Detailed Separation of C1-C5 Light Hydrocarbons on CP-Al2O3/Na2SO4 PLOT GC Column

Applications | 2010 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies

Importance of the Topic



Detailed separation of C1–C5 light hydrocarbons is essential for quality control, safety and process optimization in refinery, natural gas and LPG streams. Accurate quantification of impurities at trace to percent levels supports regulatory compliance and efficient downstream processing.

Objectives and Study Overview



This application demonstrates the use of a CP-Al2O3/Na2SO4 PLOT GC column to resolve 18 C1–C5 hydrocarbons, including structural and positional isomers, within a 20-minute run. The goal is to showcase selectivity, peak shape and robustness for routine hydrocarbon profiling.

Methodology and Instrumentation



The analysis employed gas chromatography with flame ionization detection (GC-FID) using the following setup:
  • Column: CP-Al2O3/Na2SO4 PLOT, 50 m × 0.32 mm ID, film thickness 5 µm
  • Carrier gas: hydrogen at constant pressure (100 kPa)
  • Oven program: 70 °C hold, 3 °C/min to 170 °C
  • Injection: split 1:50, injector at 250 °C, 5 µL gas sample
  • Detector: FID at 275 °C

Main Results and Discussion



The CP-Al2O3/Na2SO4 column achieved baseline separation of saturated and unsaturated C1–C5 species in under 20 minutes. Peak shapes were generally sharp, with slight tailing observed for acetylene and propyne due to stronger interactions with the polar Na2SO4 deactivation layer. Molar concentrations ranged from 0.199 % to 24.9 %, demonstrating reliable quantitation across a wide dynamic range.

Benefits and Practical Applications



The polar Na2SO4 deactivation on alumina enhances retention of unsaturated and branched hydrocarbons, improving isomeric resolution over non-polar columns. The column tolerates bake-out to 200 °C without loss of performance, offering stability for high-throughput quality assurance in petrochemical plants, natural gas monitoring and LPG specification testing.

Future Trends and Opportunities



Advancements may include novel deactivation chemistries to reduce tailing of polar species, integration with mass spectrometry for definitive identification and development of shorter, high-efficiency PLOT columns for faster cycle times. Online GC-FID systems are expected to further automate process monitoring in real time.

Conclusion



The CP-Al2O3/Na2SO4 PLOT GC column provides a robust and selective solution for routine detailed analysis of C1–C5 hydrocarbons. Its high resolving power and thermal stability make it well suited for petrochemical and gas industry requirements.

Reference


  • Provoost L. Detailed Separation of C1–C5 Light Hydrocarbons on CP-Al2O3/Na2SO4 PLOT GC Column. Agilent Technologies Application Note SI-02644, 2010.
  • Gasterra. What Is Natural Gas? http://www.gasterra.com/naturalgas/Pages/Whatisnaturalgas.aspx
  • WiseGeek. What Is Refinery Gas? http://www.wisegeek.com/what-is-refinery-gas.htm
  • Australian Government Dept. of Environment. LPG Fuel Quality Standards FAQ. http://www.environment.gov.au/atmosphere/fuelquality/standards/lpg/faq.html#lpg

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Solutions for light hydrocarbons and gasses: PLOT columns
Solutions for light hydrocarbons and gasses: PLOT columns
2019|Agilent Technologies|Presentations
PLOT Columns Separation Solutions for  Light Hydrocarbons  & Gases
Parallel GC for Complete Refinery Gas Analysis
Parallel GC for Complete Refinery Gas Analysis
2007|Agilent Technologies|Applications
Refinery Gas Analysis Using the Agilent 990 Micro GC
Refinery Gas Analysis Using the Agilent 990 Micro GC
2019|Agilent Technologies|Applications