Quantitative Analysis of Selected Per- and Polyfluoroalkyl Substances (PFAS) in Cosmetic Products by LC-MS/MS

Applications | 2026 | ShimadzuInstrumentation
LC/MS, LC/MS/MS, LC/QQQ, Sample Preparation
Industries
Environmental
Manufacturer
Shimadzu

Significance of the topic


The presence of per- and polyfluoroalkyl substances (PFAS) in consumer goods, including cosmetics, raises environmental and health concerns because of PFAS persistence, bioaccumulation potential and specific functional roles (e.g., water/oil repellency and film formation) that make them intentionally used ingredients in formulations. Reliable, low-level quantitative methods are therefore essential for regulatory monitoring, product safety assessment and quality control in the cosmetics industry.

Objectives and study overview


This application study describes a targeted LC–MS/MS workflow for the quantitative determination of 29 PFAS in a range of cosmetic matrices (foundation, sunscreen, lip balm, lipstick, moisturizing cream, serum cream). The main objectives were to: (1) develop and validate an extraction/cleanup and analysis procedure that minimizes background PFAS contamination; (2) achieve robust quantitation via isotope-dilution internal-standard calibration over 0.5–25 ng/mL; and (3) demonstrate applicability to real cosmetic products.

Methodology


Sample extraction combined QuEChERS and solid-phase extraction (SPE) cleanup to reduce matrix interferences and background contamination. Key procedural steps included:
  • Weigh ~1.0 g sample into PFAS-free tube, spike isotopically labelled internal standard (ISTD) at 5 ng/mL.
  • Extract with 10 mL acetonitrile (1% acetic acid), shake and sonicate, add 10 mL water, then QuEChERS salts; centrifuge and collect upper organic layer.
  • dSPE cleanup (QuEChERS II) followed by dilution to 25 mL with water.
  • SPE cleanup using Biotage EVOLUTE PFAS 533 (500 mg/6 mL) on a positive-pressure manifold: condition with methanol and phosphate buffer (0.1 M, pH 7), load diluted extract, elute with ammonium hydroxide in methanol, combine eluates, evaporate under N2 at 50 °C and reconstitute in methanol/acetonitrile (90:10, v/v).

Calibration and quantification used isotope-dilution internal-standard calibration with mixed native standards prepared in methanol/acetonitrile (90:10) over 0.5–25 ng/mL. Linearity, LOD/LOQ and QC monitoring were assessed across the sequence.

Instrumentation used


Analysis was performed using Shimadzu hardware and accessories configured to reduce PFAS background:
  • UHPLC: Nexera X3 (Nexera LC-40 platform) with PFAS-Free Kit (PFAS-free mobile phase tubing, in-line filters and delay column).
  • Analytical column: Restek Ultra IBD, 100 mm × 2.1 mm, 3 μm (P/N 9175312).
  • MS: Shimadzu LCMS-8060RX triple-quadrupole, electrospray ionization in negative mode.
  • Chromatographic conditions: column 45 °C, injection 5 μL (co-injection 15 μL water), mobile phase A: 10 mM ammonium acetate (water), B: 10 mM ammonium acetate (MeOH); flow 0.3 mL/min; gradient 5% B (0–4 min) → 90% B (23–25 min) → 5% B (25.1–30 min).
  • MS source/gas conditions: nebulizing 3 L/min, drying 5 L/min, heating gas 15 L/min, DL 150 °C, heat block 250 °C, interface 100 °C.

Main results and discussion


Analytical performance:
  • Calibration: all 29 PFAS showed excellent linearity across 0.5–25 ng/mL with R2 > 0.99; calibration accuracy within 80–120% for standards.
  • LOD/LOQ: limits were determined from signal-to-noise (compound-specific values summarized in the study); method quantitation was demonstrated at low ng/mL levels.
  • QC: periodic QC injections (5 ng/mL) showed accuracy within 80–120% for nearly all targets (one compound, 11Cl-PF3OUdS, was 77% in the last QC run), indicating acceptable sequence stability.

Application to cosmetic matrices:
  • Most tested cosmetic samples returned concentrations below the quantitation limit for the 29 targeted PFAS.
  • Spike-recovery experiments at 25 ng/mL generally produced acceptable recoveries, but matrix-dependent deviations were observed: FOSA showed reduced recovery in foundation, lip balm and sunscreen; 9Cl-PF3ONS produced elevated recovery in liquid lipstick. These anomalies point to residual matrix effects and the limitations of non-matrix-matched calibration for certain analyte-matrix combinations.
  • Use of PFAS-free materials and a delay column meaningfully reduced analytical background and improved confidence at trace levels.

Benefits and practical applications


The described workflow offers practical advantages for laboratories performing targeted PFAS monitoring in cosmetics:
  • High sensitivity and robust quantitation using isotope-dilution LC–MS/MS enables trace-level determination across diverse product types.
  • QuEChERS followed by targeted SPE provides efficient matrix reduction and cleaner extracts, lowering ion suppression and protecting MS source cleanliness.
  • Adoption of PFAS-free LC consumables and delay columns reduces false positives from laboratory contamination, a common challenge in PFAS analysis.
  • The method is adaptable: additional PFAS targets can be incorporated as regulatory and analytical needs evolve.

Future trends and opportunities


Potential developments and extensions include:
  • Integration of high-resolution mass spectrometry (e.g., LCMS-9050) for broader non-targeted PFAS screening and unknown identification, complementing targeted triple-quadrupole workflows.
  • Development of matrix-matched calibration or enhanced sample-matched surrogates to mitigate observed matrix-dependent recovery biases.
  • Standardization of PFAS-free materials and multi-laboratory validation to improve comparability across labs and support regulatory monitoring.
  • Expansion of targeted lists to include newly reported short-chain and replacement PFAS chemistries as they emerge in product formulations.

Conclusion


The presented method combining QuEChERS extraction, dSPE cleanup and targeted SPE, coupled with isotope-dilution LC–MS/MS on the LCMS-8060RX, provides a practical and sensitive workflow for quantifying a broad panel of PFAS in varied cosmetic matrices. While the approach effectively reduces matrix interference and laboratory background, some matrix-specific recovery issues were observed, underscoring the value of QC monitoring and, where needed, matrix-matched strategies. The protocol is extensible to HRMS platforms for non-targeted screening and is suitable for laboratories conducting routine PFAS surveillance in cosmetics.

Reference


  1. U.S. Food and Drug Administration. Intentional use of PFAS in cosmetics and functional properties; report summary (2025).
  2. United States Environmental Protection Agency. Increasing Our Understanding of the Health Risks from PFAS and How to Address Them; EPA informational resources on PFAS.
  3. U.S. EPA Method 533: Determination of Per- and Polyfluoroalkyl Substances in Drinking Water by Isotope Dilution Anion Exchange Solid Phase Extraction and Liquid Chromatography/Tandem Mass Spectrometry.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Drinking Water in Accordance with EPA 533
Robustness Evaluation of PFAS Analysis in Soil Using LCMS -8060RX
Breakthrough Sensitivity and Robustness for PFAS Analysis in Chicken Tissue for EPA Method 1633A
Simultaneous Analysis of 40 PFAS Components in Drinking Water Using a Triple Quadrupole LC/MS/MS