Analysis of contaminants in hemp using LC and GC coupled to MS/MS

Posters | 2022 | Restek | ASMSInstrumentation
GC/MSD, GC/MS/MS, GC/QQQ, GC columns, Consumables, LC/MS, LC/MS/MS, LC columns, LC/QQQ
Industries
Food & Agriculture
Manufacturer
Restek

Importance of the topic


Hemp, a Cannabis sativa variety low in THC and rich in CBD and CBG, can accumulate pesticides and mycotoxins that pose health risks. Analytical methods capable of detecting contaminants at parts-per-billion levels are essential to ensure product safety and regulatory compliance in both industrial and consumer applications.

Objectives and study overview


This work presents a streamlined workflow for simultaneous extraction and cleanup of hemp samples followed by analysis of a broad panel of pesticides and mycotoxins using LC-MS/MS and GC-MS/MS. The goal was to achieve low limits of quantification, high accuracy, and precision that meet or exceed state and country regulatory requirements.

Methodology and instrumentation


Sample preparation:
  • Grind 1 g hemp; spike with isotopically labeled standards.
  • Extract with 5 mL acetonitrile + 1 % acetic acid; vortex and centrifuge.
  • Cleanup via HLB SPE cartridge with water pre-equilibration, vacuum elution, and methanol rinse.
  • For GC: further clean extract using dSPE tubes containing MgSO₄ and C18; centrifuge.

LC-MS/MS analysis:
  • Instrument: Shimadzu LCMS-8045 with ESI source.
  • Column: Raptor ARC-18 (2.7 µm, 150 × 2.1 mm) with EXP guard.
  • Mobile phases: water and methanol each containing 2 mM NH₄HCOO and 0.1 % formic acid; gradient over 19.6 min.
  • Injection: 1.5 µL of sample diluted 1:1 with methanol:acetonitrile:water (1 % acetic acid).

GC-MS/MS analysis:
  • Instrument: Thermo Trace 1310-TSQ 8000 with EI source.
  • Column: Rxi-5ms, 30 m × 0.25 mm × 0.25 µm.
  • Oven program: 70 °C to 220 °C at 30 °C/min, to 240 °C at 5 °C/min, to 315 °C at 10 °C/min.
  • Injection: 1 µL splitless, inlet at 250 °C, carrier He at 1.4 mL/min.

Results and discussion


The method achieved LOQs well below 0.1 µg/g for most analytes, with calibration curves (0.005–1.5 µg/g) showing R² > 0.99 for almost all compounds. Accuracy ranged between 70–130% and precision (RSD) was under 30%. The water-methanol dilution and HLB cleanup effectively removed hydrophobic interferences common in hemp, and the MgSO₄ dSPE step eliminated residual moisture. A total of nine deuterated standards compensated for matrix and procedural variations.

Benefits and practical applications


  • Single extraction covers both LC- and GC-amenable analytes.
  • Rapid sample preparation with effective SPE and dSPE steps.
  • Robust quantification meeting California and Canadian regulatory limits.
  • Applicability to QA/QC laboratories, research, and industrial testing.

Future trends and potential applications


Advancements may include high-resolution MS for non-targeted screening, automated sample handling, miniaturized extraction devices, and online SPE-LC/MS workflows. Integration of machine learning for data processing and cloud-based monitoring could further enhance throughput and data reliability.

Conclusion


A comprehensive and efficient workflow was developed for the simultaneous analysis of a wide range of pesticides and mycotoxins in hemp. The method offers low detection limits, excellent accuracy, and precision, fulfilling stringent regulatory criteria and facilitating routine quality control.

References


  • Reyes-Garcés N, Myers C. J Sep Sci. 2021;44:2564–2576.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 