LC-MS/MS Analysis of Ethylene Glycol in Blood Using a Simple Derivatization Method

Applications | 2025 | ShimadzuInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Energy & Chemicals
Manufacturer
Shimadzu

Significance of the Topic


Ethylene glycol is a low-molecular-weight diol extensively used in antifreeze and industrial applications. Despite its utility, it poses serious health risks upon ingestion, causing metabolic acidosis and renal failure. Reliable and rapid quantification of ethylene glycol in blood is therefore critical for clinical toxicology and emergency response.

Study Objectives and Overview


This work describes the development of a streamlined LC-MS/MS method for measuring ethylene glycol in whole blood. By integrating protein removal and chemical derivatization into a single step, the approach aims to simplify sample preparation, minimize analysis time, and avoid expensive isotope-labeled standards.

Methodology and Instrumentation


The method employs 2-bromopyridine-5-boronic acid (BPBA) to form a stable cyclic boronic ester (EG-BPBA) with the diol groups of ethylene glycol. Key elements include:
  • Sample Pretreatment: 100 µL of blood is spiked with internal standard (1,2-pentanediol), mixed with 0.1% TFA in acetonitrile containing 2.5 mM BPBA, and vortexed to achieve simultaneous deproteinization and derivatization. After centrifugation, the supernatant is directly injected.
  • Chromatography: Separation is performed on a Shim-pack Velox HILIC column (2.1 × 100 mm, 2.7 µm) at 30 °C using a binary gradient of 10 mM ammonium formate with 0.1% formic acid in water (A) and methanol (B).
  • Mass Spectrometry: Analysis is carried out on a Nexera XR coupled to an LCMS-8050 RX triple quadrupole MS. Multiple reaction monitoring (MRM) transitions 228→148 and 230→148 (EG-BPBA) are used alongside internal standard transitions.

Main Results and Discussion


Calibration and sensitivity assessments demonstrated excellent linearity (R=0.99996) over 10–500 µg/mL, covering clinically relevant concentrations well below toxic thresholds. A high-concentration test confirmed negligible carryover of BPBA through standard solvent rinses. Spiked whole blood at 100 µg/mL yielded accuracy of 96.4% and repeatability (RSD) of 2.4%. Blank blood showed only 5% of the signal observed at the 10 µg/mL reference level, confirming method specificity.

Benefits and Practical Applications of the Method


This protocol offers rapid sample throughput by combining cleanup and derivatization in a single vial, reducing turnaround time to under ten minutes. It employs common mobile phases compatible with existing toxicology screening workflows, allowing ethylene glycol measurement to be integrated into routine drug screening systems by a simple column swap.

Future Trends and Opportunities


Further enhancements may focus on automation of sample addition and centrifugation to support high-volume clinical laboratories. Expansion to multiplexed assays could enable simultaneous quantification of multiple toxic diols. Emerging high-resolution MS techniques might improve selectivity for complex biological matrices.

Conclusion


A quick, cost-effective LC-MS/MS method for ethylene glycol in blood was established using BPBA derivatization. The streamlined workflow demonstrates high sensitivity, accuracy, and minimal carryover, making it suitable for emergency toxicology and routine clinical analysis.

References


  • S Shen, F Zhang, S Zeng, J Zheng, An approach based on liquid chromatography/electrospray ionization–mass spectrometry to detect diol metabolites as biomarkers of exposure to styrene and 1,3-butadiene, Analytical Biochemistry, 386 (2009), p. 186-193.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Shimadzu Solutions for Application Notebook - Clinical Research - Application Notebook
Simultaneous Analysis of 27 Antidiabetic Drugs in Whole Blood by LC–MS/MS
Quantitation of Ethylene Glycol in Postmortem Specimens by GC-MS/MS
Evaluation of Fatty Acids Profling in a Blood Drop Spotted on DBS Card by using a Robot-assisted GC Method