Increased thermal stability and maximum temperature of the Agilent J&W DB-HeavyWAX column

Applications | 2018 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
Agilent Technologies

Significance of the Topic


Gas chromatography employing 100% polyethylene glycol (PEG) columns is critical for the sensitive, reproducible analysis of industrial chemicals, flavors, fragrances, and BTEX compounds. Extending the operating temperature enhances the ability to separate higher-boiling analytes without compromising phase stability or reproducibility.

Objectives and Study Overview


This application note compares the thermal stability and retention time reproducibility of the Agilent J&W DB-HeavyWAX column against conventional WAX phases. Using a BTEX mixture, the study assesses performance at final oven temperatures up to 280 °C over extended operation periods.

Methodology and Instrumentation


A BTEX standard (100 ppm each of benzene, toluene, ethylbenzene, o-xylene, p-xylene, m-xylene in methanol) was analyzed on three 30 m × 0.25 mm × 0.25 µm PEG columns: two commercial WAX phases and the DB-HeavyWAX. Chromatographic conditions included a split-mode inlet at 250 °C, helium carrier flow at 1 mL/min, and an oven program from 70 °C (10 min) ramped to 280 °C at 20 °C/min with a 60-min hold. Detection was performed by FID at 280 °C.

Main Results and Discussion


Conventional WAX columns, operated above their MAOT (250/260 °C), exhibited significant retention time drift when held at 280 °C, with shifts increasing from slight to pronounced after 50 hours. In contrast, the DB-HeavyWAX column maintained stable retention for all BTEX peaks even after 100 hours at 280 °C, showing only minimal shifts of a few seconds.

Benefits and Practical Applications


The extended temperature limit and enhanced thermal stability of the DB-HeavyWAX column allow reliable high-temperature separations of heavy industrial analytes without phase degradation. This translates into improved quantitation precision, longer column lifetimes, and the ability to develop faster or higher-temperature methods for QA/QC, environmental and petrochemical analysis.

Future Trends and Applications


Advances in stationary phase chemistry will continue to push temperature and chemical resistance, enabling simultaneous analysis of volatile and semi-volatile compounds in a single run. Integration with fast GC and high-throughput autosamplers may further boost productivity in industrial and environmental laboratories. Emerging applications in complex matrices, such as polymers and biofuels, will benefit from robust, high-temperature PEG phases.

Conclusion


The Agilent J&W DB-HeavyWAX column delivers a reliable increase in maximum operating temperature to 280/290 °C while preserving retention time stability over extended use. This performance surpasses traditional WAX phases and broadens the analytical scope for challenging industrial and environmental applications.

Reference


  • ASTM D2306. Standard Test Method for C8 Aromatic Hydrocarbon Analysis by Gas Chromatography.
  • ASTM D6563. Standard Test Method for Benzene, Toluene, Xylene (BTX) Concentrates Analysis by Gas Chromatography.
  • Reese A.; Vickers A.; George C. GC Column Bleed: A MASS PerSPECtive. Agilent Technologies, publication B-0442, 2001.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Now There’s A WAX Column That Can Take The Heat
Analysis of Oil and Fat containing foods by fully automated sample preparation using a PAL3 coupled with a 7890 GC and a 5977 MSD system according to AOAC 996.01
New High Temperature PEG GC Column with Increased Temperature Limit and Ultra Low Bleed Level
WAX Column Technology Update
WAX Column Technology Update
2018|Agilent Technologies|Presentations