Analysis of Saccharides Using Integrated HPLC and Size Exclusion-Ligand Exchange (Na Type) Column

Applications | 2025 | ShimadzuInstrumentation
HPLC
Industries
Food & Agriculture
Manufacturer
Shimadzu

Significance of the Topic


Saccharides are key contributors to sweetness and essential energy sources in foods. Accurate measurement of these compounds is critical for quality control, nutritional management, and mitigating health risks associated with excessive sugar intake.

Objectives and Study Overview


This work presents a streamlined approach to simultaneously quantify maltotriose, sucrose, glucose, and fructose in fruit juices using an integrated HPLC system equipped with a size exclusion–ligand exchange (Na type) column and differential refractive index detection.

Methodology and Instrumentation


  • Column: Shim-pack SUR-Na (250 mm × 7.8 mm I.D., 8 µm) with guard column.
  • Mobile phase: water only, eliminating the need for complex solvent preparation.
  • Detector: RID-20A (refractive index detector) set at 40 °C.
  • Column temperature: 80 °C; flow rate: 0.6 mL/min; injection volume: 10 µL.
  • Sample preparation: mixed standards and 50× diluted fruit juices filtered through 0.2 µm PVDF membranes.


Main Results and Discussion


Baseline separation of the four saccharides was achieved, with calibration curves from 20 to 2000 mg/L showing excellent linearity (r2 ≥ 0.9999). Six consecutive injections of 50 mg/L standards yielded retention time RSDs under 0.02% and peak area RSDs below 0.6%. Fruit juice analyses quantified sucrose (absent in grape juice), glucose, and fructose levels in orange, mandarin, grape, pineapple, and grapefruit samples. Spike‐and‐recovery tests at 500 mg/L produced recoveries between 86% and 100%.

Benefits and Practical Applications


  • Simultaneous profiling of major saccharides across various fruit juices.
  • Simplified operation using only water as the mobile phase.
  • High reproducibility and linearity ideal for routine QA/QC in the food industry.
  • Compact integrated HPLC footprint with straightforward RID integration.


Future Trends and Potential Applications


Potential expansions include adaptation to other carbohydrate‐rich matrices, coupling with mass spectrometry for enhanced structural characterization, and automation of sample preparation to increase throughput and analytical depth.

Conclusion


The described integrated HPLC method with size exclusion–ligand exchange (Na type) column and refractive index detection provides a robust, efficient, and reliable platform for saccharide analysis in fruit juices, demonstrating outstanding linearity, precision, and recovery.

Reference


  • World Health Organization (2015) Guideline: Sugars intake for adults and children

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Analysis of Saccharides Using Integrated HPLC and Size Exclusion-Ligand Exchange (Ca Type) Column
Fermentation Monitoring of Yeast Using Size Exclusion-ligand Exchange (Na-type) Column
Profiling and Quantification of Mono and Disaccharides and Selected Alditols in Juice, Beer, Wine, and Whiskey Using the ACQUITY Arc System with Mass Detection
Simultaneous Determination of 15 Saccharides Without Interference by Salt in Seasonings