LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
IndustriesProteomics
ManufacturerThermo Fisher Scientific
Significance of the topic
Plasma proteomics is central to translational research and biomarker discovery because plasma is minimally invasive to collect and contains proteins spanning a large dynamic range that reflect physiological and pathological states. Hybrid acquisition strategies that combine targeted quantitation and data-independent acquisition (DIA) address the common trade-off between broad proteome coverage and the highest-sensitivity measurement of predefined low-abundance peptides. By integrating targeted MS2 (tMS2/PRM-style) or internal-standard-triggered PRM (SureQuant) with DIA discovery scans on a high-sensitivity instrument, researchers can simultaneously quantify prioritized biomarkers and retain unbiased discovery capability across large clinical cohorts.Objectives and study overview
This technical note developed and evaluated two hybrid-DIA strategies on the Thermo Scientific Orbitrap Astral Zoom MS: a tMS2 hybrid-DIA approach (scheduled targeted MS2 list combined with DIA) and a SureQuant hybrid-DIA approach (heavy-peptide–triggered PRM combined with DIA). Primary goals were to assess (1) proteome depth and peptide identification trade-offs as targeted panel size increases (30–300 peptides), and (2) analytical performance metrics (linearity, LOD, LOQ, precision, accuracy) for targeted peptides across wide dynamic ranges using a PQ500 reference peptide set spiked into pooled neat human plasma digest.Materials and methods
Key experimental design elements and workflows:- Two complementary hybrid-DIA workflows were implemented: (A) tMS2 hybrid-DIA: MS1 → scheduled targeted MS2 (PRM-style) → DIA; (B) SureQuant hybrid-DIA: Survey/triggering step for heavy peptides → SureQuant PRM triggered on heavy/light → DIA.
- Performance tests included (i) panel-size experiment with PQ500 heavy peptides spiked 1:20 into pooled plasma digest and panels of 30, 60, 120, 200, and 300 targeted peptides; (ii) serial dilution experiment (1:3 dilutions across 10 levels) to define linearity, LOD, LOQ, accuracy and precision for 300 targeted peptides.
- LC gradient: trap-and-elute scheme on a 150 µm × 15 cm, 2 µm C18 EASY-Spray column at 50 °C; 32-minute gradient optimized for throughput with a 0.8 µL/min flow.
- MS acquisition: Orbitrap Astral Zoom used for all experiments; DIA windowing, MS1 and targeted MS2 parameters were tuned for balance between sensitivity and scan speed (examples: MS1 resolution up to 240,000, DIA precursor range 350–980 m/z, tMS2 isolation widths and injection times adjusted between methods to accommodate targeted panels).
- Quantitation panel and samples: PQ500 reference peptides (Biognosys) spiked into pooled human plasma digest; disease plasma samples sourced from BioIVT for matrix context.
Used instrumentation
- Thermo Scientific AccelerOme Automated Sample Preparation Platform
- Thermo Scientific Vanquish Neo UHPLC System with trap-and-elute
- Thermo Scientific Orbitrap Astral Zoom Mass Spectrometer
- Thermo Scientific Easy-Spray Source and EASY-Spray HPLC column (2 µm C18, 150 µm × 15 cm)
- PQ500 Reference Peptides Kit (Biognosys) and PepMap Neo trap cartridge
Data analysis
- DIA raw files were converted with HTRMS converter and analyzed by Spectronaut (library-free directDIA) for protein/peptide identification and quantitation.
- Targeted quantitation, calibration curves and LOD/LOQ determination were performed using Skyline; LOD/LOQ defined via a three-times signal-to-noise criterion and serial dilution accuracy/precision assessments (triplicate injections).
- Summary statistics, plots and correlation analyses were generated in Python.
Main results and discussion
- Proteome depth: Hybrid-DIA retained substantial discovery depth despite adding targeted scans. For tMS2 hybrid-DIA the number of protein groups decreased by ~9.5% (from 692 to 593–645) and peptides by ~4.2% compared to standard DIA. For SureQuant hybrid-DIA the protein group decrease was ~12.8% and peptide IDs decreased by ~6.7% versus standard DIA.
- Dynamic range and sensitivity: Both hybrid methods detected protein intensities spanning more than six orders of magnitude. Analytical sensitivity on the Orbitrap Astral Zoom was notable—using tMS2 hybrid-DIA, >92% of 300 peptides had LOD <25 amol and >82% had LOQ <50 amol.
