Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS

Technical notes | 2007 | Agilent TechnologiesInstrumentation
ICP/MS
Industries
Environmental
Manufacturer
Agilent Technologies

Importance of the Topic


The determination of trace elements in environmental samples is critical for regulatory compliance, pollution monitoring, and public health protection. Traditional multi-mode ICP-MS methods require complex optimization and multiple cell conditions, limiting throughput and increasing operational complexity. The Agilent 7500cx equipped with an Octopole Reaction System (ORS) operating in a single helium collision mode addresses these challenges by simplifying interference removal and dramatically improving sample throughput.

Objectives and Study Overview


This study evaluates the capability of the Agilent 7500cx ICP-MS in helium mode to:
  • Analyze a complete suite of environmental elements, including mercury and major matrix constituents, in a single analysis mode.
  • Eliminate or minimize polyatomic matrix interferences without multiple cell gas exchanges.
  • Achieve high throughput (less than 2.5 minutes per sample) across a 12-hour, 300-sample sequence while maintaining accuracy and precision.

Methodology


The analytical protocol involved:
  • An initial calibration with blank and standards at 1, 10, 50 and 100 ppb (1 ppb for Hg).
  • A sequence of 22 repeated blocks, each containing:
    • Certified reference water (NIST 1640).
    • USEPA Interference Check Solution (ICS-A) and a modified spike version (ICS-AB) at 20 ppb.
    • Two high total dissolved solids mineral water samples.
    • Continuing calibration blank (CCB) and continuing calibration verification (CCV) samples after each block.
  • No mode switching or recalibration was performed during the 12-hour run.

Instrumentation Used


This evaluation employed a standard Agilent 7500cx ICP-MS with:
  • Glass concentric nebulizer (MicroMist) and quartz 2.5 mm torch.
  • Nickel sample and skimmer cones.
  • ORS operated exclusively in helium collision mode at 5.0 mL/min.
  • Plasma conditions: 1550 W RF power, 0.80 L/min carrier gas, 0.23 L/min makeup gas, and 8.5 mm sampling depth.
  • Integration time of 0.1–0.3 seconds per isotope, one point each.

Main Results and Discussion


Interference Removal and Sensitivity:
  • Helium mode reduced background equivalent concentrations (BECs) and detection limits (IDLs) for key isotopes by an order of magnitude compared to no-gas mode, effectively removing Ar-, Ca-, S- and Cl-based polyatomic overlaps.
  • Sub-ppt IDLs achieved for most elements despite minor signal losses from ion–gas collisions.

Calibration Stability and Precision:
  • Fifty ppb CCV measurements over 25 replicates remained within ±10% of target, with no calibration failures.
  • ICS-AB spike recoveries (20 ppb, 1 ppb Hg) across 44 replicates ranged from 97 to 104%, with RSDs below 5%.
  • NIST 1640 certified water analyses (n=44) yielded recoveries of 92–99% for major and trace elements, RSDs typically under 5%.

Throughput and Run-to-Run Times:
  • Single helium mode eliminated cell evacuation and repressurization steps required by multi-mode CRC systems.
  • Pre-emptive and intelligent rinse functions reduced idle time between samples.
  • Measured run-to-run time averaged 2.46 minutes per sample over 300 analyses.

Benefits and Practical Applications


The universal helium mode offers:
  • Consistent removal of polyatomic interferences without prior knowledge of sample matrix.
  • Simplified tuning and method development by avoiding multiple cell gases.
  • High productivity suitable for contract environmental and industrial laboratories requiring rapid turnaround.
  • Reliable quantification of mercury, major and trace elements in high-matrix matrices with minimal carryover.

Future Trends and Applications


Advances in collision cell design and software are expected to further reduce analysis times and improve sensitivity. Potential developments include:
  • Integration of real-time interference monitoring and adaptive collision gas control.
  • Expanded use of helium-only modes for food, clinical and geochemical analyses.
  • Further miniaturization and automation of sample introduction to support high-throughput screening and on-site environmental monitoring.

Conclusion


The Agilent 7500cx ICP-MS operating exclusively in helium ORS mode delivers rapid, interference-free multi-element analysis for high-matrix environmental samples. By combining robust interference removal with pre-emptive rinse and stable calibration, the system achieves routine sample throughput under 2.5 minutes, maintaining excellent accuracy and precision over extended sequences.

References


  • Wilbur S. Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS, Application Note 5989-7297EN, Agilent Technologies, 2007.
  • Agilent Application Note 5989-6663EN, Performance Characteristics of the Agilent 7500cx ICP-MS.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Agilent 7900 ICP-MS simplifies drinking water analysis
Agilent 7900 ICP-MS simplifies drinking water analysis
2014|Agilent Technologies|Applications
Maximizing Productivity in High Matrix Samples using the Agilent 7700x ICP-MS with ISIS Discrete Sampling
High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System – Discrete Sampling (ISIS–DS)
Agilent ICP-MS Journal (June 2007 – Issue 31)
Agilent ICP-MS Journal (June 2007 – Issue 31)
2007|Agilent Technologies|Others