ICP/MS, ICP/MS/MS
IndustriesSemiconductor Analysis
ManufacturerAgilent Technologies
Significance of the topic
Accurate detection of metallic nanoparticles in semiconductor manufacturing reagents is critical to avoid wafer defects and yield loss. Even trace levels of iron-based particles can cause cone defects and electrical shorting, emphasizing the need for a sensitive, interference-free analytical technique.
Objectives and Study Overview
This study evaluates the use of single particle ICP-MS (spICP-MS) on an Agilent 8900 triple quadrupole ICP-MS operating in MS/MS mode to determine 25 and 30 nm Fe3O4 nanoparticles in low-particle concentration solutions of isopropyl alcohol (IPA), propylene glycol methyl ether acetate (PGMEA), and butyl acetate (BuAc). Key goals included size characterization, particle number concentration measurement, and assessment of long-term stability.
Methodology
Sample Preparation:
- Fe3O4 nanoparticle standards (25 nm and 30 nm) spiked into IPA, PGMEA, and BuAc.
- Preparation of ionic Fe standards by diluting a 1000 ppm Fe solution in each solvent for calibration of ionic sensitivity.
- Direct introduction of organic solvents into the ICP-QQQ with oxygen addition to the sample gas to prevent carbon deposition.
- Operation in MS/MS mode with Q1 and Q2 both set to m/z 56 to select Fe ions and remove spectral interferences using NH3 reaction chemistry.
- Fast time resolved analysis at 100 µs dwell time over 60 s acquisition to capture single-particle signals.
- Data processed using the Single Nanoparticle Application Module in MassHunter software for threshold setting, size, and count determination.
Used Instrumentation
- Agilent 8900 Triple Quadrupole ICP-MS in semiconductor configuration with MS/MS operation.
- Agilent SPS 4 autosampler with continuous rinse system and PFA nebulizer (self-aspiration).
- Quartz spray chamber and torch with 1.5 mm i.d. injector and platinum-tipped cones.
- Option gas mass flow controllers delivering O2 (20% in Ar) and NH3 (10% in He) into the collision/reaction cell.
Main Results and Discussion
- 30 nm Fe3O4 nanoparticles spiked at 5 ppt in IPA, PGMEA, and BuAc show clear separation of particle signals from background, with measured mean sizes around 30 nm.
- Nebulization efficiency calculated by size was approximately 30% across all solvents.
- Ionic Fe sensitivity ranged from 1500 to 2400 cps/ppt; background equivalent diameter of blanks was around 6 nm.
- Linearity (R2 = 0.998) demonstrated for 25 nm Fe3O4 nanoparticles spiked between 0.1 and 2 ppt in IPA.
- Long-term stability tests over 12 hours showed consistent particle number concentration and size (%RSD within acceptable limits) in all three solvents.
Benefits and Practical Applications
This spICP-MS method on the Agilent 8900 provides low detection limits, high sensitivity, and effective removal of spectral interferences, making it suitable for quality control of semiconductor-grade chemicals and ensuring reliable device fabrication.
Future Trends and Potential Applications
- Expansion to multi-element nanoparticle analysis by leveraging QQQ capabilities for simultaneous detection of various metal NPs.
- Integration with inline process monitoring to detect nanoparticle contamination in real time.
- Application to emerging nanomaterials and other industrial reagents requiring high-purity specifications.
Conclusion
The study confirms that spICP-MS using an Agilent 8900 ICP-QQQ in MS/MS mode effectively determines and characterizes iron oxide nanoparticles in low-concentration organic solvents, with reliable sensitivity, linearity, and temporal stability.
References
- Hagiwara T. et al. Study on cone-defects during the pattern fabrication process with silicon nitride. Journal of Photopolymer Science and Technology. 2015;28(1):17–24.
- Yamanaka M., Wilbur S. Measuring multiple elements in nanoparticles using spICP-MS. Agilent Technologies. Publication 5994-0310EN.
- Shimamura Y., Hsu D., Yamanaka M. Multielement nanoparticle analysis of semiconductor process chemicals using spICP-QQQ. Agilent Technologies. Publication 5994-0987EN.
- Yamanaka M., Wilbur S. Accurate determination of TiO2 nanoparticles in complex matrices using the Agilent 8900 ICP-QQQ. Agilent Technologies. Publication 5991-8358EN.