Analysis of Elemental Impurities in Pharmaceutical Products Following USP <232>/<233> on ICPMS-2030

Applications | 2018 | ShimadzuInstrumentation
ICP/MS
Industries
Pharma & Biopharma
Manufacturer
Shimadzu

Significance of the Topic


Elemental impurities in pharmaceutical products can pose serious risks to patient safety and drug efficacy. Since the traditional USP<231> colorimetric test lacks the sensitivity to detect toxicologically relevant concentrations, the implementation of USP<232>/<233> guidelines mandates precise trace-level analysis. ICP–MS has emerged as the method of choice due to its high sensitivity, low detection limits, and multi-element capability.

Objectives and Study Overview


The aim of this study was to establish and validate an ICP–MS method on the Shimadzu ICPMS-2030 for quantifying 24 elemental impurities in three dosage forms (tablet, capsule, liquid) in accordance with USP<232>/<233> and ICH Q3D guidelines. The study evaluates sample preparation, calibration strategies, method performance, and application to real drug samples.

Methodology and Used Instrumentation


Sample Preparation:
  • Weighed 0.4 g of each drug sample
  • Microwave digestion using a Milestone Ethos Easy system with a mixture of 65% HNO3, 37% HCl, and 30% H2O2 in closed vessels
  • Diluted digests to 50 mL with deionized water; internal standards (Sc, Y, Ho, Bi) added at 10 µg/L
  • Resulting dilution factor 125

ICP–MS Conditions:
  • Shimadzu ICPMS-2030 with AS-10 autosampler and Octaplate collision cell
  • Helium collision mode at 6.0 mL/min to reduce polyatomic interferences
  • RF power 1.20 kW; plasma gas 8 L/min; auxiliary gas 1.10 L/min; carrier gas 0.70 L/min
  • Mini-torch, specific nebulizer, and cyclone chamber at 5 °C


Main Results and Discussion


Calibration and Linearity:
Calibration standards at 0.5J, 1.0J, and 2.0J (derived from PDE values and maximum daily dose) exhibited correlation coefficients >0.9995 across all element classes, covering concentrations from low µg/L to high mg/L.

Method Performance:
  • Spike recovery for Drug 2 at 1.0J between 88% and 110%, within USP acceptance (70–150%)
  • Precision with RSD <2% for nearly all elements
  • LODs and LOQs well below required J-based limits, enabling higher dilution factors


Analysis of Real Samples:
All 24 elements in three drug samples were either below the detection limit or present at levels far below USP <232> PDE thresholds, confirming the method’s robustness and compliance.

Benefits and Practical Applications


  • High sensitivity and low detection limits ensure reliable trace analysis
  • Wide dynamic range allows single set of calibration standards
  • Helium collision mode supports HCl use in sample prep, improving element stability (e.g., Hg)
  • Efficient closed-vessel digestion prevents volatile losses
  • Low argon consumption and streamlined operation enhance cost-effectiveness


Future Trends and Opportunities


Advancements in automation and high-throughput sample preparation, coupling ICP–MS with separation techniques, real-time monitoring in manufacturing, and integration of data analytics and chemometrics are expected to further improve impurity profiling. Ongoing regulatory updates and harmonization will drive broader adoption of sensitive multi-element analysis in quality control.

Conclusion


The validated method on the Shimadzu ICPMS-2030 demonstrates excellent sensitivity, accuracy, and precision for quantifying 24 elemental impurities in pharmaceutical products, fully meeting USP <232>/<233> requirements. Its robust performance and operational efficiency support routine QC and regulatory compliance.

References


  1. United States Pharmacopeia General Chapter <232> Elemental Impurities – Limits
  2. United States Pharmacopeia General Chapter <233> Elemental Impurities – Procedures

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Validation of quantitative method for determination of elemental impurities in pharmaceutical products following USP 232/233 on ICPMS-2030
Analysis of Pharmaceutical Raw Materials Compliant with ICH Q3D Guideline Using ICP-MS with Collision/Reaction Mode
Analysis of Pharmaceutical Raw Materials Compliant with ICH Q3D Guideline Using ICP-MS with Collision Mode
Analysis of Artificial Tear Eye Drops For Elemental Impurities