ICP/MS
IndustriesOther
ManufacturerThermo Fisher Scientific
Significance of the Topic
Collision/reaction cell (CRC) technology coupled with a pre-cell mass filter addresses challenging isobaric interferences in multicollector ICP-MS, especially for complex matrices such as laser ablation samples. By preselecting a narrow mass window, the Neoma MS/MS MC-ICP-MS significantly enhances control over reaction chemistry, leading to more reliable and interference-free isotope ratio measurements.
Objectives and Study Overview
This study introduces the Neoma MS/MS MC-ICP-MS, featuring a double Wien filter ahead of the hexapole CRC. Key objectives include demonstrating how pre-cell mass filtering improves reaction predictability, reduces unwanted molecular interferences, and expands the range of practical isotope applications, from solution analyses to in-situ laser ablation.
Methodology and Instrumentation Used
The instrument integrates three main stages:
- Filter: A double Wien filter with an adjustable slit disperses ions by m/z, allowing only a defined mass window (e.g., m/z 29 for silicon) to enter the CRC.
- React: A hexapole collision/reaction cell introduces gases such as O₂, SF₆, H₂, or He to shift or neutralize interferences.
- Separate: A magnetic sector and multicollector array measure isotope ratios under low- or high-resolution conditions.
This configuration maintains high ion transmission and predictable mass bias following the exponential fractionation law.
Main Results and Discussion
• Si in HNO₃: Pre-filtering around m/z 29 suppressed 28Si interferences, doubling Si sensitivity compared to unfiltered CRC.
• Titanium isotopes: Clean TiO⁺ spectra (m/z 62–66) achieved by pre-filtering and O₂ reaction gas, eliminating Ca, V, Cr, and Fe interferences.
• Rb-Sr in situ dating: A narrow 103–107 amu window removed matrix peaks (e.g., 103Rh, 107Ag), isolating SrF reaction products for interference-free 87Sr/87Rb ratios.
• Potassium isotopes: Using H₂/He to neutralize ArH, low-resolution K isotope analysis on 25 ppb samples achieved 0.03‰ external reproducibility, outperforming traditional high-resolution methods.
• Sensitivity and Abundance Sensitivity: Pre-filtering reduces space‐charge effects and unwanted scattering, yielding sensitivity equal or superior to prior MC-ICP-MS models and thermal ionization standards.
Benefits and Practical Applications
- Predictable CRC reactions for robust interference removal
- Improved precision in isotope ratio measurements across Fe, Sr, Nd, Hf, Pb, U, and others
- Extension to challenging in-situ and laser ablation–MC-ICP-MS applications
- Capability to measure traditionally high-resolution systems (e.g., K isotopes) at low resolution
- Enhanced sensitivity enables analysis of smaller sample sizes
Future Trends and Applications
Ongoing developments will likely focus on integrating MS/MS pre-filtering with advanced reaction gases and laser sources to expand in situ geochronology, environmental tracer studies, and nano-scale analyses. Further miniaturization and automation may enable real-time monitoring in industrial and field settings.
Conclusion
The Neoma MS/MS MC-ICP-MS with its double Wien pre-cell mass filter represents a transformative advance in CRC-MC-ICP-MS technology. By simplifying CRC chemistry and eliminating major interferences, it achieves higher sensitivity, superior abundance sensitivity, and broader application scope without sacrificing mass bias control.
References
- Moynier F, Hu Y, Wang K, Zhao Y, Gérard Y, Deng Z, et al. Potassium isotopic composition of various samples using a dual-path collision cell-capable multiple-collector inductively coupled plasma mass spectrometer, Nu instruments Sapphire. Chemical Geology. 2021;571:120144.
- Ku Y, Jacobsen SB. Potassium isotope anomalies in meteorites inherited from the protosolar molecular cloud. Science Advances. 2020;6(41).
- Schwieters J, Jung G. Double Wien Filter. EP 3769334B1. 2022.
- Schwieters J, Elliott TR, Coath CD. Multi detector mass spectrometer and spectrometry method filter. US 10867780B2. 2020.
- Bevan D, Coath CD, Lewis J, Schwieters J, Lloyd N, Craig G, et al. In situ Rb–Sr dating by collision cell, multicollection inductively-coupled plasma mass-spectrometry with pre-cell mass-filter, (CC-MC-ICPMS/MS). Journal of Analytical Atomic Spectrometry. 2021;36(5).
- Bevan D, Coath C, Lewis J, Schwieters J, Lloyd N, Craig G, et al. Detrital K-Feldspar Geochronology by Collision Cell MC-ICPMS/MS. Goldschmidt Abstracts. 2020.
- Pfeifer M, Lewis J, Coath C, Schwieters J, Elliott T. In situ Titanium Isotope Measurements in Meteorites Using the Collision Cell MC-ICPMS, Proteus. Goldschmidt Abstracts. 2019.
- Craig G, Wehrs H, Bevan DG, Lewis J, Coath C, Elliott T, et al. Project Vienna: A novel pre-cell mass filtre for a collision/reaction cell MC-ICPMS/MS. Analytical Chemistry. 2021;93(30).
- Télouk P, Albalat E, Tacail T, Arnaud-Godet F, Balter V. Steady analyses of potassium stable isotopes using a Thermo Scientific Neoma MC-ICP-MS. Journal of Analytical Atomic Spectrometry. 2022;6.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.