Thermal desorption, GC/MSD, GC/MS/MS, GC/QQQ
IndustriesEnvironmental
ManufacturerAgilent Technologies, Markes
Significance of the topic
Indoor air is an increasingly recognized vector for human exposure to volatile and semivolatile neutral PFAS (per‑ and polyfluoroalkyl substances) such as fluorotelomer alcohols (FTOHs), fluorinated acrylates, perfluoroalkyl sulfonamides (FOSAs) and sulfonamide ethanols (FOSEs). Because these species have higher vapor pressures and long atmospheric lifetimes compared with ionic PFAS, reliable air sampling and trace-level analysis are essential for exposure assessment, source identification, and regulatory monitoring. Analytical workflows must address compound volatility, extensive laboratory background contamination, and the need for sensitive, selective detection at pg–ng per tube levels.
Objectives and study overview
This application study evaluated and extended the ASTM D8591-24 TD‑GC/MS/MS workflow by:
- Assessing the performance of Agilent UltiMetal PFAS thermal desorption (TD) tubes for volatile and semivolatile PFAS analysis.
- Expanding the D8591 target list beyond the four FTOHs to include 8:2 and 10:2 fluorinated acrylates, MeFOSA, EtFOSA, MeFOSE, and EtFOSE.
- Demonstrating method sensitivity, linearity, repeatability, and low-level detection capability using a Markes TD100‑xr desorber coupled to an Agilent 8890 GC and Agilent 7000E triple quadrupole MS.
Methodology and used instrumentation
The study followed TD‑GC/MS/MS procedures consistent with ASTM D8591-24 and included careful contamination mitigation. Key methodological points:
- Calibration: multi‑point (typically five points) curves prepared by solution loading onto TD tubes; calibration range used for method validation 0.1–2.5 ng on tube (100–2,500 pg per tube in other units); extended experiments down to 10 pg on tube to probe trace detection.
- Sample/application loading: manual gas‑tight syringe deposition (1 µL test solution + 1 µL ISTD), short purge (3 min on CSLR) prior to analysis (note: Markes recommends 60 min purge for some analyte classes).
- Contamination control: methanol washes, PFAS‑optimized polypropylene vials/inserts/caps, process and instrument blanks, and UltiMetal passivated surfaces to minimize background.
Used instrumentation (separate section):
- Thermal desorber: Markes TD100‑xr with UNITY‑xr PFAS focusing trap (MKI‑U‑T24PFAS‑2S).
- Thermal desorption tubes: Agilent UltiMetal PFAS TD tubes (p/n 5190‑1137) preconditioned on a Markes TC‑20 multitube conditioner.
- GC: Agilent 8890 gas chromatograph; column Agilent J&W DB‑WAX Ultra Inert (30 m × 0.25 mm, 0.25 µm, p/n 122‑7032UI).
- MS: Agilent 7000E triple quadrupole GC/TQ operated in dynamic MRM mode; Agilent InertPlus Extractor ion source; transfer line 260 °C, source 280 °C, quadrupoles 150 °C; collision gas nitrogen.
- Typical TD and GC conditions: tube desorption 300 °C for 10 min; trap desorption 300 °C for 4 min; carrier gas helium, split outlet 6 mL/min (6:1); oven program starting 35 °C with ramps to 250 °C.
Main results and discussion
Chromatography and selectivity:
- The DB‑WAX Ultra Inert column provided superior peak shape and reduced tailing versus a DB‑FFAP equivalent; baseline separation was achieved for most targets. Some coelutions were observed (MFOET with 8:2 FTOH, MFDET with 10:2 FTOH, and partial FOSE coelution), but compound‑specific MRM transitions enabled unambiguous identification and quantitation of coeluting analytes.
- MS source temperature was critical: raising the source from 250 °C to 280 °C markedly improved peak tailing for FOSA/FOSE analytes.
Precision, linearity and sensitivity:
- Inter‑ and intra‑day repeatability for the four FTOHs at 5 ng on tube: all %RSDs < 6% (n = 7 across four days), demonstrating strong short‑term and bench‑storage repeatability.
- Five‑point calibration curves over 0.1–2.5 ng on tube met ASTM D8591‑24 performance criteria. Average response factor %RSDs (AVG RF %RSD) ranged 3.3%–8.8% across analytes.
- Precision at 250 pg on tube ranged from 3% to 11% depending on analyte.
- Instrument detection limits (IDLs) calculated from replicate measurements spanned ~13–39 pg per tube for the target analytes; observed quantified signals were detected at 10, 25 and 50 pg on tube for many compounds, though the authors recommend that individual labs verify trace‑level performance for their systems.
Flow path inertness and contamination control:
- UltiMetal Plus CVD surface deactivation applied to stainless steel TD tubes reduced active sites and adsorption/catalysis, improving peak shape, recovery and lowering background. The coating is thermally stable (reported up to 400 °C) and hydrophobic, decreasing carryover.
- Rigorous blanking, methanol rinsing, dedicated PFAS‑compatible consumables, and testing of syringes/vials verified negligible instrument or consumable contributions to target analytes in this work.
