LC/HRMS, LC/MS, LC/MS/MS, LC/Orbitrap
IndustriesProteomics
ManufacturerThermo Fisher Scientific
Importance of the Topic
Monoclonal antibodies represent a dominant class of biotherapeutics with significant therapeutic impact and commercial importance. Fast and reliable peptide mapping is critical for confirming amino acid sequence, identifying post-translational modifications (PTMs) and ensuring consistent product quality throughout development and manufacturing. High-throughput methods that maintain comprehensive sequence coverage and PTM characterization enable rapid process optimization and support quality control.
Objectives and Overview
This study demonstrates the capabilities of the Thermo Scientific™ Orbitrap Exploris™ 480 mass spectrometer coupled with a Vanquish™ UHPLC system for rapid, high-resolution peptide mapping of the therapeutic antibody golimumab. By employing shortened chromatographic gradients (5–40 minutes) and automated sample preparation, the work aims to maintain 100% sequence coverage, detect and quantify key PTMs, and illustrate the method’s applicability to process development and high-throughput screening.
Methodology
Golimumab samples were subjected to forced oxidation (500 ppm H2O2, 24 h) or analyzed in their native state. Automated digestion was performed using SMART Digest™ magnetic resin and a Thermo Scientific™ KingFisher™ Duo Prime system controlled by BindIt™ software. Peptides were separated on an Acclaim™ VANQUISH™ C18 column (2.1 × 250 mm) using the Vanquish Horizon UHPLC platform under gradients from 5 to 40 min (2–40% acetonitrile in 0.1% formic acid). Data were acquired on the Orbitrap Exploris 480 (MS1 resolution 60 000, data-dependent MS2 Top15 at 15 000) and processed using Chromeleon™ 7.2 CDS and BioPharma Finder™ 3.2 software.
Instrumentation Used
- Vanquish Horizon UHPLC system with Binary Pump H, Sampler HT and Column Compartment H
- Acclaim VANQUISH C18, 2.2 µm, 2.1 × 250 mm column
- Orbitrap Exploris 480 mass spectrometer with BioPharma option
- SMART Digest magnetic trypsin kit and KingFisher™ 96 Deepwell plate
- Chromeleon™ 7.2.10 CDS and BioPharma Finder™ 3.2 software
Main Results and Discussion
Peptide mapping across all gradients achieved 100% sequence coverage for both heavy and light chains, even with a 5-minute separation. Key PTMs were consistently identified and quantified:
- Oxidation at methionine sites (M261 ~100%, M437 ~50%) and lower-abundance sites (M51, M113)
- Deamidation and succinimide formation at N43 (~60% Asn→Asp, ~10% Asu) and low-abundance deamidation at N370
- N-terminal pyroglutamate conversion of HC Q1 (~100%) and absent at LC E1
- C-terminal lysine clipping (~15–20%) evidenced by distinct charge states and retention times
- N-linked glycosylation profile at N306 featuring G0F (~38–43%), G1F (~34–39%), G2F (~9–11%) and minor high-mannose species
Benefits and Practical Applications of the Method
- Shortened analysis times increase throughput for process support and clone or media screening
- Automated digestion and data acquisition ensure reproducibility and reduce hands-on time
- High-resolution, fast scan performance preserves data quality in rapid UHPLC separations
- Comprehensive PTM profiling supports critical quality attribute monitoring during development
Future Trends and Potential Applications
Further enhancements may include:
- Integration of multiplexed or multiplex-augmentation strategies to analyze several samples simultaneously
- Advanced AI-driven data analysis for accelerated PTM discovery and trend prediction
- Expanded workflows for deeper glycan characterization using complementary hydrophilic separations
- Real-time process analytical technology (PAT) applications for in-line monitoring during bioreactor runs
Conclusion
The combination of the Orbitrap Exploris 480 mass spectrometer and Vanquish UHPLC system delivers rapid, high-throughput peptide mapping without compromising sequence coverage or PTM characterization. This robust workflow, supported by automated digestion and compliance-ready software, is well suited for biopharmaceutical process development, quality control, and biosimilar comparability studies.
References
- Evaluate Pharma World Preview 2018, Outlook to 2024.
- Wang W. et al., Mol. Immunol., 2011, 48, 806–866.
- Guan Z. et al., JASMS, 2003, 14, 605–613.
- Haberger M. et al., mAbs, 2014, 6(2), 327–339.
- Goetze A. M. et al., Glycobiology, 2011, 21(7), 949–959.
- Hossler P. et al., Glycobiology, 2009, 19(9), 936–949.
- Huang K. F. et al., PNAS, 2005, 102(37), 13117–13122.
- Dick L. W. Jr. et al., Biotechnol. Bioeng., 2007, 97(3), 544–553.
- Liu Z. et al., J. Pharm. Sci., 2019, 108(10), 3194–3200.
- Beyer B. et al., Biotechnol. J., 2018, 13, 1700476.
- Brorson K. and Jia A. Y., Curr. Opin. Biotechnol., 2014, 29, 140–146.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.