Introduction to Gas Phase FTIR Spectroscopy

Brochures and specifications | 2022 | Thermo Fisher ScientificInstrumentation
FTIR Spectroscopy
Industries
Energy & Chemicals
Manufacturer
Thermo Fisher Scientific

Importance of the Topic


Fourier transform infrared (FTIR) gas-phase spectroscopy is a versatile analytical method widely used for gas identification and quantification across environmental monitoring, industrial process control, semiconductor manufacturing, and speciality gas quality assurance. Its ability to measure many IR-active species simultaneously, across a broad dynamic range (from parts-per-trillion/parts-per-billion up to percent concentrations), with minimal sample preparation and highly stable calibrations, makes it a practical choice for continuous emissions monitoring, contaminant detection in high-purity gases, and fast transient measurements.

Objectives and Overview


This brochure introduces the principles of FTIR gas spectroscopy, outlines how a typical FTIR gas analysis system operates, describes practical sample-conditioning considerations, and presents Thermo Scientific product solutions for gas-phase FTIR analysis. The material aims to explain capabilities, limitations, and best practices so practitioners can select and configure FTIR instrumentation for target analytes and applications.

Methodology and Instrumentation


FTIR gas analysis uses broadband infrared radiation from a blackbody source passed through an interferometer to encode all spectral frequencies simultaneously into an interferogram. A reference laser provides mirror-position calibration and timing. The interferogram is recorded by a detector, digitized, and converted by Fourier transform into an absorbance spectrum. Sample spectra are obtained by ratioing a sample scan to a background (evacuated cell or inert gas such as N2) to remove instrument-specific features and drift.

Key methodological points:
  • FTIR measures molecular vibrational–rotational transitions; each molecular structure yields a characteristic IR ‘‘fingerprint.’li>
  • Gas-phase spectra consist of sharp, narrow bands; high spectral resolution (typically ~0.5 cm-1) is required to resolve overlapping lines.
  • Time resolution can be high (up to ~5 scans per second) for detectors such as MCT; DTGS detectors are common for slower, longer-integrated measurements.
  • Sensitivity improves by co-adding repeated scans; pathlength selection (e.g., 10 cm, 2 m, 10 m) tailors dynamic range for percent down to ppb detection.
  • Diatomic molecules and noble gases (O2, N2, H2, He, Ar) are infrared-inactive and therefore not directly measurable by IR absorption.

Used Instrumentation


The brochure describes a family of Thermo Scientific FTIR analyzers and accessories for gas analysis, including:
  • MAX-iR FTIR Gas Analyzer — real-time, rugged analyzer with single-digit ppb detection for many gases and enhanced sensitivity using StarBoost technology (capable of pushing to low ppt in specific cases).
  • Antaris IGS FTIR Gas Analyzer — platform optimized for dedicated industrial/compliance monitoring.
  • Nicolet iG50 FTIR Spectrometer — customizable laboratory spectrometer for flexible sampling configurations.
  • MAX-ASC-10 Automated Sampling Console — heated, multiplexed sampling and zero/calibration selection for hot and wet streams.
  • MAX-Bev CO2 Purity Monitoring System — turnkey system for beverage-grade CO2 purity and impurity quantification, configurable with O2 sensor.
  • EMS-10 Continuous Emissions Monitoring System — integrated CEMS with sample multiplexer and factory automation software.
  • Antaris FTIR Gas Conditioning and Analysis System — rack-mounted turnkey solution including conditioning manifold.
  • Optical and detector components: interferometer with beamsplitter and moving mirror, reference laser, detectors such as DTGS and MCT, and white (multi-pass) gas cells with gold mirrors to extend pathlength.

Main Results and Discussion


The brochure highlights practical performance characteristics rather than new experimental data. Representative spectral examples show that FTIR can simultaneously monitor common gases such as CO2, CH4, HCl, NO, and SO2 within the same full-spectrum scan. Key operational conclusions include:
  • FTIR effectively quantifies dozens of IR-active species in a single measurement, enabling multi-component analysis where traditional single-analyte sensors or gas chromatography would be slower or less convenient.
  • Appropriate selection of spectral resolution and optical pathlength allows analysis from percent concentration levels down to low ppb, with detection limits improved by signal averaging and advanced instrument modes.
  • Background acquisition (evacuated cell or inert gas) removes instrument- and optical-path-related features, ensuring that spectral peaks in the final spectrum are attributable to the sample alone and allowing long-term calibration stability without daily span gases.
  • Sample conditioning (heated lines, pressure regulation, filtration, multiplexing) is essential for reproducible results, especially for hot, wet, or particulate-laden streams.

Benefits and Practical Applications


Principal practical benefits of gas-phase FTIR:
  • Simultaneous multi-species quantification—useful for combustion exhaust analysis, CEMS, catalyst reaction monitoring, process control, and specialty gas contaminant detection.
  • Large dynamic range—tailorable pathlengths and resolution permit routine measurement of both high-concentration matrix gases and trace contaminants in a single platform.
  • Non-destructive measurements with minimal sample preparation and compatibility with hot/wet samples when properly conditioned.
  • Fast transient response for second-by-second monitoring and capability to integrate with DCS/PLC systems for automated control and reporting.
  • Stable calibrations—background updates compensate for instrument drift, reducing the need for frequent external calibration gases.

Practical Considerations


For accurate and repeatable FTIR gas measurements, the brochure emphasizes:
  • Maintaining constant sample temperature and pressure to avoid density-related spectral changes.
  • Using heated sample lines and cells for combustion exhausts to prevent condensation.
  • Selecting appropriate cell pathlength and detector type (DTGS for slower, longer integrations; MCT for fast, high-sensitivity measurements).
  • Implementing sample filtration to protect optics from particulates and using multiplexed sampling consoles for multiple inputs and automatic zero/span checks.

Future Trends and Potential Uses


Anticipated developments and expanding applications for gas-phase FTIR include:
  • Further improvements in detection limits via enhanced optics, cryogenic detectors, and advanced noise-reduction algorithms, enabling routine ppt-level detection for some analytes.
  • Tighter integration with automation, cloud data systems, and process control networks for real-time decision-making and compliance reporting.
  • Application of machine learning and chemometric deconvolution to improve quantification in complex, overlapping spectral regions and to accelerate method development.
  • Broader adoption in emerging sectors such as renewable energy (biogas/anaerobic digestion gas profiling), indoor air quality/multipoint ambient monitoring, and beverage/gas purity verification.
  • Increased modularity and portability of FTIR systems for field-based source testing and rapid site surveys.

Conclusion


Gas-phase FTIR spectroscopy is a robust, flexible analytical technique capable of simultaneous multi-component measurement across a wide concentration range with high temporal resolution. When paired with appropriate sample conditioning and instrument configuration, FTIR offers reliable, stable quantitative performance for environmental monitoring, process control, high-purity gas qualification, and research applications. Modern FTIR platforms from Thermo Scientific demonstrate that the technique can meet stringent detection limits and regulatory needs while providing operational advantages over single-species sensors and slower chromatographic methods.

References


The summary is based on product and application information provided in a Thermo Fisher Scientific brochure for gas-phase FTIR spectroscopy (Thermo Fisher Scientific, BR52338_E, 12/2022).

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Advantages of a Fourier Transform Infrared Spectrometer
Advantages of a Fourier Transform Infrared Spectrometer
2015|Thermo Fisher Scientific|Technical notes
A tutorial on spectral resolution for the Nicolet iS5 FTIR Spectrometer
FTIR talk letter vol. 17
FTIR talk letter vol. 17
2012|Shimadzu|Others
Shimadzu FTIR talk letter vol. 36