LC/MS, LC/MS/MS, LC/HRMS, LC/Orbitrap, Software
IndustriesProteomics
ManufacturerThermo Fisher Scientific
Importance of the Topic
Proteomics analyses by LC–MS require high sensitivity, robustness and throughput. Tandem direct injection workflows combining two separation columns with a single electrospray emitter optimize mass spectrometer utilization approaching 100% of cycle time. This enhances robustness by avoiding emitter clogging, increases sample throughput, and maintains quantitative precision across capillary and micro-flow regimes.
Objectives and Study Overview
This study demonstrates the configuration of the Thermo Scientific Vanquish Neo UHPLC system in a tandem direct injection arrangement for capillary (150 µm I.D.) and micro-flow (300–1000 µm I.D.) separations. The aim is to achieve high-throughput bottom-up proteomics with near 100% MS utilization, low carryover, and reproducible quantification.
Methodology and Instrumentation
The workflow employs two columns connected via two low-dispersion switching valves and a single electrospray ionization emitter (Nanospray Flex, EASY-Spray or OptaMax NG) coupled to an Orbitrap Exploris 240 mass spectrometer. Fluidics use nanoViper Fingertight fittings (20 µm for capillary, 50 µm for micro-flow) and PepMap Neo columns (150 µm×150 mm, 300 µm×15 cm, 1000 µm×5 cm/15 cm). Gradient delivery is driven by two Binary Pump N modules—one for separation and one for reconditioning. Data acquisition utilizes data-independent acquisition (DIA) processed with CHIMERYS DIA or Spectronaut 19 software at FDR < 1%.
Main Results and Discussion
- Capillary flow (1–5 µL/min) methods delivered 48–225 samples/day (SPD) with ≥ 90% MS utilization. A 150 µm×150 mm column at 1.5–3.8 µL/min identified 3 111–4 281 protein groups from 50–1000 ng HeLa digest, achieving < 20% CV for over 3500 proteins at 200 ng.
- Micro-flow (15 µL/min, 300 µm×15 cm) achieved 240 SPD and ~ 2 200 proteins at 200 ng with 90% MS utilization; injection of 2000 ng increased identifications to > 3400.
- Milliflow (100 µL/min, 1000 µm×5 cm) reached 277 SPD with 95% MS utilization, yielding ~ 2400 proteins from 1 µg digest.
- Retention time variation across eight columns was < 4%, protein abundance CV < 6%, and carryover remained low (< 0.03% at 180 SPD).
- An inline 75 µm×2 cm trap column extended column lifetime with minimal impact on proteome depth.
Benefits and Practical Applications of the Method
- Easy configuration with standardized fluidic connections.
- Intelligent method creation and valve timing ensure robust, reproducible operation.
- Single-emitter design extends emitter life and eliminates inter-emitter variability.
- High MS utilization maximizes instrument efficiency for high-sensitivity proteomics.
- Low carryover and robust reproducibility suit large-scale studies, QA/QC and industrial analytics.
Future Trends and Potential Applications
- Integration of trap-and-elute schemes for online desalting and further throughput gains.
- Miniaturization of flow regimes for ultra-high sensitivity analyses.
- Automated and AI-driven method optimization to tailor workflows to diverse sample types.
- Extension to other omics fields such as metabolomics and lipidomics.
Conclusion
The tandem direct injection workflow on the Vanquish Neo UHPLC system offers a versatile, high-throughput solution for capillary and micro-flow LC–MS proteomics. It achieves near 100% MS utilization, robust performance, low carryover, and excellent reproducibility, meeting the demands of modern proteomics and industrial applications.
Reference
- Zheng R, et al. Maximizing sample throughput and sensitivities in nano and capillary LC–MS by leveraging a tandem direct injection workflow and a double-barrel ion source. Thermo Fisher Scientific Technical Note 003335; 2024.
- Bian Y, et al. Robust, reproducible and quantitative analysis of thousands of proteomes by micro-flow LC–MS/MS. Nature Communications 2020;11:157.
- Thermo Fisher Scientific. Vanquish Neo System VN-S10 and modules user guide Revision 2.0; 2024.
- Staes A, et al. High-throughput nanoflow proteomics using a dual-column electrospray source. Analytical Chemistry 2024;96(17):6534–6539.
- Zheng R, et al. Fast, sensitive, and reproducible nano- and capillary-flow LC–MS methods for high-throughput proteome profiling using the Vanquish Neo UHPLC and Orbitrap Exploris 480 MS. Technical Note 000138; 2021.
- Meding S, et al. Ultra-robust micro-flow LC–MS/MS for targeted high-throughput peptide quantification using the Vanquish Neo UHPLC system. Technical Note 74161; 2021.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.