Analysis of Adsorbable Organic Fluorine (AOF) in Accordance with EPA 1621 Using Combustion Ion Chromatograph(CIC) —Efficient Workflow Using C-IC Solution—

Applications | 2026 | ShimadzuInstrumentation
Ion chromatography
Industries
Environmental
Manufacturer
Shimadzu

Significance of the topic


The determination of Adsorbable Organic Fluorine (AOF) in aqueous matrices is an important screening approach for assessing the total burden of organically bound fluorine, including per- and polyfluoroalkyl substances (PFAS), their degradation products and other fluorinated organics. Methods that combine selective adsorption and total fluorine measurement—such as combustion ion chromatography (CIC) following EPA 1621—offer a practical way to quantify aggregate organic fluorine rather than relying solely on targeted compound lists. Reliable, automated CIC workflows improve laboratory throughput, reproducibility, and compliance with regulatory screening requirements.

Objectives and study overview


This application note demonstrates an automated CIC workflow that complies with EPA Method 1621 by combining an AQF-5000H automatic quick furnace for sample combustion (Nittoseiko Analytech) with a Nexera IC ion chromatograph (Shimadzu) and integrated C-IC Solution software. Key goals were to: (1) implement the EPA 1621 screening sequence from adsorption of AOF onto granular activated carbon (GAC) through combustion and fluoride measurement by IC; (2) verify method performance metrics required by EPA 1621 (calibration, retention time window, initial precision and recovery (IPR), method detection limit (MDL), minimum level (ML), and QC checks); and (3) demonstrate a streamlined, automated workflow for routine AOF screening.

Methodology


The analytical scheme follows EPA 1621: aqueous samples are passed through a GAC cartridge (TXA-04) to trap adsorbable organic fluorine; inorganic fluoride is washed from the cartridge; the GAC is transferred to a ceramic combustion boat and combusted; combustion products are trapped in an absorption solution; and the captured fluoride is quantified by suppressed ion chromatography. Controls and QC steps (calibration standards, method blanks, IPR, Ongoing Precision and Recovery (OPR), matrix spike (MS) and matrix spike duplicate (MSD)) are performed according to EPA 1621 to validate batch results. The combustion-calibration approach subjects calibration standards to the same combustion and adsorption steps as samples (i.e., “combustion calibration”). Data processing converts IC fluoride results to original-sample AOF concentrations per EPA 1621 formulas.

Used Instrumentation


The following instrumentation and principal operating settings were employed (overview rather than table-for-table reproduction):
  • Combustion unit: AQF-5000H automatic quick furnace (Nittoseiko Analytech) with ceramic sample boats and pyrolysis tubes. Typical furnace temperatures: inlet ≈1000–1100 °C, outlet ≈1100 °C; carrier gases: oxygen and argon with humidified argon stream; absorption solution: ultrapure water; final absorption volume after adsorption: 11.1 mL.
  • Adsorption: TXA-04 absorption unit containing granular activated carbon (GAC) for trapping AOF from sample solutions.
  • Ion chromatograph: Nexera IC (Shimadzu) configured for anion analysis. Column: Shim-pack IC-SA2 (250 mm × 4.0 mm, 9 µm) with guard; carbonate-bicarbonate eluent (~1.8 mmol/L Na2CO3 and ~1.7 mmol/L NaHCO3); flow 1.0 mL/min; column temperature 30 °C; injection volume 50 µL; suppressor unit ICDS-Ai; detection by conductivity.
  • Software: C-IC Solution integrated control and QC software to coordinate combustion runs and IC acquisition, construct EPA 1621 analytical sequences, perform automatic conversions and QC checks, and monitor run status.

Main results and discussion


Calibration and linearity
  • Combustion calibration used five levels prepared by combusting 200 µL aliquots of sodium fluoride solutions; calculated fluoride concentrations in the adsorbent solution covered the working range. Recoveries across calibration levels fell between 95 and 100%, complying with EPA 1621 acceptance (80–120%).
  • Retention time window was established from triplicate standard injections and set to ±0.1 min.

