AAS
IndustriesEnvironmental
ManufacturerAgilent Technologies
Importance of the Topic
In semiconductor manufacturing, trace metal contaminants in ultrapure rinse water can lead to electrical shorts and reduced device yield. Continuous monitoring of sodium, calcium and silicon at sub-ng/mL levels is critical for maintaining process integrity and product reliability.
Objectives and Study Overview
This application note applies graphite furnace atomic absorption spectroscopy (GFAAS) to quantify sodium, calcium and silicon in ultrapure water. The study aims to establish sensitive calibration procedures, determine detection limits under standard protocols (IUPAC and DIN 38 402), and evaluate reproducibility, including the effect of multiple injections.
Methodology and Instrumentation
Instrumentation:
- Atomic absorption spectrometer: Agilent SpectrAA-30A
- Graphite furnace atomizer: GTA-96 with PSD-96 programmable sampler
- Hollow cathode lamps for Na, Ca, Si
- Argon (99.999%) as purge gas
- Ultrapure water system: ion-exchange filter (Servo Mischbettfilter) and Nanopure Reinstwasseranlage D-1797
Results and Discussion
Calibration for sodium and calcium was linear (r>0.996) with RSDs below 9%. Silicon required a second-order polynomial fit due to non-linearity. IUPAC detection limits were 0.06 ng/mL (Na), 0.13 ng/mL (Ca) and 2.1 ng/mL (Si). DIN 38 402 limits for Na and Ca were 0.05 ng/mL and 0.20 ng/mL, respectively. Multiple injections enhanced signal proportionally, suggesting a 10–20× improvement in detection capability.
Benefits and Practical Applications
The described GFAAS method delivers high sensitivity and reproducibility for ultratrace analysis in ultrapure water. Automated sampling and multiple injection features support routine QA/QC in semiconductor fabs, reducing manual handling and increasing throughput.
Future Trends and Potential Applications
Future developments may focus on automated multi-element furnace sequences, tighter integration with inline process control, and further sensitivity gains via advanced chemical modifiers or laser-based atomization. Extending the approach to additional critical metals (e.g., iron) will broaden its utility in high-purity applications.
Conclusion
A robust GFAAS protocol has been validated for measuring sodium, calcium and silicon in ultrapure rinse water at sub-ng/mL levels. Optimized furnace programs, chemical modification and multiple injection strategies achieve the precision and detection limits required for semiconductor process monitoring.
Reference
- K. Marquardt, Marine Technology 16(4), 155–160 (1985)
- W. Frech, A. Cedergren, Anal. Chim. Acta, 113, 227–235 (1980)
- H. Kaiser, Z. Fresenius Anal. Chem., 209, 1–18 (1965)
- H. Kaiser, Z. Fresenius Anal. Chem., 216, 80–93 (1966)
- G. L. Long, J. D. Winefordner, Anal. Chem. 55, 713A–724A (1983)