Benefits and practical applications
The demonstrated TD‑GC/MS/MS workflow with UltiMetal PFAS TD tubes offers several operational advantages:
- High sensitivity and selectivity for volatile and semivolatile neutral PFAS at sub‑ng and low‑pg levels, enabling both research and routine monitoring of indoor air and controlled chamber experiments.
- Extensible target list beyond the ASTM D8591 FTOHs to include acrylates, sulfonamides and FOSEs while still meeting standard performance requirements.
- Reduced analytical uncertainty from surface interactions and laboratory background due to UltiMetal passivation and strict contamination controls.
Practical implications: laboratories can adopt this TD‑GC/MS/MS approach for exposure assessments, indoor air surveillance, product emission testing, and controlled chamber studies; however, method transfer requires local verification of LOD/LOQ and contamination control tailored to each laboratory's environment.
Future trends and potential applications
Areas for further development and likely near‑term trends:
- Validation and standardization for a broader suite of volatile and semivolatile PFAS classes, including more highly polar perfluorinated acids (which may require different sorbents and extended purge times).
- Routine inter‑laboratory comparisons and proficiency testing to harmonize TD protocols, blanking procedures, and calibration strategies for sub‑ng/pg quantitation.
- Automation of sample loading and trap/regeneration workflows to increase throughput while preserving low background levels.
- Integration with ambient and indoor monitoring networks to map exposure gradients and product emission sources; coupling TD‑GC/MS/MS data with inhalation exposure models.
- Continued development of inert coatings and low‑background consumables to push reliable detection to lower pg levels with verified robustness.
Conclusion
The combination of Agilent UltiMetal PFAS thermal desorption tubes with a Markes TD100‑xr desorber, Agilent 8890 GC and Agilent 7000E GC/TQ MS operated in dynamic MRM produces a robust, sensitive, and extensible TD‑GC/MS/MS method for volatile and semivolatile PFAS in air. The method met ASTM D8591‑24 performance criteria for an expanded list of analytes, delivered IDLs in the low‑10s of picograms per tube, and showed strong repeatability (%RSD < 6% for FTOHs at 5 ng on tube). Laboratories seeking to implement trace‑level PFAS air monitoring should apply rigorous contamination control, verify trace LOD/LOQ performance locally, and consider analyte‑specific adjustments (e.g., purge times, sorbents) when extending the method to additional PFAS classes.
References
- Vieira, H. G.; Canela, M. C.; Urban, R. C.; Cabrero, B. S. Current Knowledge about Per‑ and Polyfluoroalkyl Substances (PFAS) in the Atmosphere: Fate, Analytical Methods and Research Priorities. Chemosphere 2025, 389, 144703.
- Kim, M.; Li, L. Y.; Grace, J. R.; Yue, C. Selecting Reliable Physicochemical Properties of Perfluoroalkyl and Polyfluoroalkyl Substances (PFASs) Based on Molecular Descriptors. Environ. Pollut. 2015, 196, 462–472.
- ASTM International. ASTM D8560‑24: Standard Guide for Determination of Airborne PFAS in the Indoor Air Environment; ASTM International, 2024.
- Morales‑McDevitt, M. E.; Becanova, J.; Blum, A.; et al. The Air That We Breathe: Neutral and Volatile PFAS in Indoor Air. Environ. Sci. Technol. Lett. 2021, 8, 844–850.
- ASTM International. ASTM D8591‑24: Standard Test Method for Determination of Fluorotelomer Alcohols in Test Chamber Air by Thermal Desorption–GC–Triple Quadrupole MS; ASTM International, 2025.
- Sunderland, E. M.; Hu, X. C.; Dassuncao, C.; et al. A Review of the Pathways of Human Exposure to Poly‑ and Perfluoroalkyl Substances (PFASs) and Present Understanding of Health Effects. J. Expo. Sci. Environ. Epidemiol. 2019, 29(2), 131–147.
- Shoeib, M.; Harner, T.; Wilford, B. H.; et al. Perfluorinated Sulfonamides in Indoor and Outdoor Air and Indoor Dust: Occurrence, Partitioning, and Human Exposure. Environ. Sci. Technol. 2005, 39(17), 6599–6606.
- Zhao, Z.; Tang, J.; Mi, L.; et al. Perfluoroalkyl and Polyfluoroalkyl Substances in the Lower Atmosphere and Surface Waters of the Chinese Bohai Sea, Yellow Sea, and Yangtze River Estuary. Sci. Total Environ. 2017, 599–600, 114–123.
- Markes International. PFAS in Air: Validating ASTM D8591 for the Measurement of Airborne Fluorotelomer Alcohols; Application Note 179, Markes International, 2026.
- Markes International. Measuring PFAS Pollution in Ambient Air Using TD–GC–MS/MS; Application Note 166, Markes International, 2022.
- Agilent Technologies. UltiMetal Plus – Advanced Chemistry for Stainless Steel Surface Deactivation. Technical overview, publication 5991‑3357EN, 2014.