ICP-MS Elemental Impurity Analysis of Pharmaceuticals Following USP and ICH Protocols — Summary
Importance of the topic:
- Elemental impurities in pharmaceuticals can present toxicological risks and must be controlled to meet regulatory limits set by ICH Q3D(R2) and USP General Chapters <232> and <233>.
- Accurate, interference‑free measurement at low ppb levels is critical for quality control of finished drug products, driving demand for robust ICP‑MS methods that are suitable for routine high‑throughput laboratories as well as regulated release testing.
Objectives and study overview:
- Develop and validate a single‑mode ICP‑QQQ method to quantify the 24 elemental impurities listed in ICH Q3D(R2) in an over‑the‑counter (OTC) oral drug matrix, following USP <233> analytical and validation criteria.
- Demonstrate that the Agilent 9500 Triple Quadrupole ICP‑MS operated in Advanced Helium Mode (AHM) can suppress polyatomic interferences, maintain sensitivity across the mass range, and meet detectability, precision, specificity, and accuracy acceptance limits.
- Assess the need for complementary Air cell (O2 reactive gas) mass‑shift measurements for specific doubly charged interferences from rare‑earth elements (REEs).
Methodology and workflow:
- Regulatory basis: Permitted Daily Exposures (PDEs) from ICH Q3D(R2) were converted to target solution concentrations (J values) assuming Option 1 (10 g/day max oral dose) and a total dilution factor; J values guided calibration and spike levels.
- Sample preparation: 0.36 g of OTC oral drug was microwave‑digested (CEM MARS 6) with 4 mL HNO3, 1 mL HCl and 5 mL ultrapure water, followed by addition of 0.1 g ascorbic acid to stabilize osmium and prevent Os loss. Final dilution produced a 500× overall dilution.
- Calibration and QC: Calibration standards prepared in the same acid matrix (2% HNO3 / 0.5% HCl) at 0.5, 0.8, 1.0 and 1.5 J using Agilent USP 232 Oral Kit_v2; internal standards (Sc, Ge, In, Te, Lu, Bi) supplied and diluted for online addition.
- Analytical strategy: Agilent 9500 ICP‑QQQ operated in Advanced Helium Mode (AHM) as the primary cell‑gas setting to remove polyatomic interferences through KED and CID mechanisms. Air cell mode (O2) with MS/MS mass‑shift was used selectively to mitigate doubly charged REE++ interferences on As and Se when required.
- Performance testing: USP <233> procedures were followed for detectability (spikes at 0.8 and 1.0 J), repeatability (six replicates at each spike level), intermediate precision (12 results across different days/analysts), and quantitative accuracy across spike levels 0.3, 0.8, 1.0 and 1.5 J.
Instrumentation used:
- Agilent 9500 Triple Quadrupole ICP‑MS with Dual‑Cell System (DCS) capable of Advanced Helium Mode (AHM) and Air cell (O2) modes.
- Sample introduction: MicroMist glass concentric nebulizer, quartz spray chamber, quartz torch with 5 mm ID injector; nickel sampler and skimmer cones.
- Autosampler: Agilent SPS 4. Software: Agilent OpenLab ICP‑MS with preset USP/ICH batch templates.
- Digestion system: CEM MARS 6 microwave digestion system. Calibration kits: Agilent ICH Q3D & USP 232 Orals Kit_v2 and associated internal standard solutions.
Key results and discussion:
- Sensitivity and detection limits: AHM delivered low detection limits (DLs) and low background equivalent concentrations (BECs) for all 24 target elements, including traditionally interfered species (V, Cr, As, Se) even in a chloride‑rich matrix introduced by HCl for Hg stabilization.
- Calibration and linearity: Representative calibration curves (Li, Se, Cd, Hg) showed excellent linearity (R2 > 0.999) across the calibration range tied to J values.
- Throughput gains: AHM provided substantially improved low‑mass sensitivity, enabling a typical integration time reduction for low‑mass elements from ~6 s to ~0.3 s (≈20×), and an overall reduction in total integration time by ~50% versus conventional multitune approaches. Eliminating gas switching further reduced cycle time.
- Specificity and interference control: AHM suppressed many polyatomic ions (e.g., 35Cl16O+, 36Ar16O+, 40Ar35Cl+), yielding low BECs on 51V, 52Cr, 75As and 78Se. For matrices containing elevated REEs producing REE2+ interferences, Air cell mode with MS/MS mass‑shift (e.g., As→AsO, Se→SeO) effectively removed problematic doubly charged interferences.
- Accuracy and recoveries: Spike recoveries at 1.0 J ranged from 92 to 106%; recoveries across 0.3–1.5 J were within the USP <233> acceptance window (70–150%), including acceptable results at 0.3 J (30% of J), demonstrating the method’s quantitative robustness below the target concentration.
- Precision: Instrumental repeatability (six replicates at 1.0 J) produced primary isotope RSDs <3%; intermediate precision across 12 measurements (different days/conditions) met USP <233> criteria (RSD <25%).
Practical benefits and applications:
- Single‑mode (AHM) analysis simplifies method setup and routine operation, reducing operator burden and minimizing risk of errors from gas‑mode switching.
- Reduced integration times and elimination of multitune modes increase sample throughput, important for high‑volume QC laboratories.
- Robust interference control in chloride‑containing digests enables the use of HCl for stabilization of Hg and Os without compromising results for V, As and Se.
- Flexible workflow: Air cell mass‑shift is available as a complementary tool for matrices with REE interference issues, extending method applicability across diverse pharmaceutical and related sample types.
Future trends and potential uses:
- Broader adoption of entry‑level ICP‑QQQ systems in regulated QC labs to provide triple‑quadrupole benefits (superior interference removal) alongside simplified operation and lower cost of entry.
- Further automation of digestion, online QC, and data‑processing templates to accelerate compliance with regulatory workflows and reduce manual steps.
- Extension of single‑mode AHM workflows to other dosage forms and complex matrices, with selective use of reactive gas modes where doubly charged or unusual polyatomic interferences are present.
- Continued integration of software‑based compliance templates (e.g., USP/ICH presets) to streamline method validation, data review, and audit trails for GMP environments.
Conclusion:
- The Agilent 9500 ICP‑QQQ operated in Advanced Helium Mode provided a rapid, robust, and regulatory‑compliant method to quantify the 24 elemental impurities defined by ICH Q3D(R2) in an OTC oral drug matrix following USP <233>.
- AHM achieved effective polyatomic interference removal while preserving sensitivity across the mass range, enabling single‑mode operation with improved throughput and simplified workflows.
- All acceptance criteria for detectability, precision, intermediate precision, specificity, and accuracy were met; recoveries and precision values demonstrate suitability for routine pharmaceutical elemental impurity testing. Air cell mass‑shift offers a targeted remedy for REE++ interferences when necessary.
References:
- International Council for Harmonisation (ICH). ICH Q3D (R2): Guideline for Elemental Impurities. Step 4, April 2022.
- United States Pharmacopeia General Chapter <232> Elemental Impurities—Limits. USP–NF. United States Pharmacopeial Convention.
- United States Pharmacopeia Elemental Impurities—Procedures, General Chapter <233>, USP‑NF. United States Pharmacopeial Convention.
- Agilent Technologies. USP/and ICH Q3D Elemental Impurities Analysis: The Agilent ICP‑MS Solution, Agilent publication.