GD/MP/ICP-AES, Elemental Analysis
IndustriesMaterials Testing
ManufacturerLECO
Bulk Analysis of Cast Iron Using the GDS900 — Application Note Summary
Importance of the topic
The control of chemical composition in cast irons is critical for achieving target mechanical and functional properties (strength, wear resistance, corrosion resistance). Rapid, accurate bulk elemental analysis in foundry and QC environments enables process control, alloy verification, and troubleshooting. The GDS900 glow discharge atomic emission spectrometer (AES) addresses routine multi-element determination in a range of cast-iron types with high throughput and reduced matrix bias.
Objectives and overview
This application note documents use of the LECO GDS900 for bulk elemental analysis of various cast-iron types (chilled ductile, chilled, as-cast gray). It describes sample preparation, calibration approach, analysis timing, and presents representative analytical results comparing measured averages and precision to certified reference values. The goal is to demonstrate the instrument's suitability for routine production/laboratory workflows.
Methodology
The GDS900 employs a glow discharge source to sputter material from conductive solid samples and excite atoms in a plasma separated from the sample surface. Emission is recorded by a CCD-based spectrometer. Key methodological attributes:
- Sputtering produces a uniform, surface-removed sample region that minimizes effects of prior metallurgical history.
- Spectra are dominated by neutral atomic lines, reducing spectral complexity and interferences compared with some other AES techniques.
- Typical analytical response is linear across concentration ranges, simplifying calibration and wavelength selection.
Instrumentation used
- LECO GDS900 glow discharge atomic emission spectrometer (CCD-based detection)
- Accessories for surface preparation: LECO BG belt grinder and LECO PX polisher
Sample preparation and accessories
Preparation varied by sample form:
- Chilled cast iron: surface prepared using a 120-grit zirconium oxide belt.
- As-cast samples (gray/ductile): polished to a 600-grit finish.
Calibration and drift control
Calibrations are factory-installed and built from certified reference materials (CRMs) and reference materials (RMs) supplied by vendors such as Brammer, CKD, CTIF, and MBH. Customer-supplied calibration pieces can supplement the factory set to tailor calibrations to specific composition ranges. Drift correction is performed using homogeneous, non-certified set-up standards (SUSs); when SUS materials are not suitable, RMs and CRMs may be used for drift control.
Analysis times and throughput
The instrument supports multiple consecutive burns at the same spot without dropping the sample, enabling high throughput. Typical timing examples reported:
- Single burn workflow: total ~70 seconds (including start-up/pre-burn and analyze steps).
- Three burns without dropping sample (consecutive burns option): total ~90 seconds, enabling roughly three analyses in ~1.5 minutes.
- As-cast single burn sequence: approximately 190 seconds (longer pre-burn or stabilization requirements).
Typical results and discussion
Representative data for chilled ductile, chilled, and as-cast gray iron standards show strong agreement with certified values for major elements (Fe, C, Si) and alloying elements (Cr, Ni, Mn, Mo, Cu, etc.). Precision is generally good: many elements show RSDs below 1–2% across replicate runs. Elements present at very low concentrations (e.g., P, Pb, S in some cases) exhibit higher relative variability (RSDs up to several percent), which is typical at trace levels. Key observations:
- Measured averages closely match certified concentrations, indicating accuracy of factory calibrations and the effectiveness of glow-discharge excitation to minimize matrix effects.
- Low spectral complexity in glow discharge spectra simplifies wavelength selection and reduces inter-element interference, contributing to consistent linear response.
- Repeatability across three runs demonstrates stable sputtering behavior and reliable detector performance (CCD).
Benefits and practical applications
- Fast throughput suitable for production QC and metal testing laboratories—multiple analyses without sample movement reduce cycle time.
- Reduced matrix influence and fewer spectral interferences relative to some other emission techniques, improving robustness across different cast-iron microstructures (chilled, ductile, gray).
- Simplified calibration strategy due to linear analyte response, and flexibility to use vendor CRMs/RMs plus customer standards.
- Minimal, straightforward sample preparation (grinding/polishing) compatible with routine workflows.
Future trends and potential applications
Expected developments and opportunities include:
- Expanded calibration libraries and automated calibration transfer to improve accuracy across more alloy grades and trace levels.
- Integration with manufacturing process control (inline or near-line sampling) to provide faster feedback for foundry process optimization.
- Enhanced data processing techniques (multivariate corrections, machine learning) to further reduce residual matrix effects and improve detection limits for challenging trace elements.
- Improved automation for sample handling and surface preparation to increase throughput and reduce operator variability.
Conclusion
The LECO GDS900 glow discharge AES provides a robust, high-throughput solution for bulk elemental analysis of cast irons. Its glow-discharge excitation, combined with CCD detection, yields accurate and precise multi-element results across chilled, ductile, and gray iron types with relatively simple sample preparation. The system is well suited for production laboratories and QA/QC environments seeking fast, repeatable alloy verification and process monitoring.