Significance of the Topic
Per and polyfluoroalkyl substances PFAS are persistent environmental contaminants linked to serious health concerns They accumulate in ecosystems and can be transported over long distances in the atmosphere Understanding and monitoring PFAS levels in ambient and indoor air is essential for assessing human exposure and guiding regulatory strategies
Objectives and Study Overview
This work aims to extend and adapt the US EPA draft method OTM 45 originally designed for source emission sampling to ambient and indoor air contexts The study evaluates accelerated solvent extraction ASE for sample preparation assesses analytical performance metrics and demonstrates a streamlined workflow combining sampling resin cleanup and extraction in a single apparatus
Methodology and Instrumentation
The sampling system follows the OTM 45 design integrating a particulate filter XAD 2 resin trap and impingers to capture PFAS compounds from C4 to C18 classes A compact sampler housing 8 grams of ultra clean resin and a 20 micrometer frit was constructed for ambient and indoor sampling Extraction of the resin fraction was performed using a Dionex ASE 350 with a 4 to 1 methanol acetonitrile mixture under 1500 psi at 120 degrees C Two sequential static cycles of 15 minutes each yielded extracts for analysis HPLC MS MS analysis utilized a Shimadzu Nexera X2 system with an 8045 tandem quad detector and a Restek Force C18 column 50 by 2.1 millimeters operated at 40 degrees C The gradient elution runs in 8.5 minutes with water ammonium acetate and methanol phases
Main Results and Discussion
Accelerated solvent extraction achieved 1.4 to 55 times greater analyte response and reduced sample preparation time from 32 hours to under one hour compared to traditional shakeout approaches ASE met accuracy criteria within 30 percent for 37 of 49 target PFAS while shakeout met it for 34 ASE precision results showed relative standard deviations below 20 percent for 40 compounds versus 35 for shakeout Method detection limits calculated for a 3 cubic meter sample volume met or improved on published OTM 45 values for all but two analytes Resin cleaning by ASE yielded blank levels below detection limits across compounds in the range of 0.01 to 0.32 nanograms per gram of resin Field tests collecting nearly 30 liters of air over 16.5 hours showed no breakthrough of target PFAS or volatile precursors such as fluorotelomer alcohols
Benefits and Practical Applications
- Significantly reduced extraction time and solvent consumption enables higher throughput
- Improved extraction efficiency enhances sensitivity for trace level PFAS
- Integrated sampling and extraction workflow simplifies laboratory procedures
- Adaptable to both ambient outdoor and indoor air monitoring scenarios
Future Trends and Opportunities
Further method validation will focus on recovery precision and detection limits specific to ambient and indoor matrices Expanded breakthrough studies will ensure resin capacity under varying environmental loads Real world sampling campaigns can provide spatial and temporal PFAS distribution data in urban rural and indoor settings Emerging instrumentation such as high resolution mass spectrometry may offer deeper insight into novel PFAS species
Conclusion
The adaptation of the OTM 45 method with accelerated solvent extraction and a compact resin based sampler offers a robust approach for quantifying PFAS in ambient and indoor air This streamlined technique meets performance criteria for accuracy precision and detection limits while facilitating practical laboratory implementation