Significance of the topic


Oxygen heterocyclic compounds (OHCs) — principally coumarins, furocoumarins and polymethoxyflavones — occur mainly in the non-volatile fraction of cold‑pressed Citrus essential oils. They have notable bioactive properties but may present safety concerns (phototoxicity, systemic toxicity) when present at elevated concentrations in foods, fragrances or cosmetics. Sensitive, selective and fast methods for their routine monitoring are therefore essential for product quality control, regulatory compliance and safety assessment, while greener analytical approaches reduce environmental impact and operational costs.


Objectives and study overview


The study aimed to develop and validate a rapid, low‑solvent method for the separation and quantification of 28 OHCs in cold‑pressed Citrus essential oils using supercritical fluid chromatography coupled with triple quadrupole tandem mass spectrometry (SFC‑QqQ‑MS). Key goals included column screening and chromatographic optimization, construction of matrix‑matched calibration curves, method validation according to EURACHEM guidance, and application to real Citrus oil samples to produce quantitative OHC profiles.


Methodology


  • Sample set: five cold‑pressed Citrus essential oils (lemon, bergamot, sweet orange, bitter orange, mandarin); standards for 28 OHCs were procured and stored at −18 °C.
  • Sample preparation: oils were analyzed without extraction; dilutions in ethanol at 1:1,000 and 1:20,000 (v/v) were prepared and injected in triplicate.
  • Chromatography: a pentafluorophenyl (F5) core‑shell column (150 × 2.1 mm, 2.7 μm) was selected after screening eight stationary phases. Mobile phase: CO2 (solvent A) with methanol modifier (solvent B) using a linear gradient from 2% to 10% MeOH over 10 min. Flow rate 1.0 mL·min−1, injection volume 2 μL, oven 40 °C, backpressure regulator (BPR) 120 bar. Separation of the 28 targets achieved in under 8 minutes.
  • Mass spectrometry: Shimadzu LCMS‑8050 triple quadrupole with APCI source; full SFC flow directed to MS; MRM acquisition for selective quali‑quantitative detection. Typical MS settings: interface 350 °C, DL 250 °C, heat block 200 °C, N2 nebulizing 3 L·min−1, drying 5 L·min−1; event time 0.024 s per MRM transition.
  • Calibration and validation: matrix‑matched calibration on distilled lemon oil with 10 concentration levels spanning 0.001–3 mg·L−1 (and extended ranges per compound). Linearity, LoD/LoQ (Eurachem approach), accuracy, intra‑ and inter‑day precision were assessed. Regression intercepts were tested and curves forced through zero when appropriate.

Used instrumentation


  • SFC system: Shimadzu Nexera‑UC SFC platform (CO2 pump, dual solvent pumps, backpressure regulator, column oven, autosampler).
  • Detector: Shimadzu LCMS‑8050 triple quadrupole mass spectrometer with APCI source operating in MRM mode.
  • Column: Pentafluorophenyl (F5) core‑shell column, 150 × 2.1 mm, 2.7 μm.
  • Solvents and gases: CO2 (4.8 grade), MeOH and 2‑propanol (LC–MS grade), ethanol (HPLC grade) for dilutions.

Main results and discussion


Column screening identified the pentafluorophenyl (F5) stationary phase as the best performer, enabling baseline or near‑baseline separation of the 28 OHCs within 8 minutes while using <10% MeOH. MRM detection provided the selectivity needed to overcome partial coelutions observed for some isobaric or closely eluting pairs.


  • Validation metrics: coefficients of determination R2 ranged from 0.9952 to 0.9996. Limits of quantification (LoQs) ranged between approximately 0.0015 and 0.1536 mg·L−1 depending on compound and ionization efficiency. Intra‑ and inter‑day peak area CVs were ≤8% for all analytes. Accuracy (expressed as R’%) fell between 80.0% and 118.6% (within ±20% acceptance criteria).
  • Critical pairs and selectivity: Although certain pairs exhibited close retention times (e.g., citropten/isopimpinellin; bergamottin/5‑geranyloxy‑7‑methoxycoumarin), the MRM transitions enabled unambiguous quantification. Some coelutions were observed (e.g., cnidicin/phellopterin; cnidilin/isopimpinellin; epoxybergamottin/byakangelicin) but did not prevent selective measurement by MRM. Several isomeric couples were well separated by SFC.
  • Ecological/operational advantages: SFC achieved shorter runtime and lower organic solvent consumption versus a comparable reversed‑phase LC method. Estimated solvent use per analysis was ~10.7 mL MeOH for the SFC workflow (including make‑up solvent) versus ~16 mL (MeOH plus THF) for the LC method — a meaningful reduction for high‑throughput QC laboratories.
  • Application to real samples: quantitative profiles were coherent with literature patterns. Bergamot and bitter orange oils contained representatives of all OHC classes; lemon oil was dominated by coumarins and furocoumarins; mandarin and sweet orange oils were characterized primarily by polymethoxyflavones (PMFs). Representative abundant analytes included bergamottin, 5‑geranyloxy‑7‑methoxycoumarin and citropten in lemon/bergamot, and tangeretin as the main PMF in mandarin.

