Titration
IndustriesEnergy & Chemicals
ManufacturerMetrohm
Importance of the Topic
The accurate quantification of calcium in cement is critical for quality control, compliance with construction standards, and ensuring the performance of cement-based materials. Photometric titration provides a rapid, sensitive, and automatable approach to determine calcium levels following standardized digestion methods.
Aim and Study Overview
This application note describes a photometric titration procedure for calcium determination in cement according to the DIN EN 196-2 digestion protocol. After digestion, calcium is complexed with EDTA and the endpoint is detected photometrically using a chromogenic indicator at 610 nm.
Methodology and Instrumentation
The cement sample is digested per DIN EN 196-2 and an aliquot (10 mL) is adjusted to pH 12.5 with NaOH. A murexide indicator is added, and the solution is titrated with standardized Na2EDTA. The photometric endpoint is monitored at 610 nm, where the color shifts from pink to violet.
- Digestion: Acid digestion following DIN EN 196-2
- Titrant: 0.03 mol/L Na2EDTA (titered against Ca standard)
- Indicator: Murexide in NaCl matrix
- pH Adjustment: Saturated NaOH
- Endpoint Detection: Optrode sensor at 610 nm
Used Instrumentation
- 907 Titrando with 800 Dosino units
- 5 mL, 10 mL, 50 mL dosing units
- 802 Rod Stirrer, 815 Robotic USB sample processor
- Optrode photometric sensor (610 nm) with Ecotrode plus electrode
Key Results and Discussion
The method yields highly reproducible results: CaO 48.81 % ± 0.209 (n=6) and Ca 34.89 % ± 0.151 (n=6). The photometric detection at 610 nm ensures a clear color transition and precise equivalence point determination. Measurement drift and endpoint criteria are optimized for minimal noise and accurate titration curves.
Benefits and Practical Applications
- High precision and repeatability for cement quality control
- Automatable workflow with robotic sample handling
- Compliance with international standards (DIN EN 196-2)
- Rapid results suitable for routine laboratory and production monitoring
Future Trends and Potential Applications
Advances may include integration with inline process analytics for real-time monitoring, miniaturized sensors for on-site testing, and coupling with chemometric tools to broaden multi-element analysis in construction materials.
Conclusion
Photometric titration using EDTA and murexide at 610 nm provides a robust, standardized method for calcium determination in cement. The approach delivers reliable, traceable results and supports efficient laboratory workflows aligned with industry standards.
Reference
- Application Note T–79: Photometric Determination of Calcium in Cement by EDTA Titration (DIN EN 196-2)
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.