Acidity in volatile solvents and chemical intermediates

Applications | 2020 | MetrohmInstrumentation
Titration
Industries
Energy & Chemicals
Manufacturer
Metrohm

Importance of Topic


The acidity level in volatile organic solvents directly impacts their storage stability and can cause material corrosion in industrial equipment. Accurate monitoring of acid content is essential to prevent asset damage and process interruptions in chemical manufacturing and quality control environments.

Objectives and Overview of the Study


This study demonstrates the determination of very low acid concentrations in solvents such as 4-methyl-2-pentanone (MIBK), 2-ethoxyethyl acetate, and 2-ethoxyethanol. The aim is to apply ASTM D1613 via an automated photometric titration system to deliver reliable and reproducible acidity measurements.

Methodology and Instrumentation


Solvent samples are titrated with standardized sodium hydroxide in the presence of a phenolphthalein indicator. The end point is detected photometrically using an Optrode sensor. Water-insoluble samples are dissolved in CO₂-free ethanol, while water-compatible solvents use deionized water.

Used Instrumentation


  • OMNIS Advanced Titrator with magnetic stirrer for endpoint titration (ASTM D1613).
  • Optrode optical sensor with selectable wavelengths for objective photometric detection.
  • OMNIS Sample Robot and Dosing Module for automated sample handling and reagent delivery.

Main Results and Discussion


Repeated analyses (n=6) yielded mean acidity values of 0.0151% for MIBK, 0.012% for 2-ethoxyethyl acetate, and 0.0003% for 2-ethoxyethanol. Relative standard deviations ranged from 0.4% to 5.2%, demonstrating high precision even at trace acid levels. The Optrode’s objective end point recognition improves reproducibility compared to visual indication.

Benefits and Practical Applications


  • Enhanced accuracy in determining low-level acidity in volatile solvents.
  • Reduced operator subjectivity due to photometric end point detection.
  • Increased throughput and automation reduce labor and potential errors.
  • Versatile wavelength selection permits adaptation to diverse titration chemistries.

Future Trends and Potential Applications


The flexible Optrode sensor is suited for expanding applications such as nonaqueous titrations in pharmaceutical and petrochemical analyses, and monitoring of functional end groups. Ongoing developments may integrate multichannel parallel titration and advanced data analytics for real-time process monitoring.

Conclusion


The automated OMNIS system paired with photometric indication meets ASTM D1613 requirements for acidity determination in volatile solvents, achieving reliable, low-level acid quantification. The approach offers significant improvements in precision and operational efficiency for industrial and laboratory settings.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Acidity in aviation turbine fuel according to ASTM D3242
Determination of Lauryl Sulfate
Determination of Lauryl Sulfate
2020|Metrohm|Applications
Photometric titration of acid value in biodiesel according to EN 14104
Kjeldahl nitrogen in waste water
Kjeldahl nitrogen in waste water
2020|Metrohm|Applications