Initial performance and precision
  • Relative standard error (RSE) calculated for calibration response was 2.7% (well below the 20% EPA criterion).
  • IPR testing (four replicate analyses of PFHxS standard) produced an RSD of 2.07% and average recovery of 94.8%, both well within EPA 1621 requirements (RSD ≤ 20%, recovery 80–120%).

Method blank, MDL and ML
  • Method blank (ultrapure water passed through GAC and processed) measured 0.57 µg/L fluoride—below the EPA 1621 blank acceptance limit of 4.0 µg/L.
  • MDL was determined following EPA/40 CFR Appendix B procedures. The reported MDL was 1.16 µg/L and the ML (rounded to a reporting-friendly value according to EPA rules) was set to 5 µg/L.

Matrix spike and ongoing QC
  • Matrix spike (river water spiked to 25 µg/L) and a 2.5× diluted MSD were analyzed. MS recovery = 92.6%; MSD recovery = 98.7%; RPD between MS and MSD = 6.3%, satisfying EPA performance criteria for spiked samples (recovery and RPD limits).
  • C-IC Solution automated QC checks for retention time, recovery, RPD and other criteria, simplifying batch acceptance decisions.

Benefits and practical applications


The integrated CIC solution offers several practical advantages:
  • End-to-end automation: linkage of the AQF-5000H and Nexera IC under C-IC Solution software enables automated batch creation, combustion scheduling, IC acquisition and QC evaluation, reducing manual steps and risk of human error.
  • Regulatory compliance: procedures and QC criteria were implemented to follow EPA 1621, facilitating standardized AOF screening in environmental and water-quality labs.
  • Comprehensive screening: AOF measurement by CIC quantifies total organically bound fluorine (including PFAS mixtures and transformation products) which complements targeted PFAS analyses for risk prioritization.
  • Robust analytical performance: demonstrated low method blanks, acceptable MDL/ML, tight precision and good recoveries support routine screening use.

Future trends and potential applications


Areas likely to advance and opportunities for wider use include:
  • Lower detection limits: improvements in trapping efficiency, combustion design and IC sensitivity to push MDLs lower and better detect trace AOF in challenging matrices.
  • Workflow integration: expanded interfaces between CIC platforms and laboratory information management systems (LIMS) for automated reporting and regulatory submissions.
  • Coupling total-fluorine screening with targeted speciation: using AOF as a screening gate followed by non-targeted or targeted mass-spectrometric techniques to identify priority fluorinated compounds.
  • Online and high-throughput sampling: adapting adsorption and combustion hardware for automated or semi-continuous sample processing to increase throughput for monitoring programs.
  • Standardization and QA materials: development of certified reference materials for AOF and inter-laboratory comparison schemes to harmonize reporting and improve comparability.

Conclusion


The combined AQF-5000H combustion unit, Nexera IC ion chromatograph and C-IC Solution software provide a validated, automated workflow for AOF screening in accordance with EPA 1621. The system demonstrated acceptable calibration performance, low blanks, reproducible precision and recoveries, and convenient QC automation. This integrated CIC approach supports reliable environmental screening of total organic fluorine and can streamline laboratory operations where AOF screening is required.

References


  • EPA 1621 Screening Method for the Determination of Adsorbable Organic Fluorine (AOF) in Aqueous Matrices by Combustion Ion Chromatography.
  • U.S. Code of Federal Regulations, 40 CFR Appendix B to Part 136—Definition and Procedure for the Determination of the Method Detection Limit, Revision 2.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Analysis of AOF (Adsorbable Organic Fluorine) According EPA Method 1621
Screening of PFAS compounds in wastewater using adsorbable organic fluorine with combustion ion chromatography (CIC)
Analysis of Adsorbable Organic Fluorine (AOF) by Combustion Ion Chromatography (CIC)
Enhanced screening of PFAS compounds in wastewater: Implementing U.S. EPA Method 1621 with improved combustion-ion chromatography