Smart solutions for high analyte concentrations and carryover in LC-MS monitoring

Posters | 2025 | Thermo Fisher Scientific | MSACLInstrumentation
Software, LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Industries
Forensics
Manufacturer
Thermo Fisher Scientific

Importance of Topic


Liquid chromatography mass spectrometry methods often face samples with analyte levels above their calibration range leading to inaccurate quantification and risk of carryover. Integrating real time software controls safeguards data integrity and maintains analytical performance in clinical toxicology laboratories.

Study Objectives and Overview


This work demonstrates the System Suitability Test and Intelligent Run Control features in Chromeleon software. Using a panel of 31 drugs of abuse in oral fluids, analyzed on a Vanquish Flex UHPLC coupled to an Orbitrap Exploris 120, the study showcases automated responses when analyte concentrations exceed upper limits of quantification.

Methodology


Samples were prepared by dispersive solid phase extraction using DPX XTR mixed mode tips and analyzed by UHPLC MS with data dependent MS2 acquisition. Chromatographic separation employed a column wash protocol, while calibration standards ranged from 5 to 1000 ng/mL. Data processing and test case setup were performed in Chromeleon version 7.3.2 MUe.

Used Instrumentation


  • Thermo Fisher Vanquish Flex UHPLC system
  • Orbitrap Exploris 120 mass spectrometer
  • OptaMax NG HESI ion source
  • DPX XTR INTip dispersive SPE tips

Key Results and Discussion


Two test cases, AutoDilution_3x and AutoDilution_7x, compared unknown sample peak amounts to three and seven times the highest calibration level. When thresholds were exceeded, the software automatically inserted a blank with a column wash and reinjected the sample at reduced volumes of one third or one seventh. Sequence data confirmed proper pass and fail actions and prevention of carryover.

Benefits and Practical Applications


  • Automated dilution and reinjection reduce manual intervention and operator error
  • Real time threshold checks maintain quantification accuracy
  • Customizable test cases support diverse analytical workflows
  • Applicable across Thermo Fisher mass spectrometry platforms

Future Trends and Potential Applications


Emerging developments may include advanced adaptive control algorithms, integration with laboratory information management systems, machine learning to predict carryover risk, and expansion into environmental monitoring and pharmaceutical quality control.

Conclusion


The Chromeleon SST IRC functionality effectively detects analyte concentrations exceeding quantification limits and automatically applies wash and dilution steps. This innovation enhances data reliability and operational efficiency in clinical toxicology LC MS workflows.

References


  • Patterson C Hassell K Thermo Fisher Scientific Technical Note TN003851 2025
  • Patterson C et al Thermo Fisher Scientific Technical Note TN002363 2023

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 