Consumables, HPLC, LC columns
IndustriesOther
ManufacturerShimadzu
Significance of the Topic
Mouthwash formulations often include multiple active antimicrobial and anti-inflammatory agents to maintain oral hygiene and health. Reliable quantification of these components is essential for quality control, regulatory compliance, and consumer safety. High-speed chromatographic methods reduce analysis time and resource consumption while maintaining accuracy and precision.Objectives and Study Overview
This study describes a rapid reversed-phase liquid chromatography method for simultaneous determination of dipotassium glycyrrhizate (GK2), cetylpyridinium chloride (CPC), and isopropyl methylphenol (IPMP) in mouthwash products. The goal is to achieve baseline separation within a short run time (<7 minutes) with robust performance suitable for routine quality control.Methodology
- Column: Shim-pack Arata C18, 75 mm × 3.0 mm i.d., 2.2 μm
- Mobile Phase: A) 20 mmol/L sodium phosphate buffer (pH 2.6), B) methanol; gradient from 60% B to 80% B (0–4.50 min) returning to 60% B (4.51–7.00 min)
- Flow Rate: 0.8 mL/min; Column Temperature: 45 °C
- Detection: UV at 250 nm (GK2), 258 nm (CPC), 280 nm (IPMP)
- Injection Volume: 5 μL; Mixer Volume: 180 μL
Used Instrumentation
- Shimadzu Nexera XR UHPLC system
- SPD-M40 UV detector with standard cell
- Shimadzu LabTotal™ glass vials (1.5 mL)
Main Results and Discussion
The developed method achieved clear separation of GK2, CPC, and IPMP within 7 minutes with resolution factors >1.5. Retention times were reproducible (RSD <1.0%) across multiple injections. Calibration curves for each analyte were linear over the tested concentration range (5–100 mg/L) with correlation coefficients (r2) exceeding 0.999. Method precision and accuracy met industry standards for quantitative analysis.Benefits and Practical Applications
- High sample throughput for routine quality control laboratories
- Reduced solvent consumption and analysis time
- Reliable performance for regulatory compliance in oral care product manufacturing
Future Trends and Possibilities
Advancements in UHPLC and sub-2 µm particle columns can further shorten run times and enhance resolution. Emerging detection technologies such as mass spectrometry enable multi-residue screening and structural confirmation. Adoption of green chemistry principles may lead to alternative solvents and buffer systems to minimize environmental impact.Conclusion
The presented high-speed LC method provides an efficient, accurate, and robust approach for simultaneous quantification of key active ingredients in mouthwash, supporting quality assurance and regulatory requirements in oral care product analysis.References
- Shimadzu Corporation. Application News an_01-00505 (JP, ENG), 2023.