Determination of N-Methylpyrrolidine in Cefepime Using a Reagent-Free Ion Chromatography System

Applications | 2016 | Thermo Fisher ScientificInstrumentation
Ion chromatography
Industries
Pharma & Biopharma
Manufacturer
Thermo Fisher Scientific

Significance of the topic


Cephalosporins are widely used for treating bacterial infections; cefepime, a fourth-generation cephalosporin, degrades at ambient temperatures to produce N-methylpyrrolidine (NMP), which is potentially toxic. Monitoring NMP levels is essential for ensuring drug purity and patient safety.

Objectives and study overview


Develop a rapid, sensitive, and simplified ion chromatography method for quantifying NMP in cefepime formulations using reagent-free methanesulfonic acid eluent and suppressed conductivity detection. Compare performance to current USP compendial methods and assess applicability to raw material and finished product samples.

Methodology and instrumentation


  • Instrumentation: Dionex ICS-3000 RFIC-EG system with SP/DP pump, EGC II MSA cartridge, CR-CTC II trap column, CSRS 300 detector, autosampler, Chromeleon software
  • Column: IonPac CG17 guard (2×50 mm) and CS17 analytical (2×250 mm)
  • Eluent: Electrolytically generated MSA gradient (6 mM for 0–7.5 min, 85 mM for 7.5–20 min, 6 mM for 20–30 min)
  • Flow rate: 0.4 mL/min; Column temp: 40–50 °C; Detector temp: 30 °C
  • Detection: Suppressed conductivity, recycle mode, 100 mA
  • Samples: Cefepime hydrochloride, simulated injection (with arginine), USP system suitability reference material

Main results and discussion


  • NMP elutes at ~5.3 min; total run time is 35 min for cefepime hydrochloride, reducing analysis from USP’s 55 min to 35 min
  • Linearity: r2=0.9999 across 0.45–200 µg/mL; LOQ 0.10 µg/mL (0.001%); LOD 0.03 µg/mL
  • Precision: Peak area RSD <1.3%, retention time RSD <0.8% intra-day and <0.5% inter-day
  • Accuracy: Spike recoveries between 97.8% and 102.0%
  • Method extends to cefepime for injection (with 7.25 mg/mL arginine) and system suitability sample, achieving baseline separation (RS>1.3) and runtimes around 45 min
  • Stability study: At 25 °C, NMP in cefepime increases from 0.23% to ~2% over 3 days; refrigerated and frozen storage maintain stability over 96 hours

Benefits and practical applications


  • Faster throughput and lower solvent consumption compared to USP methods
  • Reagent-free eluent generation simplifies preparation and avoids pH-induced degradation
  • High sensitivity allows quantitation well below pharmacopeial limits
  • Suitable for QC of raw materials, formulated products, and stability testing

Future trends and possibilities for application


The approach may be adapted to other β-lactam degradation products and combined with mass spectrometry for comprehensive impurity profiling. Miniaturized columns and automated platforms could further increase efficiency and reduce waste.

Conclusion


The developed RFIC method with IonPac CS17 and suppressed conductivity detection offers a rapid, accurate, and sensitive quantification of NMP in cefepime products, meeting and exceeding USP requirements while improving lab productivity.

Reference


  1. Sader HS, Jones RN. Historical Overview of the Cephalosporin Spectrum: Four Generations of Structural Evolution. Antimicrob. Newsl. 1992;8:75-84.
  2. Yahav D, Paul M, Fraser A, Sarid N, Leibovici L. Efficacy and Safety of Cefepime: A Systematic Review and Meta-Analysis. Lancet Infect Dis. 2007;7:338-348.
  3. Sader HS, Fritsche TR, Jones RN. Potency and Spectrum Trends for Cefepime Tested Against 65746 Clinical Isolates. Diagn Microbiol Infect Dis. 2005;52:265-273.
  4. FDA. Maxipime (Cefepime Hydrochloride) for Injection. MedWatch 2003.
  5. Sprauten PF, Beringer PM, Gill MA, Louie S, Synold TW. Stability and Antibacterial Activity of Cefepime During Continuous Infusion. Antimicrob Agents Chemother. 2003;47:1991-1994.
  6. USP 31–NF 26. Cefepime Hydrochloride and Cefepime for Injection Monographs. 2008.
  7. Dionex Application Note 157: Comparison of Suppressed to Nonsuppressed Conductivity Detection. 2004.
  8. Chen G, Liu G, Qin F, Wang Y. A Simple and Sensitive GC Method for Determination of N-Methylpyrrolidine in Cefepime. J Pharm Biomed Anal. 2003;33:797-801.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Determination of N-Methylpyrrolidine in Cefepime with Nonsuppressed Conductivity Detection
Determination of Dimethylamine in Metformin HCl Drug Product Using IC with Suppressed Conductivity Detection
Determination of Methylamine in Drug Products
Determination of Methylamine in Drug Products
2016|Thermo Fisher Scientific|Applications
Application Note 205 Determination of Cefepime and Cefepime- Related Substances Using HPLC with UV Detection