Oxygen and Nitrogen in Nitride Material

Applications | 2003 | LECOInstrumentation
Elemental Analysis
Industries
Materials Testing, Energy & Chemicals
Manufacturer
LECO

Importance of the Topic


Accurate determination of oxygen and nitrogen in nitride ceramics is critical for ensuring optimal mechanical, thermal and electrical properties. These elements significantly influence the performance and reliability of components used in automotive, aerospace, electronic and industrial applications.

Objectives and Overview


This application note presents a validated inert-gas fusion method for simultaneous quantification of oxygen and nitrogen in silicon nitride and related materials. It describes sample preparation, instrument calibration and key analysis parameters using a LECO TC500/TC600 analyzer.

Methodology and Instrumentation


Samples are weighed (20–50 mg) into folded tin capsules and placed in nickel baskets. A graphite crucible completes the assembly. The analyzer operates in automatic mode with a typical analysis delay of 20 s, outgas/furnace power set at 6200 W (outgas) and 5200 W (analysis), and minimum analysis times of 30 s for oxygen and 60 s for nitrogen. Calibration is performed with NIST 8983 silicon nitride standards. Key parameters include:
  • Outgas cycles: 3
  • Analysis delay comparator: 1.0000
  • Integration delays: 5 s (O) and 15 s (N)
  • Purge time: 15 s
  • Significant digits: 5

Used Instrumentation


  • LECO TC500/TC600 inert-gas fusion analyzer
  • Calibration standard: NIST 8983 silicon nitride
  • Accessories: 782-720 graphite crucible, 782-721 electrode tip, 501-059 tin capsules, 502-344 nickel baskets

Main Results and Discussion


Analysis of NIST 8983 yielded an average of 1.145 % O (±0.007 %) and 40.16 % N (±0.26 %) over multiple runs. Precision is within 0.14 % for oxygen and 1.06 % for nitrogen, demonstrating excellent reproducibility. The automated furnace control and blank correction procedures ensure low detection limits and reliable baseline stability.

Benefits and Practical Applications


This method offers rapid, fully automated analysis with minimal sample handling. It is ideally suited for routine QA/QC in ceramic manufacturing, research laboratories and process development. High precision and accuracy facilitate compliance with industry specifications and support material optimization.

Future Trends and Applications


Advances may include integration of real-time process monitoring, enhanced detection limits through improved sensor technology and coupling with chemometric software for predictive quality control. Miniaturized analyzers and remote operation capabilities could further extend applications in field-based testing.

Conclusion


The described inert-gas fusion technique on the LECO TC500/TC600 platform provides a robust, accurate and reproducible approach for simultaneous determination of oxygen and nitrogen in nitride materials, supporting stringent quality requirements in advanced ceramic applications.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Oxygen/Nitrogen Determination in Nitrides
Determination of Oxygen and Nitrogen in Reactive/Refractory Metals and Their Alloys*
Determination of Oxygen and Nitrogen in Reactive/Refractory Metals and Their Alloys* (TC600)
Oxygen and Nitrogen Analysis in Calcium Fluoride