Consumables, ICP/MS
IndustriesManufacturerAgilent Technologies, Savillex
Significance of the Topic
The Savillex C-Flow S-Type PFA nebulizers deliver enhanced sensitivity and rapid washout, crucial for trace element determination in inductively coupled plasma mass spectrometry (ICP-MS). Their low dead volume and demountable uptake lines streamline cleaning cycles and reduce cross-contamination, supporting high-throughput and low-sample-volume applications in industries such as semiconductors, environmental monitoring, and clinical analysis.
Aims and Study Overview
This technical evaluation compares three S-Type PFA concentric nebulizer variants (C50s, C100s, C250s) with nominal uptake rates of 50, 100, and 250 µL/min. The goal is to assess sensitivity, precision (RSD), oxide formation (CeO/Ce), and background interferences (80Ar2) under normal and cool plasma conditions, thereby determining their suitability for demanding analytical tasks.
Methodology and Instrumentation Used
- ICP-MS System: Agilent 7700s ICP-MS with quartz torch and Scott-type spray chamber at 2 °C.
- Normal Plasma Mode: 1500 W forward power, 13 mm sampling depth, 0.7 SLPM nebulizer gas, adjusted make-up gas for ~2 % CeO/Ce.
- Cool Plasma Mode: 600 W forward power, 18 mm sampling depth, optimized make-up gas to minimize 80Ar2 interferences.
- Uptake Rate Measurement: Gravimetric determination by aspirating water over time at constant gas flow.
Main Results and Discussion
- Uptake Rates: All models met specifications within ±20 % (C50s: 48 µL/min; C100s: 107 µL/min; C250s: 239 µL/min).
- Normal Plasma Sensitivity: C250s showed highest counts/sec across Li, Co, Y, Ce, Tl with RSDs ~1.5–3.5 %. C100s and C50s exhibited sensitivity 10 % lower but outperformed expectations based on uptake alone, indicating improved nebulization efficiency.
- Cool Plasma Performance: Sensitivity decreased for lower uptake rates due to reduced solvent loading and necessary gas flow adjustments; C100s and C250s achieved effective interference removal (59Co/80Ar2 ratio ≥1/2000), while C50s required relaxed criteria (80Ar2 <20 cps).
- Oxide Ratio: CeO/Ce maintained near 2 % across all configurations, demonstrating consistent plasma conditions.
Benefits and Practical Applications
The S-Type nebulizers combine high sensitivity, fast washout, and low sample consumption. Ideal applications include:
- Semiconductor impurity profiling.
- Limited-volume biomedical or forensic analyses.
- High-matrix samples where plasma loading must be controlled.
- Routine QA/QC with rapid method turnover.
Future Trends and Possibilities
Advancements may focus on integrating microfluidic sample introduction, novel low-temperature materials for enhanced chemical resistance, and automated gas-flow control for real-time optimization. The trend toward miniaturization and multiplexed ICP-MS systems could further leverage low-volume nebulizer designs.
Conclusion
Savillex C-Flow S-Type PFA nebulizers demonstrate superior sensitivity and precision with minimal dead volume. Their versatile performance in both normal and cool plasma modes, coupled with easy maintenance, positions them as valuable tools for modern ICP-MS workflows.
Instrumentation Used
- Agilent 7700s ICP-MS.
- Quartz torch and Scott-type spray chamber (2 °C conventional).
- Nebulizer gas flow controllers and analytical balance for uptake measurement.
References
- Savillex. C-Flow S-Type PFA Nebulizers Technical Note TN026, January 2020.