In situ density determination of polyethylene in multilayer polymer films using Raman microscopy

Applications | 2018 | Thermo Fisher ScientificInstrumentation
RAMAN Spectroscopy, Microscopy
Industries
Materials Testing
Manufacturer
Thermo Fisher Scientific

Importance of the Topic


Polyethylene (PE) is one of the most produced plastics worldwide, with tunable density directly affecting mechanical, barrier, and processing properties. Accurate density measurement in multilayer films is critical for quality control, product development, and regulatory compliance in sectors such as packaging, pharmaceuticals, and consumer goods.

Objectives and Study Overview


This study aims to develop a non-destructive, in situ method for quantifying PE density within single-layer and multilayer films using confocal Raman microscopy combined with Partial Least Squares (PLS) regression. The approach targets rapid analysis without the labor-intensive sample preparation required by traditional ISO and ASTM methods.

Methodology and Instrumentation


A total of 25 PE samples (12 pellets, 13 films) spanning densities from 0.8598 to 0.9620 g/cm3 were analyzed. A Thermo Scientific DXR2 Raman Microscope equipped with a 532 nm laser (2 mW for pellets/films; 5 mW for depth profiling) and objectives of 10× (50 µm slit) and 50× (25 µm confocal pinhole) was used. Spectra were acquired for 30 s per point and processed in OMNIC software. TQ Analyst software normalized spectra using CH2 bending (1422–1452 cm–1) peak area and developed a 3-factor PLS model over 1400–1500 cm–1. Model validation employed five independent standards.

Main Results and Discussion


  • Distinct spectral differences in CH2 stretching (2848 vs 2882 cm–1) and bending/twisting (1416 vs 1440 cm–1) enable discrimination of HDPE, LDPE, and LLDPE.
  • The 3-factor PLS calibration achieved R²=0.9914, RMSEC=0.00360 g/cm3 and RMSEP=0.00432 g/cm3, with all samples within ±0.81% of actual density.
  • Application to an unknown film predicted 0.9014 g/cm3 versus the actual 0.9008 g/cm3.
  • Confocal depth profiling of a transdermal nicotine patch identified two distinct PE layers: Layer 1 at 0.9150 g/cm3 (LDPE/LLDPE) and Layer 2 at 0.9583 g/cm3 (HDPE), matching expected functional roles.

Benefits and Practical Applications


  • In situ density determination eliminates tedious microtoming and solvent-based layer separation.
  • Applicable to both pellet and film samples, including complex multilayer structures.
  • Supports rapid, non-destructive quality control and reverse engineering workflows.

Future Trends and Potential Applications


  • Integration with advanced chemometric algorithms and machine learning for broader polymer systems.
  • Extension to other physical properties such as crystallinity and molecular orientation in situ.
  • Potential use in real-time process monitoring and high-throughput screening in industrial settings.

Conclusion


Confocal Raman microscopy combined with PLS regression provides an accurate, efficient, and non-destructive method for quantifying PE density in multilayer films. The approach simplifies workflows, reduces sample preparation, and offers robust performance across diverse sample formats.

Reference


  1. Piringer O.G. and Baner A.L., Plastic Packaging: Interactions with Food and Pharmaceuticals, 2nd ed., Wiley-VCH, 2008.
  2. The Essential Chemical Industry, Polyethylene, 2017.
  3. ISO 1183-1:2012; ASTM D792-13, Methods for Density and Specific Gravity of Plastics.
  4. ISO 1183-2:2004; ASTM D1505-10, Density-Gradient Technique for Plastics.
  5. ASTM D4883-08, Density of Polyethylene by Ultrasound Technique.
  6. Mieth A., Hoekstra E., Simoneau C., EUR 27816 EN, 2016.
  7. Sato H. et al., J. Appl. Polym. Sci., 2002, 86, 443–448.
  8. Strobl G.R., Hagedorn W., J. Polym. Sci. B Polym. Phys., 1978, 16, 1181–1193.
  9. Williams K.P.J., Everall N.J., J. Raman Spectrosc., 1995, 26, 427–433.
  10. Ibrahim M., He H., Thermo Scientific Application Note AN52301, 2017.
  11. Thermo Scientific, TQ Analyst Software Chemometric Algorithms, 2009.
  12. Kearney C.J., Mooney D.J., Nat. Mater., 2013, 12, 1004–1017.
  13. Allen L.V., Ansel H.C., Ansel’s Pharmaceutical Dosage Forms and Drug Delivery Systems, 10th ed., Lippincott Williams & Wilkins, 2013.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Classification of polyethylene by Raman spectroscopy
Classification of polyethylene by Raman spectroscopy
2022|Thermo Fisher Scientific|Applications
Confocal Raman Microscopy Analysis of Multilayer Polymer Films
Confocal Raman Microscopy Analysis of Multilayer Polymer Films
2008|Thermo Fisher Scientific|Applications
Density and Copolymer Content in Polyethylene Samples by FT-NIR Spectroscopy
XRD investigation of three types of polyethylene films with ARL EQUINOX 100 X-ray Diffractometer