Analysis of Tequila Carbohydrates

Applications | 2011 | Agilent TechnologiesInstrumentation
HPLC
Industries
Food & Agriculture
Manufacturer
Agilent Technologies

Significance of the topic


Tequila is a popular spirit whose carbohydrate profile reflects both raw material quality and production parameters. Monitoring sugars such as raffinose, sucrose, lactose, glucose, fructose and sugar alcohols supports quality control, ingredient authenticity and process optimization in the beverage industry.

Objectives and overview


This application note demonstrates a straightforward chromatographic approach to separate and quantify major sugars and the trisaccharide raffinose in tequila. Using a single-column, ligand-exchange system with refractive index detection, the study aims to achieve baseline resolution of all analytes in a 30-minute run time.

Methods and instrumentation


The chromatographic system employed:
  • Column: Agilent Hi-Plex Ca, 7.7 × 300 mm, 8 μm particle size
  • Mobile phase: 100% deionized water
  • Flow rate: 0.6 mL/min
  • Column temperature: 85 °C
  • Injection volume: 20 μL
  • Sample concentration: 20 mg/mL of tequila sugars
  • Detector: Refractive index (RI)

No organic modifiers or salt additives were required, and isocratic operation simplified instrument maintenance.

Main results and discussion


The method achieved distinct peaks for six sugar components between 8.3 and 17.2 minutes. Key performance metrics included:
  • Retention times: raffinose (8.27 min), sucrose/maltose (9.13 min), lactose (9.45 min), glucose (10.92 min), fructose (13.43 min), mannitol (17.24 min)
  • Resolution: baseline separation with values exceeding unity for adjacent peaks
  • Plate counts: up to 43,750 theoretical plates for mannitol, indicating high column efficiency
  • Peak symmetry: asymmetry factors close to 1 for all analytes

The water-only mobile phase and elevated temperature optimized ligand-exchange interactions between calcium ions and hydroxyl groups, providing sharp, well-resolved peaks without interference from ethanol.

Benefits and practical applications


This protocol delivers reliable quantitation of sugars in tequila with minimal sample preparation and no complex gradient programming. It supports:
  • Quality assurance of agave spirit production
  • Verification of raw material authenticity and identification of adulteration
  • Routine monitoring in QA/QC laboratories

The simplicity of an isocratic method reduces downtime and reagent costs.

Future trends and possibilities


Emerging directions include coupling the Hi-Plex Ca method with mass spectrometric detection for enhanced sensitivity and structural confirmation. Miniaturized or high-throughput formats may enable real-time process monitoring in distilleries. Adaptation to other complex matrices, such as fermented foods or biofuels, could extend the application scope.

Conclusion


The Agilent Hi-Plex Ca ligand-exchange column paired with RI detection offers a robust, easy-to-operate solution for comprehensive sugar profiling in tequila. The method’s high efficiency, reproducibility and low operational complexity make it ideally suited to routine quality control in the beverage industry.

References


Stephen Ball and Linda Lloyd, Analysis of Tequila Carbohydrates, Application Note SI-01677, Agilent Technologies, 2011.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Agilent Hi-Plex Columns for Carbohydrates, Alcohols, and Acids
Analysis of the Isomeric Forms of Methyl-D-Glucopyranose
Analysis of the Isomeric Forms of Methyl-D-Glucopyranose
2011|Agilent Technologies|Applications
Agilent Hi-Plex Ligand-Exchange HPLC Columns
Agilent Hi-Plex Ligand-Exchange HPLC Columns
2011|Agilent Technologies|Brochures and specifications
Fruit Wine Analysis by HPLC
Fruit Wine Analysis by HPLC
2011|Agilent Technologies|Applications