phases complement existing IP-RPLC and HILIC approaches and expand options for oligonucleotide QC and LC–MS workflows.

Reference:
  1. Tsalmpouris A., Largy E., Guillarme D. Systematic evaluation of phenyl stationary phases for oligonucleotide analysis without ion-pairing reagents. Journal of Chromatography A. 2025;1762:466395. DOI: 10.1016/j.chroma.2025.466395.
  2. Studzińska S., Bocian S., Rivoira L., Faden E., Faden G. Separation and identification of oligonucleotides impurities and degradation products by reversed phase UHPLC using phenyl-bonded stationary phases without ion pairs. Journal of Chromatography A. 2024;1736:465380. DOI: 10.1016/j.chroma.2024.465380.
  3. Studzińska S., Pietrzak L., Buszewski B. The Effects of Stationary Phases on Retention and Selectivity of Oligonucleotides in IP-RP-HPLC. Chromatographia. 2014;77:1589–1596. DOI: 10.1007/s10337-014-2766-x.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.