Measurement of total nitrogen (TN) for protein estimation in pharmaceutical vaccines on the Shimadzu TOC-L with TNM-L

Applications | 2020 | ShimadzuInstrumentation
TOC
Industries
Pharma & Biopharma
Manufacturer
Shimadzu

Significance of the Topic


The accurate determination of total nitrogen content is essential for protein quantification in vaccine development and quality control. Nitrogen measurement underpins estimation of antigen concentration in biopharmaceutical products, ensuring consistent potency and safety across production stages.

Objectives and Study Overview


This study evaluated a high-temperature combustion method coupled with chemiluminescence detection for total nitrogen (TN) analysis using the Shimadzu TOC-L with TNM-L module. Conducted at the Serum Institute of India, it compared results against the conventional Kjeldahl method for tetanus, diphtheria and pertussis antigens.

Used Instrumentation

  • Shimadzu TOC-L analyzer equipped with TNM-L module
  • High-temperature combustion furnace operating at 720°C
  • Chemiluminescence detector for nitrogen monoxide
  • Standard solutions of potassium nitrate for calibration


Methodology


Samples containing organic and inorganic nitrogen compounds were combusted at 720°C to generate nitric oxide. The resulting gas was dehumidified, cooled and mixed with ozone in a chemiluminescence analyzer. Emitted light from excited NO2 species was measured photoelectrically, producing a signal proportional to nitrogen content. Calibration was performed using potassium nitrate standards, and recovery was tested with L-glutamic acid (1000 mg/L) and L-histidine (500 mg/L). Comparative analyses were carried out on three batches each of purified tetanus toxoid, diphtheria toxoid, pertussis toxin and FHA antigen.

Results and Discussion


A calibration curve for potassium nitrate demonstrated excellent linearity (r ≥ 0.9997). Amino acid recoveries fell within 100 ± 10 percent, matching values obtained by the Kjeldahl method. Total nitrogen estimates for vaccine antigens were comparable between the two techniques, confirming the accuracy and precision of the TOC/TN approach.

Benefits and Practical Applications


The TOC-L with TNM-L method offers significant advantages over the traditional Kjeldahl technique:
  • Faster analysis time and higher sample throughput
  • Elimination of corrosive concentrated acids and associated safety hazards
  • Reduced need for protein-specific correction factors
  • Compatibility with diverse pharmaceutical and biopharmaceutical matrices


Future Trends and Possibilities


Advancements may include integration of real-time process analytical technology for in-line monitoring of nitrogen content, miniaturized mobile analyzers for field use, and expanded application to complex biologics and environmental samples. Automation and data connectivity will further streamline quality control workflows.

Conclusion


The high-temperature combustion chemiluminescence method on the Shimadzu TOC-L with TNM-L provides a reliable, efficient and safer alternative to the Kjeldahl method for total nitrogen determination in vaccines. Its strong correlation with established methods and operational benefits support its adoption in pharmaceutical quality assurance.

References

  • ASTM D8-83-16: A New Method for Total Nitrogen (TN) & Total Kjeldahl Nitrogen (TKN)

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Quality Control of Vinegar by TN Measurement
Cleaning Validation of Food Production Equipment by TOC and TN Measurement
Solutions for Vaccine Testing
Solutions for Vaccine Testing
2020|Shimadzu|Brochures and specifications
Solutions for Vaccine Testing
Solutions for Vaccine Testing
2020|Shimadzu|Guides