Importance of the topic
Electrospray ionization (ESI) coupling between liquid chromatography (LC) and mass spectrometry (MS) remains a critical and often rate-limiting interface in high-sensitivity proteomics and other LC–MS workflows. Variability in emitter positioning, spray stability, emitter lifetime, and manual fluidic handling are persistent sources of analytical imprecision, reproducibility loss, and throughput constraints. Automating and standardizing the LC–MS ion source while preserving compatibility with advanced column technologies and front-end options (e.g., FAIMS) addresses a major practical need for robust large-cohort and high-sensitivity analyses.Goals and study overview
This work presents the design and evaluation of the OptiSpray automated low-flow ion source and a replaceable-emitter cartridge platform intended to reduce user-dependence, improve spray stability and emitter lifetime, and integrate leading column technologies for nanoflow and capillary-flow LC–MS. The system goals include: automated emitter positioning, real-time spray monitoring, low-variance replaceable emitters, and seamless integration with different column types for reproducible, high-throughput proteomics.Methodology and analytical approach
- Automation: A 3-axis motorized stage positions emitters automatically using two optimization algorithms — one tailored for nano-flow (<1 µL/min) and another for capillary flow (>1 µL/min) with nebulizing gas. Positioning routines include adaptations for different MS inlets and FAIMS Pro interfaces.
- Spray monitoring: Visual (multiview cameras with white and green LEDs) and electrical (emitter current) sensors provide objective metrics to detect good spray, sputtering, droplet ejection, or spray loss.
- Replaceable emitters and cartridges: Plug-and-play cartridges support replaceable integrated emitters in two formats (nano and capillary), integrated sheath gas, on-board memory, and temperature-controlled columns to minimize post-column volume and extend consumable lifetime.
- Column integration: Compatibility demonstrated with Thermo Scientific PepMap Neo (high pressure tolerance) and µPAC Neo (high reproducibility and longevity) column families. The µPAC cartridge design electrically “floats” the column to improve chromatographic performance and proteome coverage compared to a standalone grounded µPAC implementation.
- Performance evaluation: Multiple instruments and workflows were used, including Orbitrap Lumos, Orbitrap Ascend, Orbitrap Exploris 480, and Orbitrap Astral MS platforms paired with Vanquish Neo UHPLC. Sample types included HeLa digests, peptide standards (PRTC), and FFPE tissue digests from historical pandemics. Data were processed using tools such as Spectronaut, Proteome Discoverer with CHIMERYS, Skyline, and DIA-NN.
Used instrumentation
- OptiSpray automated low-flow ion source with 3-axis motorized stage, multiview cameras, LEDs, integrated sheath gas, and replaceable-emitter cartridge architecture.
- Thermo Scientific Vanquish Neo UHPLC for nanoflow and capillary flow delivery.
- Mass spectrometers: Orbitrap Lumos, Orbitrap Ascend, Orbitrap Exploris 480, Orbitrap Astral (with FAIMS Pro Duo where indicated).
- Columns: Thermo Scientific PepMap Neo (75 µm × 15 cm and 75 µm × 50 cm) and µPAC Neo (50 cm, µPAC Neo HT), temperature-controlled cartridge formats.
- Software/processing: Spectronaut 19, Proteome Discoverer with CHIMERYS, Skyline, DIA-NN 2.1.
Main results and discussion
- Automated emitter positioning reproducibly matched analytical proxy ions and improved quantitative precision. Positioning routines were successfully adapted for FAIMS, yielding increased peptide and protein coverage compared with unoptimized/manual positioning.
- FAIMS integration: Optimization for FAIMS demonstrated up to ~10% increase in protein groups and ~20% increase in peptide identifications in comparisons shown, with improved median peak-area %CVs across replicates and reduced variability in FAIMS-enabled runs.
- Inter-cartridge and inter-emitter reproducibility: Low variability was observed across cartridges and emitters. Median inter-cartridge and inter-emitter CVs for peak area and identifications were reported below ~7% (examples show median CVs around 4–5%), indicating tight manufacturing and installation reproducibility.
- Long-term robustness: Across 2000 injections distributed over multiple cartridges, OptiSpray exhibited consistent protein identifications and quantitative stability. Retention-time reproducibility reported median RT CV near 0.1% (max 2.8%), and peak-area reproducibility had a maximum CV of ~10.9% across the full injection series, supporting reliability for large-cohort studies.
- High-sensitivity and deep-proteome capability: Using an Orbitrap Astral with Vanquish Neo UHPLC and a µPAC Neo 50 cm cartridge, analysis of 195 single-injection FFPE tissue digest samples (200 ng) identified 9,034 proteins, demonstrating the platform’s capacity for extensive proteome coverage in a high-value cohort study.
- Practical automation benefits: The platform effectively narrows the skill gap in emitter positioning—performance from automated routines approximated results achieved by experienced mass spectrometrists using manual positioning, across a range of sample loads (1 ng to 200 ng).
Benefits and practical applications
- Reduced operator dependence: Automated positioning and objective spray monitoring lower variability introduced by manual emitter handling and improve reproducibility across users and sites.
- Consumable life and throughput: Replaceable emitters and cartridge design extend usable lifetime, reduce downtime for emitter replacement, and lower post-column volume to preserve chromatography and sensitivity.
- Compatibility with advanced front ends: Integration with FAIMS and multiple column technologies enables optimized sensitivity, specificity, and deep proteome coverage in both discovery and large-cohort workflows.
- Suitability for high-throughput proteomics: Demonstrated stability over thousands of injections and low inter-cartridge variability supports large-scale clinical or retrospective cohort studies and routine lab operations.
Future trends and potential applications
- Deeper automation and diagnostics: Further integration of predictive maintenance, automated QC checks, cloud reporting, and remote diagnostics could enhance uptime and standardization across distributed labs.
- Broader platform compatibility: Expanding cartridge and emitter formats for a wider range of column chemistries, inner diameters, and flow regimes would increase applicability (e.g., single-cell proteomics, clinical microflow assays).
- Miniaturization and single-cell adaptation: Continued improvements in nano-flow stability and sensitivity may enable more routine single-cell proteomics and low-input clinical assays when paired with optimized sample prep.
- Standardization for regulated environments: Consistent cartridge performance and automated routines can be a foundation for translating high-sensitivity LC–MS assays toward regulated workflows, with required validation steps.
- Data-driven optimization: Closed-loop optimization where spray metrics and MS performance feed automated re-calibration could further reduce variability and improve long-run consistency.
Conclusion
The OptiSpray platform addresses four central weaknesses at the LC–MS interface—emitter positioning, spray monitoring, emitter lifetime, and column integration—through a combination of motorized automation, sensor feedback, replaceable emitters, and cartridge-based column integration. The system demonstrably improves reproducibility, reduces operator dependence, and supports deep and high-throughput proteomic workflows, including FAIMS-enabled analyses. These attributes make the platform well suited for large-cohort studies, routine laboratory deployment, and applications demanding high sensitivity and long-term stability.References