Ensuring toy safety from hexavalent chromium to meet European regulations using IC-ICPMS quantification

Applications | 2021 | Thermo Fisher ScientificInstrumentation
IC-MS, ICP/MS, Speciation analysis
Industries
Materials Testing
Manufacturer
Thermo Fisher Scientific

Importance of the Topic


The presence of hexavalent chromium in toys poses serious health risks, especially to children who may ingest small components. European regulations (EN 71-3:2019) enforce strict migration limits for Cr(VI). Reliable detection at ultra-low levels is crucial to protect consumers and ensure regulatory compliance.

Objectives and Overview of the Study


This study aimed to refine an existing ion chromatography–ICP-MS (IC-ICPMS) workflow to quantify Cr(VI) migration from sticky or liquid toy materials (category II) and scraped-off materials (category III). Key goals were compliance with EN 71-3:2019, a quantification limit down to 0.0025 mg/kg in toys, and a robust migration test protocol.

Methodology and Instrumentation


Samples and migration extracts were prepared following EN 71-3 guidelines under alkaline conditions to stabilize Cr(VI). Separation was achieved on a dual-column setup (IonPac NG1 guard + AG7 analytical) using 75 mM HNO₃ at pH 7.0, with standards adjusted to pH 7.5. The IC effluent was directly introduced into a Thermo Scientific iCAP RQ ICP-MS via a PFA-LC nebulizer. The 52Cr isotope was monitored in helium collision mode (KED) to eliminate interferences. A multi-pump configuration and an AS-AP autosampler allowed rapid sequential injections.

Used Instrumentation

  • Thermo Scientific Dionex ICS-6000 ion chromatography system
  • Thermo Scientific iCAP RQ ICP-MS
  • Thermo Scientific Dionex AS-AP autosampler
  • IonPac NG1 and AG7 columns
  • PFA-LC Nebulizer

Main Results and Discussion


Calibration over 0.02–0.16 µg/L Cr(VI) exhibited excellent linearity (R² = 0.9998) and a background-equivalent concentration of 0.002 µg/L. The instrumental detection limit was 0.35 ng/L, corresponding to 0.0025 mg/kg in toys—twenty-times below the regulatory threshold. Migration testing of glue, paint, and plastic extracts showed no native Cr(VI). Spiked samples (0.01 µg/L) yielded recoveries of 98–109% with stable retention times (~50 s) and no carry-over. Analysis time was reduced to 2 min per run.

Benefits and Practical Application of the Method


  • Meets stringent regulatory limits for toy safety
  • High-throughput, rapid analysis for QA/QC labs
  • Robust separation of Cr species without interconversion
  • Minimal sample preparation and carry-over risk

Future Trends and Potential Applications


As regulations evolve, detection limits may need further reduction. Coupling with high-resolution mass analyzers or advanced chromatographic phases could enhance sensitivity. Automation of migration procedures and data processing will streamline lab workflows and support routine screening.

Conclusion


The optimized IC-ICPMS method delivers a fast, reliable, and ultra-sensitive approach for Cr(VI) quantification in toys, ensuring compliance with current EN 71-3:2019 standards and readiness for future regulatory demands.

Reference


  1. EN 71-3:2019 Safety of Toys – Part 3: Migration of certain elements
  2. Alvarez CC, Gómez MEB, Zavala AH. Hexavalent chromium: Regulation and health effects. Journal of Trace Elements in Medicine and Biology, 2021.
  3. Thermo Fisher Scientific Application Note 43175

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

The Migration of Elements from Toys and Speciation of Chromium (VI) in Toy Material Using a Low Cost IC-ICP-MS Solution
Speciation Applications Summary Ion Chromatography
Speciation Applications Summary Ion Chromatography
2017|Thermo Fisher Scientific|Guides
Determination of Trivalent and Hexavalent Chromium in Toy Materials
Determination of chromium species using ion chromatography coupled to inductively coupled plasma mass spectrometry