Importance of the topic
Comprehensive forensic toxicology requires rapid, sensitive screening followed by robust confirmatory analyses. Traditional workflows often rely on multiple instruments and disparate software environments, increasing turn-around time, user training burden, and potential for reporting inconsistencies. Integrating high-resolution, untargeted screening with targeted confirmation in a single software platform streamlines data handling, improves traceability from putative hits to confirmed identifications, and supports retrospective review of acquired high-resolution data—important capabilities for modern forensic and clinical toxicology laboratories.
Objectives and study overview
This work demonstrates a consolidated screening-to-confirmation workflow using Shimadzu LabSolutions Insight Explore to process data from LC-QTOF screening, LC-triple quadrupole (TQ) targeted confirmation, and GC-single quadrupole (SQ) targeted confirmation. Key aims were to show rapid untargeted detection with high-resolution QTOF, efficient targeted quantitation by LC-MS/MS, and orthogonal confirmation by GC-MS, all managed within one software environment to simplify review, reporting, and retrospective queries.
Methodology and workflow
Overview of the end-to-end workflow:
- Sample types: Biological matrices (urine, blood) following extraction, cleanup, and addition of internal standards.
- Screening: LC-QTOF (untargeted MS1/MS2, data independent acquisition) to detect features and produce library-searchable spectra.
- Confirmation: Targeted LC-MS/MS (MRM) for high-sensitivity quantitation and GC-MS (Scan/SIM) for orthogonal confirmation of volatile/thermostable targets.
- Data processing: Single-platform tools for peak detection, formula finding, library searching (including reverse search), hit scoring/flagging, and quantitative calibration with review-by-exception.
Used instrumentation
- LC-QTOF: Shimadzu LCMS-9050 used for high-resolution screening with DIA (data independent acquisition) to allow untargeted MS2 collection and retrospective data mining. LC conditions employed a Shim-pack Scepter PFPP-120 column (1.9 µm, 2.1 x 50 mm), water/methanol mobile phases with ammonium formate and formic acid, 0.4 mL/min flow and 5 min run time.
- LC-TQ: Shimadzu LCMS-8050/8050RX (triple quadrupole) for fast targeted quantitation using rapid polarity switching and MRM to quantify 46 compounds with 23 isotopically labeled internal standards in under three minutes.
- GC-SQ: Shimadzu GCMS-QP2020NX for confirmation of >50 commonly encountered toxicology targets using Scan plus SIM acquisitions. Column: Rtx-5ms (30 m x 0.25 mm, 0.25 µm); oven program 60 °C (2 min) to 320 °C at 15 °C/min; split injection at 300 °C.
- Software: LabSolutions Insight Explore (single-platform environment) for combined screening, library searching (compatible with multiple library formats including NIST and vendor libraries), quantitation, review and reporting.
Main results and discussion
Key experimental outcomes:
- Chromatographic separation resolved isobaric/opiate pairs such as morphine/hydromorphone and codeine/hydrocodone, enabling reliable identification and quantitation of critical analytes.
- Rapid LC methods (5 min) and very short targeted LC-MS/MS runs (<3 min) provided high throughput while maintaining excellent linearity (many R2 > 0.995) across relevant concentration ranges.
- Limits of quantitation: Method LLOQs ranged down to low pg/ng per mL levels for several analytes (examples reported LLOQs as low as 0.005 ng/mL for some compounds), supporting sensitive detection of fentanyl analogs and emerging nitazene-class opioids.
- Library matching: QTOF MS2 spectra searched against Cayman Spectral Library and a Shimadzu Forensic Toxicology Database produced high similarity scores (e.g., Methadone 97%, Tramadol 89%), aided by reverse-search functionality for complex matrices.
- Combined Scan and SIM acquisition strategies allowed simultaneous non-targeted detection and targeted monitoring (e.g., methadone SIM chromatogram shown), which simplifies confirmation workflows without separate injections.
- Quantitation approach: Calibration used internal standards with linear curves and 1/x or 1/x2 weighting; acceptance required triplicate agreement within ±20% at each calibration level.
Benefits and practical applications
- Consolidation of screening and confirmation in one software reduces data transfer errors, shortens review times, and centralizes reporting for forensic casework.
- High-resolution QTOF screening with DIA permits retrospective interrogation for new/emerging drugs without re-running samples, an advantage when monitoring novel psychoactive substances.
- Orthogonal confirmation (LC-MS/MS and GC-MS) increases confidence in positive findings, particularly for isobaric or matrix-challenged targets.
- Fast chromatographic methods increase sample throughput for high-volume forensic and clinical toxicology laboratories while retaining required analytical performance.
Future trends and possibilities
- Broader adoption of single-platform solutions that integrate HRMS screening, targeted MS/MS confirmation, library management and automated reporting will likely continue as labs seek efficiency and traceability.
- Expanded spectral libraries and improved reverse-search algorithms will enhance identification confidence in complex biological matrices and for novel substances (e.g., nitazenes, carfentanil analogs).
- Further method validation in authentic matrices (planned urine recovery studies noted by the authors) and robustness testing across instrument generations will be important steps toward routine casework implementation and accreditation.
- Machine-learning assisted flagging and automated review-by-exception could further reduce analyst workload and improve consistency in large-scale screening programs.
Conclusion
This study demonstrates a practical, integrated screening-to-confirmation workflow combining LC-QTOF untargeted screening, LC-TQ targeted quantitation, and GC-SQ targeted confirmation within a single software environment. The approach yields sensitive, high-throughput analyses with robust library matching and streamlined data review. Adoption of this consolidated workflow can reduce operational complexity and enable retrospective HRMS-based investigations; however, full validation in relevant biological matrices is required prior to routine forensic use.
References
- Hiramatsu Y, Matos Mejías C, Monti SA, Smith JP, Wiest L, Gilles C. Screening to Confirmation: Complete Workflow for Comprehensive Forensic Toxicology Analysis Using a Single Software Platform. Shimadzu Scientific Instruments, Inc.; MP725. Authors affiliated with Shimadzu Scientific Instruments, Inc.