LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
IndustriesPharma & Biopharma
ManufacturerAgilent Technologies
Importance of the Topic
The characterization of synthetic oligonucleotides is critical in pharmaceutical research, quality assurance, and quality control. High-performance liquid chromatography coupled with high-resolution mass spectrometry (LC/Q-TOF) provides both separation and accurate mass identification, enabling the detection of product variants and impurities without extensive sample preparation or contamination by ion-pairing reagents.
Objectives and Study Overview
This work demonstrates a hydrophilic interaction chromatography (HILIC) approach for intact mass and impurity analysis of diverse oligonucleotides. The method evaluates chromatographic performance, mass spectral quality, retention time reproducibility, ability to preserve higher-order structures such as siRNA duplexes, and detection of low-abundance impurities across DNA and RNA oligos ranging from 18- to 100-mers.
Methodology and Instrumentation
Ultrapure mobile phases based on ammonium acetate (15 mM, pH 7) and acetonitrile were employed with an Agilent InfinityLab Poroshell 120 HILIC-Z column (2.1×100 mm, 1.9 μm). Separation was performed on an Agilent 1290 Infinity II LC system and online MS detection used an Agilent 6545XT AdvanceBio Q-TOF with dual-spray JetStream source. Key conditions included:
- Gradient: 15–60% B over 13 min; flow rate 0.4 mL/min; column at 30 °C.
- Injection volume: 2–5 μL; autosampler at 4 °C.
- MS settings: negative polarity, 350 °C gas temp, 400 °C sheath gas, m/z 400–3200, 1 spectrum/s.
Main Results and Discussion
- Baseline separation of 20, 40, 60, 80, and 100-mers was achieved in a 15 min gradient, with efficient deconvolution of intact masses and identification of low-level depurination/depyrimidination products.
- Charge state distributions under HILIC were narrow and shifted to lower charge, consistent with native-like conditions. Larger oligos showed reduced ion intensities, highlighting the need for extended m/z range and optimized desolvation.
- Retention time reproducibility was excellent (<0.1% RSD) after two preconditioning injections, even with short (5–10 min) re-equilibration times.
- The method preserved higher-order structures: siRNA duplexes remained intact throughout analysis, and deconvoluted mass peaks matched duplex and individual strand masses.
- Low-abundance impurities down to <0.1% of full-length product were detected using untargeted and targeted deconvolution workflows.
- Complex mixtures with mixed bases at terminal positions were resolved by mass, even when chromatographic coelution occurred.
Benefits and Practical Applications
- Avoidance of ion-pairing reagents reduces instrument cleaning and enables rapid polarity switching.
- Robust performance for a broad range of oligonucleotide sizes, chemistries, and structures.
- High dynamic range MS detection supports comprehensive impurity profiling and relative quantitation.
- Fast throughput with reproducible retention times and minimal equilibration requirements.
Future Trends and Potential Applications
- Extension to longer RNA strands and heavily modified oligos for therapeutic development.
- Automation of HILIC-MS workflows for high-throughput quality control in oligonucleotide manufacturing.
- Integration with real-time data analysis and artificial intelligence for rapid impurity screening.
- Exploration of alternative zwitterionic and polymeric HILIC phases to further improve selectivity and robustness.
Conclusion
The described HILIC-LC/Q-TOF method offers a versatile, fast, and sensitive platform for intact mass analysis and impurity characterization of synthetic oligonucleotides. It combines excellent chromatographic resolution, native-like preservation of secondary structures, and high dynamic range mass detection, making it well suited for research and quality control applications.
Reference
- Rye P., Schwarzer C. MS1 Oligonucleotide Characterization Using LC/Q-TOF with HILIC Chromatography. Agilent Technologies, Application Note 2023.
- Lobue P. et al. Oligonucleotide Analysis by HILIC-MS without Ion-Pair Reagents. J. Chromatogr. A 2019, 1595, 39–48.
- Guimaraes R. et al. Impact of Mobile Phase pH in Oligonucleotide IP-LC-MS. Future Sci. OA 2021, FSO753.
- Guo H. et al. Secondary Structural Characterization of Oligonucleotides by ESI-MS. Nucleic Acids Res. 2005, 33(11), 3659–3666.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.