Consumables, HPLC, LC columns
IndustriesPharma & Biopharma
ManufacturerThermo Fisher Scientific
Significance of the Topic
The rapid and reliable analysis of statin drugs is critical in pharmaceutical quality control, clinical laboratories, and drug screening applications. High-throughput methods enable better process efficiency, cost reduction, and faster decision making in routine assays involving multiple analytes.
Objectives and Study Overview
This work demonstrates a fast screening method for nine statins using the Thermo Scientific Hypersil GOLD VANQUISH 1.9 µm UHPLC column coupled with the Vanquish UHPLC system. The study explores non-linear gradient capabilities and evaluates the impact of increased flow rates and backpressures on separation performance and assay throughput.
Methodology and Instrumentation
Sample Preparation:
- Primary solutions of nine statins at 1 mg/mL in methanol
- Working solutions at 100 µg/mL by dilution in water
- Vials labeled via Virtuoso Vial Identification System
UHPLC Conditions:
- Column: Hypersil GOLD VANQUISH, 1.9 µm, 100 × 2.1 mm
- Mobile Phase A: 0.1% formic acid in water
- Mobile Phase B: 0.1% formic acid in acetonitrile
- Column temperature: 40 °C with eluent pre-heating
- Injection volume: 1 µL; Detection wavelength: 240 nm
- Flow rates evaluated: 0.5, 1.0, 1.2, and 1.4 mL/min
- Gradient profiles: linear versus non-linear curves (curve value = 8)
Instrumentation:
- Vanquish UHPLC System: System Base, Binary Pump H, Split Sampler HT, Column Compartment H, Active Pre-heater, Diode Array Detector HL, LightPipe flow cell
- Chromatography Data System: Thermo Scientific Dionex Chromeleon 7.2 SR2
Main Results and Discussion
Non-linear gradient at 0.5 mL/min improved resolution of critical fluvastatin and atorvastatin peaks compared to a linear gradient. Increasing flow rates up to 1.4 mL/min while maintaining gradient slope kept peak resolution within acceptable USP criteria, despite backpressures rising to 1,334 bar.
Resolution Performance (1.0–1.4 mL/min):
- Amlodipine: 2.28–2.11
- Ezetimibe: 24.31–20.98
- Fluvastatin: 10.87–9.87
- Atorvastatin: 1.69–1.41
- Lovastatin: 7.34–5.32
- Simvastatin: 6.77–4.68
Precision: Twelve replicate injections at 1.2 mL/min non-linear method yielded retention time RSD ≤ 0.20% for all analytes, underscoring system stability and reproducibility.
Benefits and Practical Applications
- Method time under three minutes at maximum flow rates
- Baseline resolution of critical statin pairs
- High retention time precision (RSD ≤ 0.2%)
- Increased sample throughput through non-linear gradient optimization
Future Trends and Potential Applications
Advances in UHPLC columns and high-pressure systems will further accelerate multi-compound screening. Emerging areas include:
- Automation and AI-driven method development for gradient optimization
- Ultra-high pressure (> 1500 bar) workflows for faster analyses
- Integration with mass spectrometry for enhanced sensitivity and selectivity
- Miniaturization and multiplexing for clinical and bioanalytical platforms
Conclusion
The combination of the Hypersil GOLD VANQUISH UHPLC column with the Vanquish UHPLC system and non-linear gradient capabilities enables rapid, high-resolution screening of statins. This approach supports sub-three-minute assays with excellent reproducibility and significant throughput gains.
References
- Hillbeck D. Application Note AN21496. Thermo Fisher Scientific; 2016.