Quantitative Analysis of Waste Oil by EDX-7000

Applications | 2013 | ShimadzuInstrumentation
X-ray
Industries
Energy & Chemicals
Manufacturer
Shimadzu

Importance of the Topic


The increasing environmental and operational demands on machinery drive the need for rapid, non-destructive elemental analysis of waste oils. Energy-dispersive X-ray fluorescence (EDX) provides a minimal-prep approach to quantify wear metals and contaminants, supporting preventive maintenance, product quality assurance, and regulatory compliance in industrial settings.

Objectives and Study Overview


This work assesses the performance of the Shimadzu EDX-7000 spectrometer for quantifying trace metals in hydrocarbon oil. Key goals include determining detection limits, calibration accuracy, and analytical precision. Results are benchmarked against a previous-generation instrument to demonstrate improvements in sensitivity and measurement throughput.

Methodology


  • Sample Preparation: Hydrocarbon oil samples spiked with Ti, V, Cr, Ni, Cu, Zn, Ag, Cd, Sn, Sb, Ba, and Pb at concentrations from 10 to 500 ppm. Eight milliliters of each sample were placed in polypropylene-film-sealed cups without further treatment.
  • Data Acquisition: Fluorescence spectra collected at integration times of 100 s and 300 s per element. Theoretical lower limits of detection (L.L.D.) calculated using net and background counts with a 3σ criterion.
  • Calibration: Working curve approach with internal standard correction for scattered radiation ensured linear responses across the concentration range.
  • Repeatability Testing: Ten consecutive measurements at 300 ppm for all elements, recording standard deviation and coefficient of variation.


Instrumentation Used


  • Spectrometer Model: Shimadzu EDX-7000 with Rh-target X-ray tube and silicon drift detector (SDD).
  • Operating Conditions: Tube voltages of 15 kV and 50 kV; automatic current control up to 30 μA; 10 mm collimator; primary filters #1, #2, #4; air atmosphere.


Main Results and Discussion


  • Detection Limits: Achieved L.L.D. between 0.3 ppm (Zn at 300 s) and 9.9 ppm (Ba at 300 s), representing a 1.5–4× sensitivity improvement over the prior model.
  • Calibration Accuracy: One-sigma uncertainties ranged from 1.0 ppm (V) to 3.3 ppm (Cr), confirming excellent linearity with internal standard corrections.
  • Precision: Coefficients of variation below 1% for most elements; Ba exhibited a higher 3% variation due to absorption effects and lower signal intensity.
  • Spectral Quality: High-resolution peak separation with overlap correction for coexisting elements ensured reliable quantification.


Benefits and Practical Applications


  • Fast, non-destructive analysis with minimal sample handling.
  • Sub-ppm sensitivity facilitates early detection of wear metals for machinery condition monitoring.
  • High precision supports lubricant production quality control and routine maintenance diagnostics.


Future Trends and Potentials


  • Integration with automated sampling and online monitoring platforms.
  • Enhanced detection of light elements through advanced detector technologies.
  • Application of chemometric algorithms to interpret complex oil degradation signatures.
  • Development of portable EDX systems for in-field diagnostic capabilities.


Conclusion


The Shimadzu EDX-7000 delivers significant improvements in sensitivity, speed, and precision for quantitative analysis of trace metals in waste oils. Its ease of use and robust analytical performance make it an essential tool for industrial maintenance, environmental monitoring, and lubricant quality assurance.

Reference


  • Shimadzu Application News No. X242

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Screening Analysis with EDX-7000 Navi Software
Multi-element Analysis of Used Lubricant Oils
Multi-element Analysis of Used Lubricant Oils
2019|Agilent Technologies|Applications
Analysis of Aqueous Solution by EDX-LE - Performance in Air Atmosphere
Improved Accuracy in the Measurement of Wear Metals and Additives in Lubricant Oils by ICP-OES