Uranium isotopic analysis for the nuclear industry

Applications | 2019 | Thermo Fisher ScientificInstrumentation
ICP/MS
Industries
Energy & Chemicals
Manufacturer
Thermo Fisher Scientific

Significance of the Topic


Precise determination of uranium isotope ratios is critical for nuclear safeguards, fuel cycle control and quality assurance within enrichment facilities.

Objectives and Study Overview


The study assesses the performance of the Thermo Scientific Neptune XT Multicollector ICP-MS for measuring low enriched uranium according to ASTM C1477-08. It aims to achieve relative uncertainties below 0.1% for 235U and evaluate throughput, stability and minor isotope measurement.

Used Instrumentation


  • Thermo Scientific Neptune XT MC-ICP-MS with central SEM and RPQ lens
  • Compact Discrete Dynode SEM for 234U detection
  • Faraday cups equipped with 10^11, 10^12 and 10^13 ohm amplifiers
  • Cetac ASX-112 FR autosampler
  • Wet plasma with standard cones and PFA spray chamber

Methodology


Certified reference materials (NBS U-045, U-030, U-010) were prepared at 100 ng/g U in 2% HNO3. Measurements follow an automated sequence of 30 samples over three days, including blank corrections and hydride adjustments. Mass bias and ion counter yield are corrected using a CRM (IRMM-187) at the start of each sequence.

Main Results and Discussion


Instrument stability over eight hours yielded 235U uncertainties better than 0.1%, while 234U and 236U were measured with precisions below 1%. Relative standard deviations across three sessions were 0.01–0.02% for 235U and 0.04–0.07% for 234U. Measured ratios closely match updated literature values, demonstrating low scatter and reliable bias correction.

Benefits and Practical Applications


  • High productivity: 30 samples and controls in eight hours
  • Fully automated data correction and sequence control
  • Fit-for-purpose precision for enrichment monitoring
  • Global service support for continuous operation

Future Trends and Applications


Integration with glovebox or fume hood containment and optional jet interface promise enhanced sensitivity for trace samples. Advances in amplifier technology and collector configurations will support lower sample amounts and broaden isotope coverage. Combined MC-ICP-MS and TIMS workflows may further improve minor isotope analysis for safeguards and fuel cycle studies.

Conclusion


The Neptune XT MC-ICP-MS delivers robust uranium isotope analysis in compliance with ASTM C1477-08, combining high precision, stability and throughput for nuclear industry applications.

References


  • Richter S., Goldberg S.A. Int. J. Mass Spectrom. 2003;229:181–197.
  • ASTM C1477-08. Standard Test Method for Isotopic Abundance Analysis of Uranium Hexafluoride and Uranyl Nitrate Solutions by MC-ICP-MS. ASTM International, 2009.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Uranium Isotopic Analysis for the Nuclear Industry Using the Thermo Scientific NEPTUNE Plus MC-ICP-MS
Highly-sensitive uranium isotopic analysis for nuclear safeguards
Increasing accuracy and precision of uranium isotope ratios by Thermal Ionization Mass Spectrometry using 1013 Ω amplifier technology
Thermo Scientific Triton XT TIMS
Thermo Scientific Triton XT TIMS
2018|Thermo Fisher Scientific|Brochures and specifications