- Linearity and quantitative performance: Calibration across serial dilutions showed excellent linearity: 89% of peptides had R2 > 0.99 and 99% had R2 > 0.90. Example peptide (LEYLLLSR) showed linearity and accuracy maintained across ~6 orders of magnitude down to ~6 amol on-column at low levels.
- Precision and reproducibility: Protein quantity CVs were generally low. Reported CVs included ~3.2% for protein group ID counts (tMS2), while quantitative CV for protein quantities across standard DIA and hybrid-DIA experiments ranged ~5–7%; peptide-intensity CVs ranged ~8–13% depending on method and panel size.
- Effect of targeted panels on scan density: Data points per chromatographic peak decreased modestly with large targeted panels. For tMS2 hybrid-DIA mean DIA data points per peak fell from ~5.0–5.3 (30–120 peptides) to ~4.5–4.7 (200–300 peptides). SureQuant hybrid-DIA showed a smaller decrease (mean points ≈5.80 to 5.56 across panel sizes) because PRM acquisition is triggered only when heavy-peptide evidence is present.
- Impact on targeted intensities: Targeted peptide peak intensities in hybrid-DIA were slightly reduced relative to dedicated PRM/SureQuant runs: approximately 8–11% decrease for tMS2 hybrid-DIA (median ratio ~1.08–1.11) and ~13% decrease for SureQuant hybrid-DIA vs standard SureQuant PRM. Despite this, LOD/LOQ and quantitative linearity were not materially affected.
- Correlation with standard methods: Strong agreement was observed between hybrid-DIA and conventional workflows. Pearson correlations for protein quantities between hybrid-DIA and standard DIA and for targeted peptides between hybrid-DIA and PRM/SureQuant were typically r > 0.97–0.99, and cross-validation between the two hybrid approaches also yielded r > 0.97.
Benefits and practical applications
- Simultaneous discovery and robust targeted quantitation: Hybrid-DIA enables concurrent unbiased DIA profiling and sensitive quantification of prioritized low-abundance peptides in the same run, improving throughput for cohort studies and biomarker verification pipelines.
- Flexible trade-offs: The two strategies offer complementary strengths—tMS2 hybrid-DIA is simpler to set up and does not require heavy standards; SureQuant hybrid-DIA increases confidence in peptide identity and quantitation via heavy-peptide triggers at the cost of heavier method complexity and requirements for standards.
- High sensitivity suitable for clinical proteomics: Achieved sub-femtomole to low-attomole detection limits make both hybrid approaches attractive for plasma biomarker studies where target peptides are rare.
- Adaptability: The hybrid methods and parameters are applicable to other sample matrices and can be tuned for panel size, throughput and sensitivity needs.
Future trends and potential uses
- Automation and scaling: Combining automated sample prep with hybrid-DIA is suited to high-throughput clinical studies; further automation of assay setup (e.g., automated transition-list generation and retention-time alignment) will accelerate adoption.
- Intelligent acquisition refinement: On-instrument decision-making (adaptive RT windows, real-time trigger optimization) can maximize data points per peak while supporting larger targeted panels.
- Standardization and multiplexing: Wider use of heavy-peptide panels and community reference kits will improve comparability across labs; hybrid acquisition is promising for multiplexed verification studies that bridge discovery and targeted validation phases.
- Extension to PTMs and modified-peptide panels: Hybrid-DIA could be adapted to targeted phosphopeptide or PTM panels while preserving discovery-level DIA for site-localization and global signaling context.
Conclusion
This study demonstrates that two hybrid-DIA strategies implemented on the Orbitrap Astral Zoom MS—tMS2 hybrid-DIA and SureQuant hybrid-DIA—successfully combine DIA discovery with sensitive targeted quantitation. Both approaches maintain deep proteome coverage, deliver excellent linearity and low attomole-level sensitivity for many peptides, and produce highly correlated protein and peptide quantitation compared to standard DIA and PRM workflows. Choice between methods depends on experimental priorities: method simplicity and adaptive RT (tMS2) versus enhanced identification confidence using internal standards (SureQuant).References
- Martínez-Val A, et al. Hybrid-DIA: intelligent data acquisition integrates targeted and discovery proteomics to analyze phospho-signaling in single spheroid. Nature Communications. 2023;14:3599.
- Gajadhar A. SureQuant intelligence-driven MS: a new paradigm for targeted quantitation. Thermo Fisher Scientific Technical Note 65873.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.