ICP/MS, ICP/MS/MS
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Significance of Topic
Chloride in crude oil can induce corrosion and fouling in refinery equipment. Maintaining chloride levels below 1 mg/L is essential for safe and efficient distillation and to extend equipment lifespan.
Objectives and Study Overview
This work aimed to establish a rapid and accurate method for determining chloride in crude oils by direct dilution in an o-xylene based diluent using an Agilent 8900 ICP-QQQ. The method was validated with a NIST fuel oil reference material and compared against Instrumental Neutron Activation Analysis (INAA).
Methodology and Instrumentation
- Sample Preparation: Crude oils diluted 1:5 or 1:10 in a diluent of 90% o-xylene and 10% mineral oil modifier with dispersant and internal standards (Sc and Y at 0.1 mg/kg).
- Calibration: Organic chloride standards in the range 1–1000 mg/kg; blank runs with diluent and modifier.
- ICP-QQQ Conditions: Agilent 8900 with platinum cones, concentric glass nebulizer, quartz spray chamber and torch; H2 cell gas for mass-shift of 35Cl to 37Cl as ClH2+; RF power 1500 W, nebulizer flow 0.4 L/min, H2 flow 4.6 mL/min.
- Interference Control: MS/MS mode to remove 16O18O H and 34S H polyatomic overlaps.
- Comparison Technique: INAA performed by an external facility.
Main Results and Discussion
- Analytical Performance: Detection limit of 0.01 mg/kg, LOQ of 0.04 mg/kg and background equivalent concentration of 0.24 mg/kg for 35Cl.
- SRM Recovery: 99% at 1:5 dilution and 107% at 1:10 dilution versus the 45 mg/kg reference.
- Crude Oil Analysis: Twelve samples with sulfur 0.38–2.01 wt% and H/C ratios 1.68–1.88 yielded chloride levels between 6 and 50 mg/kg.
- INAA Comparison: Agreement from 89% to 134%, with sample heterogeneity and incomplete dissolution cited for discrepancies.
Benefits and Practical Applications
- Fast, sensitive chloride determination in complex petroleum matrices.
- Effective interference removal using H2 mass-shift in MS/MS mode.
- Direct dilution reduces acid consumption and simplifies sample handling.
- Applicable for routine QA/QC and corrosion monitoring in refineries.
Future Trends and Opportunities
- Extension to other halogens and trace elements in petroleum by ICP-QQQ.
- Automated dilution and flow injection for higher sample throughput.
- Integration with multivariate analysis for crude oil fingerprinting and source tracking.
- Development of portable or miniaturized triple-quadrupole ICP systems for field analysis.
Conclusion
The Agilent 8900 ICP-QQQ method with H2 mass-shift enables accurate, low-level chloride measurement in crude oils via direct dilution. The approach delivers high recoveries, low detection limits, and robust interference control, supporting refinery quality control and corrosion risk assessment.
References
- Nelson J, Poirier L, Lopez-Linares F. Determination of Chloride in Crude Oils by Direct Dilution using Inductively Coupled Plasma Tandem Mass Spectrometry (ICP-MS/MS). J Anal At Spectrom. 2019; DOI 10.1039/C9JA00096H.
- ASTM D8110-17. Standard Test Method for Elemental Analysis of Distillate Products by Inductively Coupled Plasma Mass Spectrometry (ICP-MS). ASTM Int. 2017.
- Agilent Technologies. Handbook of ICP-QQQ Applications using the Agilent 8800 and 8900. Publication 5991-2802EN.
- Sugiyama N. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS. Agilent 5991-2303EN.
- Nakano K. Ultra low-level determination of phosphorus, sulfur, silicon, and chlorine using the Agilent 8900 ICP-QQQ. Agilent 5991-6852EN.
- Doyle A, Saavedra A, Tristao MLB, Nele M, Aucelio RQ. Direct chlorine determination in crude oils by energy dispersive X-ray fluorescence spectrometry: An improved method based on a proper strategy for sample homogenization and calibration with inorganic standards. Spectrochim Acta B. 2011;66:368–372.