Benefits and practical applications


  • High selectivity and sensitivity for 28 target OHCs with LoQs suitable for trace analysis in finished products (cosmetics, foods, fragrances).
  • Fast throughput (sub‑8‑min separations) enabling routine QC and high sample throughput.
  • Reduced organic solvent consumption and lower environmental footprint compared to conventional LC methods — aligns with green‑chemistry goals.
  • MRM‑based MS/MS readout mitigates issues from partial chromatographic coelution and supports robust quantification in complex matrices.

Future trends and potential uses


  • Integration with on‑line supercritical fluid extraction (SFE) to create fully automated SFE‑SFC‑QqQ‑MS workflows for direct extraction and analysis of OHCs — promising further time savings, elimination of manual sample prep, and reduced solvent use.
  • Extension of the validated panel to additional OHCs as standards become available, and adaptation to finished cosmetic and food matrices to support regulatory monitoring (IFRA, EFSA, EC cosmetic regulation).
  • Broader adoption of SFC‑MS/MS for other non‑volatile or semi‑polar natural product classes where speed and green credentials are advantageous.
  • Implementation in routine QC laboratories for batch release testing and stability studies where fast, sensitive, and selective OHC profiling is required.

Conclusions


The validated SFC‑QqQ‑MS method delivers a rapid, sensitive and greener analytical approach for simultaneous analysis of 28 oxygen heterocyclic compounds in cold‑pressed Citrus essential oils. The method meets accepted validation criteria (linearity, precision, accuracy, and low LoQs), shortens analysis time (<8 min), reduces solvent consumption relative to HPLC workflows, and provides the selectivity required to resolve problematic pairs by targeted MRM. The approach is well suited for routine quality control and has clear potential for further automation (on‑line SFE‑SFC) and extension to finished product testing.


Reference


  • Russo M, Rigano F, Arigò A, Dugo P, Mondello L. Coumarins, psoralens and polymethoxyflavones in cold‑pressed Citrus essential oils: a review. J Essent Oil Res. 2021.
  • Hung WL, Suh JH, Wang Y. Chemistry and health effects of furanocoumarins in grapefruit. J Food Drug Anal. 2017.
  • Scott BR, Pathak MA, Mohn GR. Molecular and genetic basis of furocoumarin reactions. Mutat Res Genet Toxicol. 1976.
  • Serrano‑Pérez JJ, Olaso‑Gonzáles G, Merchan M, Serrano‑Andrés L. Singlet oxygen generation in PUVA therapy studied using electronic structure calculations. Chem Phys. 2009.
  • European Parliament. Regulation (EC) No. 1223/2009 on cosmetic products. 2009.
  • International Fragrance Association (IFRA). IFRA Standard. 48th Amendment Citrus oils and other furo‑coumarins containing essential oils. 2015.
  • International Fragrance Association (IFRA). IFRA analytical method — quantitative determination of furocoumarins by HPLC‑DAD. 2013.
  • Macmaster AP et al. Quantification of selected furocoumarins by high‑performance liquid chromatography and UV‑detection. J Chromatogr A. 2012.
  • Zoccali M, Donato P, Mondello L. Recent advances in the coupling of carbon dioxide‑based extraction and separation techniques. Trends Anal Chem. 2019.
  • Dugo P et al. Rapid analysis of polymethoxylated flavones from Citrus oils by supercritical fluid chromatography. J Agric Food Chem. 1996.
  • Desmortreux C et al. Improved separation of furocoumarins of essential oils by supercritical fluid chromatography. J Chromatogr A. 2009.
  • Ellison SLR, Barwick VJ, Duguid TJ. Practical statistics for the analytical scientist: a bench guide. 2nd ed. RSC; 2009.
  • Magnusson B, Örnemark U (eds). The fitness for purpose of analytical methods: a laboratory guide to method validation and related topics. 2nd ed. 2014.
  • Arigò A et al. Oxygen heterocyclic compound screening in Citrus essential oils by linear retention index approach applied to liquid chromatography coupled to photodiode array detector. Flavour Fragr J. 2019.
  • Russo M et al. Reduced time HPLC analyses for fast quality control of Citrus essential oils. J Essent Oil Res. 2015.
  • Dugo P et al. Comprehensive two‑dimensional normal‑phase (adsorption)‑reversed‑phase liquid chromatography. Anal Chem. 2004.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Oxygen heterocyclic compounds profile evaluation in cold-pressed Citrus essential oils using supercritical fluid chromatography coupled to photodiode array detector
Qualitative and Quantitative Determination of Furanocoumarins in Citrus Oils by LC/TOF
Simultaneous Quantification of 16 Furanocoumarins in Essential Oils Using a Robust and Sensitive Triple Quad LC-MS/MS Method
Profiling of Citrus Oils and Determination of Furocoumarins in Citrus Oils Using the Agilent 1290 Infinity 2D-LC Solution